Remote Display (Option) Test
4-16
Maintenance: Checkout Procedures
1. Verify the docking station is connected to the remote display.
2. Power up the monitor and the remote display.
3. From the monitor, select MORE MENUS > PATIENT DATA >
VITAL SIGNS.
4. Verify the Vital Signs NRT displays correctly on the remote display.
See
"Adjust the Remote Display (Option) Video Signal"
Dash Port 2 Docking Station
2000966-138
on page 3-12.