Medivators VERISCAN LT User Manual

Automated endoscope leak tester

Advertisement

LT
User Manual
Automated Endoscope
Leak Tester
REPROCESSING SYSTEMS

Advertisement

Table of Contents
loading

Summary of Contents for Medivators VERISCAN LT

  • Page 1 User Manual Automated Endoscope Leak Tester REPROCESSING SYSTEMS...
  • Page 2 Minntech Corporation. Minntech reserves the right to make changes in the specifications shown herein without notice or obligation. Contact your Medivators representative or Medivators customer service for more information.
  • Page 3: Table Of Contents

    Table of Contents Introduction ..............5 Using this Manual ............5 Product Description ............ 5 Product Specifications ..........6 Safety ................7 Intended Use .............. 7 Electrical Compliance ..........7 Operator Safety ............7 General Cautions ............7 Guidelines ..............8 Cleaning and Disinfection ........
  • Page 5: Introduction

    Flexible endoscopes are highly vulnerable to damage-causing moisture that enters the endoscope’s sensitive interior through holes, tears, broken seals, cleaning cap leaks and loose contact pins. Prior to Veriscan LT, leak detection has always been hampered by testing equipment that lacked precision, and required extraordinary skills of observation beyond the capabilities of even the most experienced cleaning technician.
  • Page 6: Product Specifications

    Product Specifications Power Requirement: Universal Power Supply 100-240VAC, 50/60 Hz, .25 Ampere Fuse Rating: 250V, 1 Ampere, Type S, 3AG style. Dimensions: 13.125 H x 8.25 W x 19.25 L inches (33.3 H x 21 W x 48.9 L cm) Weight: 15.8 lbs.
  • Page 7: Safety

    Safety Intended Use The Medivators Veriscan system is intended for endoscope leak testing of flexible, immersible endoscopes. The Medivators Veriscan system has been validated by Medivators for use on Olympus, Pentax, Fujinon, and Storz flexible, immersible endoscopes. Electrical Compliance The Medivators Veriscan system was tested to stringent electrical certifications and safety standards. This device complies to UL Standard 61010-1 and CAN/CSA Standard 61010-1.
  • Page 8: Guidelines

    Guidelines Endoscope Leak Testing and Pre-cleaning An effective endoscope leak test and pre-cleaning process is an essential step in the reprocessing of an endoscope. Always follow the endoscope manufacturer’s instructions and established professional guidelines to properly leak test and clean an endoscope. If an endoscope leak is detected, the endoscope should be immediately taken out of service and sent in for repair.
  • Page 9: System Components

    • One (1) ProV-LT Link Database Software • One (1) Quick Start Guide • One (1) Veriscan Users Manual If any of the components are missing, contact Medivators Customer Support at 1-800-444-4729 or your authorized distributor. USERS MANUAL | 9...
  • Page 10: System Features

    System Features Display Screen Displays information and test results Control Keypad Touchpad used to operate Veriscan System Printer Provides printout of test results Quick Connect Port Veriscan connector for Quick Connect Tubing to endoscope Power Switch Used to turn Veriscan system ON/OFF Power Cord Outlet Receptacle for AC power cord Power Cord...
  • Page 11: System Setup

    System Setup Installation • Position Veriscan system on a flat, level surface and within six (6) feet/ two (2) meters of the electrical AC wall outlet. • Veriscan’s electrical requirements can be found under the “Introduction, Product Specifications” section. Attaching the Power Cord •...
  • Page 12 If replacement or additional tubings are needed for connecting other brands of endoscopes, contact Medivators Customer Support at 1-800-444-4729 or your authorized distributor. NOTE: Inspect Quick Connect Tubings often and replace if damaged or worn.
  • Page 13 System Setup Step 4: Push the metal housing release button down and hold. (see figure 5) Step 5: While continuing to hold the button, gently pull down on the metal bulb housing and rotate it to the left, until it releases down and away from the machine’s bulb housing structure.
  • Page 14: Performing A Leak Test

    Performing a Leak Test Attaching a flexible, immersible endoscope to the Veriscan System • Ensure the Quick Connect Tube is connected to the Veriscan system. If not, refer to the System Setup section. • Check the endoscope cleaning cap to ensure it is securely in place over the endoscope’s ETO connector.
  • Page 15 Performing a Leak Test • The leak test will now automatically begin when Veriscan is operating in the factory default “STANDARD TEST” mode, and the following information will be displayed: Leak Test Start Please standby... • If Veriscan is operating in the “MANUAL” mode, the following information will display: (The manual mode is used for selecting alternative parameter settings, other than the factory default “STANDARD TEST”...
  • Page 16 Performing a Leak Test Disconnecting the Endoscope • Disconnect the Quick Connect Tubing from the endoscope (see figure 17). The endoscope is now ready for the next step in the manual cleaning process. Shut Down and Start Up of the Veriscan system •...
  • Page 17: Performing A Screening Test

    Performing a Screening Test Veriscan’s “Screening Test” is an extremely accurate, dry endoscope leak test which is ideal for evaluating “cold” or stable temperature endoscopes. Use the “Screening Test” to: 1) Evaluate an endoscope that was not used on a patient within the last 30 minutes. 2) Evaluate a facility’s inventory of endoscopes.
  • Page 18 • If Veriscan is operating in the “MANUAL” mode, the following information will display: (The manual mode is used for selecting alternative parameter settings, other than the factory default “SCREENING TEST” settings.) Select Test Setting Screen Test 1 Screen Test 2 Screen Test 3 Use the up/down arrow keys to select the desired test, and then press “CONTINUE”...
  • Page 19: Testing A Failed Endoscope

    Testing a Failed Endoscope Majority of FAIL leak test results are due to true endoscope leaks. On occasion, a FAIL test result can be traced to other preventable and correctable issues such as: • Nicks in the endoscope cleaning cap gasket: If the rubber seal or gasket on the inside of the endoscope cleaning cap is damaged, a false FAIL leak test result can occur.
  • Page 20 2) Endoscope Distal End Does Not Inflate: Do not submerge the endoscope in fluid. This is an indication of a large endoscope leak. Press CANCEL to stop the constant air flow. The endoscope must be hand cleaned before being sent in for repair by following these steps: •...
  • Page 21: Daily Quality Assurance Test

    This test ensures the internal integrity of the Veriscan unit and its ability to properly sense an endoscope leak. Contact Medivators Technical Support or your authorized Medivators distributor if the Daily Quality Assurance Test fails.
  • Page 22: Data Access And Management

    Data Access and Management Main Menu Veriscan can be used as a productive tool for data management, asset tracking, equipment utilization and infection control. Additional system capabilites are available through the “MENU” key pad and the ProV-LT data management software. See “ProV-LT Data Management Software” located in the “Data Access and Management”...
  • Page 23 • Use down arrow to toggle to “Yes” and press CONTINUE • Press the CANCEL button until the “Connect Scope” language is displayed is on the screen. Using MENU button for setting Automatic or Manual Test Mode The Veriscan system is normally used in the Automatic Test Mode which uses factory default parameter settings for endoscope leak testing.
  • Page 24 If the Veriscan is operating in the “Manual” test mode and needs to be changed back to the “Automatic” test mode, then follow the steps listed under “Activate the Automatic or Manual Test Mode by” to select the Automatic mode. Contact Medivators Technical Support or your authorized Medivators distributor if further assistance is required. 24 | USERS MANUAL...
  • Page 25 Veriscan Warranty. Contact Medivators Technical • Press CONTINUE to select this option Support at 1-800-444-4729 or your authorized Medivators distributor if you have questions about this • Arrow down to select the “delete all” option and to area. purge all past test data. To delete all stored runs, press the CONTINUE button and the program will begin the deletion process.
  • Page 26: Printing And Quick Reports

    Data Access and Management Printing and Quick Reports The Veriscan system features a compact thermal printer that allows the operator to: 1) Print the leak test results of any test upon completion. 2) Print quick reports summarizing daily, weekly, and monthly results. 3) Print quick reports summarizing all endoscopes that failed a leak test within the last 30 days.
  • Page 27 Data Access and Management • To print a summary of the week’s procedures, use the up/down arrow keys to select “Previous 7 day summary”. When the option is selected, the printer will automatically print the previous 7 day’s report which includes: •...
  • Page 28: Prov-Lt Data Management Software

    Excel™ file. All downloaded data is not automatically ® erased from Veriscan. Data files must be erased from Veriscan if no longer needed. Contact Medivators Technical Support for further information. The screen will display “downloading data” with an active progress bar.
  • Page 29 (yes or no). Test Parameter Settings This data refers to the technical setting information that a Medivators Service Technician would access in the event of a service call. USERS MANUAL | 29...
  • Page 30: Maintenance, Handling And Care

    When the crystals turn a vivid bubble gum pink color, it is recommended to ensure you have a replacement bag of drying crystals available or on order from Medivators. As the crystals continue to absorb moisture, their color continues to change until they reach a pale pink which indicates it is time to change the crystals.
  • Page 31 The Veriscan system uses a Littlefuse PN 31300, which is a 3AG style fuse. The fuse rating is: 250V, 1 Ampere, Type S. Contact Medivators Technical Support at 1-800-444-4729 or your authorized distributor to order a new fuse or for technical assistance.
  • Page 32 Maintenance, Handling and Care Changing Printer Paper The printer will automatically detect when the thermal paper roll has run out and report this condition with a flashing light on the LED display. To replace the printer paper follow these steps: Step 1: Pull the lever on the front of the printer paper access door upwards until the lid is released from its locked position (see figure 30).
  • Page 33: Troubleshooting

    Turn the Veriscan system off. Wait 15 seconds and then turn it back on. (see figure 41). Figure 34 • If the sytem still does not power on, contact your Medivators service representative. For all other Veriscan performance inquiries, contact your local representative or Medivators Technical Support at 1-800-444-4729.
  • Page 34: Warranty

    Limited Warranty Information Limited Warranty Medivators Reprocessing Systems, a division of Minntech Corporation (the “Company”), warrants that the endoscope reprocessor and associated equipment it manufactures (collectively, the “Products”) will conform to the Company’s written specifications and will be free from defects in material and workmanship under normal use and service for a period (the “Warranty Period”) of fifteen (15) months from date of shipment from the Company or one (1) year from...
  • Page 36 Distributed in Canada by: CARSEN MEDICAL INC. 151 Amber Street Unit 17 & 18 Markham, ON L3R 3B3 1-866-677-4122 © 2009, Minntech Corporation 36 | USERS MANUAL 50096-988 Rev A...