Download Print this page
Yamato VBE204 Instruction Manual

Yamato VBE204 Instruction Manual

Vented balance enclosure

Advertisement

Quick Links

VENTED BALANCE ENCLOSURE
Model
VBE204/214/306/316/408/418/600/610
 Thank you for purchasing the Vented Balance
 To use this unit properly, read this "Instruction
Instruction Manual
– December 2019 –
Enclosure, VBE204/214/306/316/408/418/600/610,
of Yamato Scientific.
Manual" thoroughly before using the unit.
Keep this instruction manual around this unit for
reference at any time.
WARNING:
Carefully read and thoroughly understand the
important warning items described in this
manual before using this unit.

Advertisement

loading

Summary of Contents for Yamato VBE204

  • Page 1 VBE204/214/306/316/408/418/600/610 Instruction Manual – December 2019 –  Thank you for purchasing the Vented Balance Enclosure, VBE204/214/306/316/408/418/600/610, of Yamato Scientific.  To use this unit properly, read this "Instruction Manual" thoroughly before using the unit. Keep this instruction manual around this unit for reference at any time.
  • Page 2: Table Of Contents

    Contents ♦ General Information ……………………………………..……………………….. 2 ♦ Set Up Procedures ………………………………………………………………... 4 ♦ Exhaust Connection / Filter Change Out ……………………………………... 5 ♦ General Maintenance/ Cleaning and Sterilizing .…………………………….. 6 ♦ Models & Specifications …………………………………………………………. 7 ♦ Replacement Parts ……………………….……………………………………….. 8 ♦...
  • Page 3: General Information

    General Information The Vented Balance Enclosures have been designed and engineered to provide operator safety from airborne particulates during small scale laboratory weighing. The unit features a 25.4mm thick chemical resistant phenolic resin base, top-mounted HEPA filter fan assembly, horizontal and vertical airfoils, removable rear baffle, cable-pass thru ports, airflow alarm, and optional Bag Port chute for waste.
  • Page 4 General Information...
  • Page 5: Set Up Procedures

    Set-up Procedures INITIAL SET-UP 1. Remove packaging from enclosure. Remove power cord, power supply for air flow alarm and supplementary air flow alarm information from inside the enclosure. Position the enclosure in your desired location. 2. Plug main power cord (IEC 60320 -C13) into the input filter module on the back of the HEPA filter fan assembly.
  • Page 6: Exhaust Connection / Filter Change Out

    Exhaust Connection CANOPY CONNECTION • Air gap between house exhaust and VBE exhaust • Exhaust flow is greater than VBE exhaust flow • Previously referred to as a thimble connection. Filter Change-out 1. It is recommended that proper personal protection equipment (PPE) is utilized during filter change-out.
  • Page 7: General Maintenance/ Cleaning And Sterilizing

    General Maintenance Schedule Weekly: General cleaning (see cleaning procedures). Annually: Re-calibrate air flow sensor Cleaning and Sanitizing Most components consist of cast acrylic, thermoplastics and 304 stainless steel. Like any piece of fine laboratory equipment, care should be taken to avoid dropping, mishandling, and misapplication.
  • Page 8: Models & Specifications

    Models & Specifications Model # Inside dimensions Outside dimensions VBE204 24.00 w x 23.00 d x 21.50” h 26.00 w x 26.75 d x 41.00” h VBE214 609 w x 584 d x 546mm h 660 w x 679 d x 1041mm h VBE306 35.25 w x 23.00 d x 21.50”...
  • Page 9: Replacement Parts

    Replacement Parts ITEM # DESCRIPTION MS2027 Vinyl Tape, EL1131Power Cord, EU 1- 16, 2.5M long EL1622 Speed Control, 230 VAC SPL0026 Speed Control, 110 VAC EL1681 Impeller, 230VAC, 530CFM EL1682 Impeller, 110VAC, 530 CFM EL1703 Capacitor, 10 uF, for 110 VAC impeller ELl683 Capacitor, 2uF, for 230 VAC impeller 900-BE/24/HEPA...
  • Page 10: After Service And Warranty

    Yamato reserves the right to inspect the product under claim before having an obligation to repair or replace the defective unit covered by this warranty. All costs of shipping to Yamato for inspection shall be borne solely by the purchaser. Products repaired or replaced under the terms of the warranty may be refurbished or new product will be provided at the discretion of Yamato.
  • Page 11 Responsibility Please follow instructions in this document when using this unit. Yamato Scientific has no responsibility for accidents or breakdown of device due to failure to comply. Never conduct what this document forbids as unexpected accidents or breakdown may result.

This manual is also suitable for: