2.1. Set-up overview The Precise SHP is designed for easy set-up in your office, without the need for service personnel. Please read these instructions first so that you have a clear understanding of how and where to install this unit. If you are unsure of any step, please call your authorized distributor for either telephone or in-office support.
2.3. Facility Requirements 2.4 Installation and Setup In order to ensure the safe use of the Precise SHP in your facility, Consistent with established licensing and use agreements, please check to make sure that the proposed location has the it will be necessary for you to perform some final set up...
If that occurs you will need to redo these steps. 1. Look for the Precise SHP App. It will have a title of “SHPLaser”. If you don’t see it, drag your finger along the screen from right to left to show additional 2.4.1.3 Creating an App Store Account...
2.5.1. Mounting Base Module to Delivery Unit in the mounting brackets. Screw in the mounting screws and If desired, the Base Module of the Precise SHP can be mounted tighten securely. to or within the operatory furnishings, or to carts or mobile 6.
When you receive your Precise 3. If desired, the mounting bracket (sold separately) may SHP kit, the iPod touch included with that kit needs to be also serve as a handle for the Base Module to facilitate synchronized with the Base.
Further, status. The working beam cannot be activated by it may be necessary to power off either the Precise SHP or the means of the iPod touch. If any of the universal other equipment in order for the other device to work properly.
3.1. Safety Considerations stem. (Figure 3.2a) The safe use of the Precise SHP is the responsibility of the 3. Remove the protective cap from the shaft of the disposable entire operatory team including the doctor, assistants, and tip, and insert the tip shaft into the end of the handpiece.
Press the power button on the Control Pod. Touch the screen performance settings and allows you to on the Control Pod and select the Precise SHP App. The activate the Precise SHP handpiece. screen will now display the Introduction Screen.
For example, version of this manual, along with a quick if the duty cycle of the Precise SHP is set for 50, you will setup guide and other helpful documents have 1 energy pulse lasting 50 milliseconds with 1 period of and video demonstrations.
The following table indicates the list of procedural Presets loaded in the Precise SHP, with the associated settings that will auto-populate into the Operate screen when the procedure is selected. For all Presets, the aiming beam is set at an intensity level of 4.
Standby Status. device. Registration Screen – Also, from the Manuals Group you can register your Precise SHP. Registration aids in identifying your unit and facility if future communication occurs with CAO Group. Registration will also assist you in receiving useful information about the Precise SHP and CAO Group products in the future.
Lock key. Alternatively, press the ‘Home’ button of the look away from the treatment to confirm the laser status. iPod touch to close the Precise SHP App. Press and hold the power button on the Base Module for 3 seconds to turn While the working beam is active, move the tip over the the Base Module off.
Apply the whitening composition according to the manufacturer’s instructions for the composition. Ensure the appropriate settings are applied to the Precise SHP as indicated in Section 3.2.8 for the “Tooth Whitening” application, or select “Tooth Whitening” from the presets menu.
residual foreign substances on the optical fiber cloth moistened with a water or alcohol-based disinfectant termination and will cause premature burning to gently wipe the tip fiber and housing. DO NOT remove of the optical fiber. the clear cap and wipe the tip stem. DO NOT attempt to autoclave the tips.
If the care and cleaning of the device is followed as described, the Precise SHP will provide years of consistent performance. No other maintenance activities should 3. Use a Phillips head screwdriver to remove the four screws be needed or undertaken.
Health Care Facilities. • CAN/CSA-Z386-08 Laser safety in health care facilities. The Precise SHP has been tested and designed to the latest • TR IEC 60825-8 Guidelines for the Safe Use of safety standards applicable to medical lasers in the U.S., Medical Laser Equipment.
810 nm wavelength. We strongly recommend exposure levels exceed the MPE (which is the highest level of you use the glasses provided with the Precise SHP system or laser radiation to which a person may be exposed without obtained from CAO Group or one of its authorized dealers.
5.3. Training • A risk of fire and/or explosion exists when the Precise SHP is used in the presence of flammable materials, solutions The Precise SHP is intended for use by a licensed dental or gases, or in an oxygen enriched environment. Ensure professional.
FOR LOSS OF REVENUES INCURRED BY THE PURCHASER/USER DUE TO protection: Protected THE PRODUCT REQUIRING SERVICE. IN ORDER TO RECEIVE WARRANTY against direct water Contact information SERVICE, PRECISE LASERS MUST BE SHIPPED TO CAO GROUP, INC. IN ITS Date the device was spray of the manufacturer manufactured ORIGINAL TRANSPORTATION BOX.
Display with the Base (see Section 2.6). Check the upper right of the Display and ensure the Bluetooth icon is visible and is colored blue when the Precise App is started. If the icon is gray or missing completely, consult the ®...
Page 26
Try and enter it again. If it fails repeatedly, contact Customer Service to receive an override passcode and then reset the passcode. The laser suddenly stopped working and Precise SHP app is no The Display is receiving a communication, or another longer on the screen.
CAO Group may result in increased emissions or decreased immunity of the Precise SHP. The Precise SHP should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the Precise SHP should be observed to verify normal operation in the configuration in which it will be used.
Page 28
Guidance and manufacturer’s declaration – electromagnetic emissions The Precise SHP is intended for use in the electromagnetic environment specified below. The customer or the user of the Precise SHP should assure that it is used in such an environment. Immunity test...
Page 29
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Precise SHP is used exceeds the applicable RF compliance level above, the Precise SHP should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Precise SHP.
Page 30
RF communications equipment and the Precise SHP The Precise SHP is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Precise SHP can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Precise SHP as recommended below, according to the maximum output power of the communications equipment.
10. Glossary of Laser Terminology Activate - The action that prepares the laser to emit energy. Active Medium - The core material of a laser that is responsible for producing a source of electromagnetic energy when activated by a power supply. They can be a gas, liquid dye, semi-conductor chip or a man- made rod of Yttrium, Aluminum Garnet Scandium or Gallium, or some combination of those elements.
11. References 1. Leo Meserindino, D.D.S. and Robert Pick, D.D.S. Lasers in Dentistry. Quintessence Books, 1995. 2. Manni , Jeffrey G. Dental Applications of Advanced Lasers. Burlington, MA: JGM Associates. Contact at 781-272-6692 3. The Institute for Advanced Dental Technologies, Southfield, MI: “Lasers Dentistry: Clinical Training Seminars”. 1966.