Do you have a question about the ARKON and is the answer not in the manual?
Questions and answers
Subscribe to Our Youtube Channel
Summary of Contents for TIM ARKON
Page 1
Instructions for Use ARKON Data Bridge Philips Option SW 1.00.00...
Page 3
Page intentionally left blank. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Page 4
Owner of the brand IntelliVue Philips IntelliBridge Philips MIRUS Technologie Institut Medizin GmbH (TIM) ® ® Incides Ecolab ® ® Sani-Cloth Ecolab ® Mikrozid Schülke & Mayr GmbH Cleanisept® Dr. Schumacher GmbH ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Settings on the MIRUS Controller ................19 MIRUS Controller alarms ....................19 Faults ......................22 Cleaning ...................... 23 Service ......................25 Disposal ....................... 26 Sending the device ..................27 Spare parts list .................... 28 ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Page 6
0 Table of contents Technical data .................... 29 ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
1 Introduction 1 Introduction Intended use ARKON Data Bridge is a device for transferring data signals from a medical device, e.g. the TIM MIRUS Controller, to a Philips IntelliVue patient monitor with an internal or modular IntelliBridge EC10 interface. During operation, measurements and alarms are transferred unencrypted and unmodified, and the following functions are supported: ...
Data may not be transferred correctly or data transfer may fail. When in use, the device should be placed on a stable, flat surface and secured against falling down. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Proper functioning may be affected, possibly resulting in damage to property. Always follow precisely the instructions for all connected devices or device combinations before operating the Data Bridge for the first time. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Moreover, the proper functioning of the Data Bridge may be affected, possibly resulting in personal injury or damage to property. Only use the accessories specified by the manufacturer. Other accessories may only be used if they do not affect the electromagnetic compatibility. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
(e.g. Intensive Care), including the cleaning interval and cleaning method. Clean the Data Bridge as descripted in chapter 7. Clean the Data Bridge before maintenance and before it is returned for repair purposes. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Denotes a hazard with a low degree of risk. Failure to observe this warning can lead to minor or moderate CAUTION injuries. Denotes a harmful situation. IMPORTANT Failure to observe this warning can lead to damage to property. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
ARKON Data Bridge to the patient monitor. Mains adapter incl. power cable (3) with USB 2.0 connector (3.a) and micro USB connector (3.b) for connecting to the mains power supply. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
USB 1: USB interface for connecting the MIRUS Controller USB 2: USB interface for connecting the IntelliBridge EC5 module using the connecting cable (2) Input: Micro USB port for the mains adapter power cable (5V/2.4 A) ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Regulation 2014/30/EU Serial number Article number USB interface Packaging can be recycled Air pressure, limit (storage, transport) Relative humidity, limit (storage, transport) Temperature, limit (storage, transport) Keep dry Fragile, handle with care ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Philips power cable in the required length In order to use ARKON Data Bridge with a Philips information centre, the ARKON Data Bridge is connected to the information centre by the IntelliBridge EC5 ID module via the IntelliBridge EC40/80 hub instead of via the IntelliBridge EC10 interface. The procedure is the same as when connecting to a patient monitor.
Page 17
E. Connect the power cable to the corresponding input on the internal or modular IntelliBridge EC10 interface. F. Plug the USB connector (3.a) of the power cable (3) into the mains adapter. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Page 18
(duration: approx. 5 minutes). ► Once the data communication is established, the yellow data communication LED on the ARKON Data Bridge is permanently illuminated. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
5 Data signals from the MIRUS Controller 5 Data signals from the MIRUS Controller The following tables list the MIRUS Controller output signals provided by the ARKON Data Bridge that are visible on the patient monitor. The signal ranges and validity conform to the specifications of the MIRUS Controller.
Page 20
***UPS BATTERY LOW UPS battery has dropped below 25% of its capacity. (2 minutes remaining) [21] ***DEVICE INCLINED Device is tilted. [24] ***VA RES. EMPTY Level is less than 45 ml. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Page 21
MIRUS notification on the patient Controller monitor [25] ***DEVICE FAILURE A technical problem was [26] identified. [27] [29] [30] [31] [39] ***MR IFACE DISC. Reflector connector was disconnected. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
No data is displayed on Faults or errors in the Refer to the instructions for the patient the patient monitor. patient monitor. monitor. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Mikrozid® sensitive liquid Schülke & Mayr GmbH Mikrozid® sensitive wipes Cleanisept® Maxi wipes Dr. Schumacher ® Peracetic acid Mikrozid PAA wipes Schülke Other surface disinfectants are used at your own responsibility. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Page 24
B. Remove all other connecting cables. C. Carry out wipe disinfection. D. Remove any remaining disinfectant after the exposure time. E. Let the device dry. F. Check the device for any visible damage. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
The device does not require any maintenance. Repair To repair a defective device, and in the case of errors that you cannot rectify yourself, contact a manufacturer-authorized service organisation or the manufacturer. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
This device falls within the scope of the EU Directive 2012/19/EU (WEEE) on waste electrical and electronic equipment. To dispose of the device correctly, contact an authorised retailer or the manufacturer (TIM) to arrange for its return. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
WARNING Risk of infection The device may be contaminated with pathogens. Before sending clean the device as described in chapter 7. ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
ARKON Basic Unit AR-PH-003 ARKON Unit Feet AR-SC-12641 The instructions for use in digital form are provided via the manufacturer’s website. Comment: The printed documents can be requested from the manufacturer. ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Page 29
Philips IntelliBridge EC40/80 hub: PIIC iX Information Center (SW version B.0) PIC Information Center (SW version C.0 or higher) Compatible drivers OpenInterface IB-ED101-A.6 or higher ARKON Data Bridge SW 1.00.n, Instructions for Use (en-GB)
Page 30
Notes ARKON Data Bridge, SW 1.00.n, Instructions for Use (en-GB)
Need help?
Do you have a question about the ARKON and is the answer not in the manual?
Questions and answers