4.Product operation is carried out according to the instruction manual. Disclaimer Biobase shall not be liable for any equipment failure or damage, or for any direct or indirect damage that may occur during the use of the equipment. 1.Malfunction or damage due to violation of the instructions, precautions, and intended use of this manual.
1. Installation and Operating Conditions 1.1. This equipment should be placed in a room with good ventilation, shading, and cool conditions. The equipment and walls must be more than 10 cm away. Do not place flammable or explosive materials around it. 1.2.
5. Instructions of Operation The upper row of digital tubes displays the measured temperature, and the lower row displays the set temperature Press the SET key shortly, the temperature setting state will be entered The parameters can be corrected by the▼, ▲ keys, press the SET key to enter the time setting state Adjust time by After the▼, ▲...
Page 8
It need to re-adjust the temperature if the temperature fluctuations is more than 5 degrees, or can’t meet the requirements of the temperature. For example: After setting the required temperature of 100 degrees, press the button 5 seconds, AT indicator is on, then enter the auto tuning state.
Page 9
SCK,SCK= Standard value / Measure value. 3) P for setting proportional band. When take place regular oscillation of controlled temperature, the proportional band should be increased and if controlled temperature drifts irregularly we ought to decrease the proportional band. 4) I for setting the integral time. When the system takes place regular oscillation ought to I for setting proportional band.
6. Wiring Diagram 7. Trouble Shooting Common failures and Solutions: Phenomena Causation Treatment Method No power supply a. poor plug contact or line broke a. Connect the plug and line. b. Fuse protector is broken. b. Change the fuse protector. No temperature a.
8.2. Biobase would not be liable for any repair of damage caused by improper operation. 8.3. If the warranty has been expired, Biobase would still responsible for repair with relative charges. 8.4. Biobase would provide necessary technical data for maintenance companies or personnel trained...