Page 1
HIGH VOLTAGE POWER SUPPLIES PS-120 PS-120C PS2-60 PS2-60C Technical Description Rev. 2202 2022 Lithuania...
Page 3
CHAPTER 4 SAFETY ..........................6 CHAPTER 5 QUICK START GUIDE ......................7 5.1. I PS-120, PS2-60 ......................7 NSTALLATION OF 5.1.1. Ground the HV Power Supply ....................7 5.1.2. Connect DC Power Supply ......................7 5.1.3. Set HV Voltage ........................7 5.1.4.
IST OF IGURES 1. O ..............4 IGURE UTLINE DRAWING AND DIMENSIONS OF THE POWER SUPPLY 2. T ............4 IGURE OP VIEW OF THE POWER SUPPLY WITH POTENTIOMETER CONTROL 3. T ) ........5 IGURE OP VIEW OF THE POWER SUPPLY WITH INTERFACE CONTROL MODELS...
Page 5
Any evidence of improper use or unauthorized repair attempts voids the warranty. For service/warranty requests, please contact: EKSMA OPTICS, UAB Phone: +370 5 272 99 00 c/o EKSMA Optics, UAB Fax.: +370 5 272 92 99 Dvarcioniu st. 2B E-mail: info@eksmaoptics.com...
The table below lists standard options. Custom voltage modifications may be delivered on request. Table 1. Models Catalog number HV output Maximal HV current, Output voltage tuning range, kV control PS-120-1.8 +0.8 … +1.8 By potentiometer PS-120C-1.8 +0.7 … +1.8 By CAN interface PS-120-2.6 +1.6 … +2.6 By potentiometer PS-120C-2.6...
Table 3. Technical specifications Parameter Value(s) PS-120(C) PS2-60(C) Output voltage polarity Positive Bipolar Maximum HV output power at maximal output 2×60 voltage, W Output voltage ripple, % <0.5 Internal potentiometer Output voltage control CAN interface (-C models) DC supply requirements 23.5…26 V, 6500 mA...
Chapter 3 D E VICE AYOU T Figure 1. Outline drawing and dimensions of the HV power supply Figure 2. Top view of the HV power supply with potentiometer control www.eksmaoptics.com...
4 Pin HV output off (HV output is off when pin connected to GND) 5 HV tuning potentiometer 6 Pin for HV measurement. 1:1000 output. 7 CAN interface connector 8 CAN terminator 9.1 +HV output 9.2 GND for HV output (two pins of connector) -HV output for PS2-60(C) +HV output for PS-120(C) www.eksmaoptics.com...
Page 10
Chapter 4 S AFE T Y Equipment is designed to be safe under normal environmental conditions according to 1.4.1. 61010-1@IEC:2010 (Safety requirements for electrical equipment, control and laboratory use): 1. indoor use; 2. altitude up to 2000 m; 3. temperature 5˚C to 35˚C; 4.
Page 11
Chapter 5 Q U ICK T ART U IDE Provide proper ground connection to the HV module. In order to prevent electrical breakdown, ensure that the distance between the bottom of HV module and the surface below is more than 6 mm.
CAN-USB converter is recommended for operation. It can be ordered from Eksma Optics. It comes together with software and a set of connection cables. Figure 4. CAN-USB converter Provide proper ground connection to the HV module. In order to prevent electrical breakdown, ensure that the distance between the bottom of HV module and the surface below is more than 6 mm.
Figure 5. Control panel application Use CANBrowser software to read-out actual status of HV power supply and save CSV file while need for technical service or consultation of manufacturer. CSV file includes list of CAN control registers and actual settings. Switch the DC power on.
Need help?
Do you have a question about the PS-120 and is the answer not in the manual?
Questions and answers