Document 900034, Rev B 1. GENERAL INFORMATION 2. APPROVALS The DIGITEL MPCq is shown to meet the intent of Directive 89/336/ The DIGITEL™ MPCq is an ion pump power supply controller. EEC for Electromagnetic Compatibility and Low Voltage Directive It is designed to operate up to two independent ion pumps along 73/23/EEC for product Safety.
Document 900034, Rev B 4. UNPACKING THE CONTROLLER 6. INSTALLATION The DIGITEL MPCq is a full 19 inch rack controller. No additoinal 4.1 Inspect For Any Obvious Damage hardware needed to install in a standard 19 inch wide device rack.
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 7. CONTROLLER CONFIGURATION 8. HV CONTROL / ION PUMP OPERATION Following configuration options are specified at order time: 8.1 Prior to using an ion pump Confirm following: 7.1 Input Mains AC Voltage (110Vac vs. 220Vac) - Rough pump down to 1x10 Torr or less (the lower the better).
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 9. PRESSURE MEASUREMENT Pressure value is determined using following equation Where: I - Current in amps. V - Voltage in volts. U - Pressure units conversion factor (1 for Torr, 1.33 for mBar and 133 for Pascal).
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 11. TOUCHSCREEN USER INTERFACE 11.2 Main Menu (Page 1) 11.1 Home Screen Touch anywhere on the screen to invoke button controls. Figure 5. Main Menu (Page 1)) Page 1 of main menu is accessed by pressing the “PAGE ½” button.
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 11.4 Supply Setup 11.6 System Setup Figure 7. Supply Setup Screen Figure 9. System Setup Screen Available setup: Available setup: Pump name - Configures pump name. Max of 15 characters.
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 11.7 Analog Output 11.8 Digital Input Setup Figure 11. Digital Input Setup Screen Figure 10. Analog Output Screen Total of four digital inputs on connector J104. Total of four analog outputs on connector J104:...
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 11.9 Digital Output 11.10 Relay Setup Figure 13. Relay Setup Screen Figure 12. Digital Output Setup Screen Total of four relays on connector J104: Total of four digital outputs on connector J104:...
Pump size set to 0L/s. HV cannot run. Supply not calibrated. Current metering might not be accurate. Supply over-temperature detected. HV cannot run. NOTE: If error code is not listed in the table please contact Gamma Vacuum for more information. Page 14 of 28...
Use high current cable to connect to Titatium Sublimation Pump Standard Sub-D 9-pin female connector. Connection to Profibus (TSP). interface. For Profibus information reference Profibus user manual, document 900033, available on Gamma Vacuum website under download section. Page 15 of 28...
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 13.7 Connector J104, Misc I/O 13.8 Connector J112, SER Standard Sub-D 9-pin female connector. Connection to serial Standard Sub-D 37-pin female connector. interface. Supported serial standars are RS-232, RS-485 full duplex Table 2.
2 ASCII hex characters Range 00 through FF See “Table 9. Gamma Protocol Commands” on page 20 for list of available command codes. The command code must be two hex digits, even if the first is a zero. 4. DATA field(s)
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 14.2 Gamma Protocol Response Packet Structure Over Serial Connection The response packet is made up of at least five fields and contains information to let the controlling computer know the command requested was either recognized and accepted (STATUS = “OK”), or an error condition occurred (STATUS = “ER”).
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 14.5 Gamma Protocol Commands Table 9. Gamma Protocol Commands 9.1 Hex 9.2 Description 9.3 Data 9.4 Response 9.5 Data/Response Command Field Description SYS GET MODEL DIGITEL MPCQ Get controller model string.
Page 21
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 9.1 Hex 9.2 Description 9.3 Data 9.4 Response 9.5 Data/Response Command Field Description HV SET SUPPLY AUTO RESTART S, X Where: S is supply (1-2) Sets supply auto restart option.
Page 22
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 9.1 Hex 9.2 Description 9.3 Data 9.4 Response 9.5 Data/Response Command Field Description TSP TURN ON Where: R is TSP (1-2) Turns on TSP programmed mode. TSP GET ACTIVE FILAMENT Where: Reads active TSP filament.
Page 23
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 9.1 Hex 9.2 Description 9.3 Data 9.4 Response 9.5 Data/Response Command Field Description TSP GET TSP STATUS R, X Where: Reads TSP status. R is TSP (1-2) X is option. Currently option 0 is only supported.
2 ASCII hex characters Range 00 through FF See “Table 9. Gamma Protocol Commands” on page 20 for list of available command codes. The command code must be two hex digits, even if the first digit is a zero. 4. DATA...
Gamma Vacuum DIGITEL MPCq Users Manual Document 900034, Rev B 17. TECHNICAL SPECIFICATIONS Dimensions (max) ......472 x 432 x 131mm, length x width x height (fits standard 19 inch wide device rack) Shipping Weight (max) ....32 (66) kg lbs (fully configured with two supplies) Operating Temperature ....0 to 40C Free airflow around the unit is required.