Micro Gear Pump Drive Operating Manual Impor tant Infor mation: Please read operating manual carefully before operation. War ning: T o prevent damage to pump, do not run the pump dry. If the power line or the plug are worn or damaged please pull out the plug (Hold the plug not the power line when pulling out).
Table of Contents Instruction Basic Operation Running Interface Menu Navigation Diagram Pump Head Operation P u m p He a d I n s t a l l a t i o n P u m p He a d F l o w R a t e s P u m p He a d S e t t i n g Flow Mode Operation E n t e r F l o w M o d e...
Micro Gear Pump Drive Operating Manual Instruction Acceptable pump heads for WT3000-1FB are MG204, MG209 and MG213. The pump delivers flow rates from 83 to 2750 mL/min and dispensing volume is from 0.1 mL to 999 L. Combining 128 32 LCD display with membrane keypad and rotary coded switch makes the operation easy and prompt.
Micro Gear Pump Drive Operating Manual Basic Operation LCD Display Knob Exit Key Prime Key Stop Key Run Key Power Switch Pump Head Connecting Board Run Key In stop state, press Run Key , the pump starts to run. Stop Key In non-prime state, press Stop Key , the pump stops.
Micro Gear Pump Drive Operating Manual Knob Function 1: Turn the Knob to adjust the flow rates in flow rates display state. Function 2: Turn the Knob to select the menu and set the parameters, press Knob for confirmation in menus selection state. Power Switch Turn on or off the pump.
Micro Gear Pump Drive Operating Manual 【 】 【 】 d. Running status: means the pump runs; means the pump 【 】 stops; means the pump is in prime state. e. Control mode: INT means the pump is controlled internally; V means the pump is controlled by external analog voltage signal;...
Micro Gear Pump Drive Operating Manual Pump Head Operation Pump Head Installation: The pump head is mounted on drive before the pump leaves the factory. Follow below procedure for changing another pump head. 1.Loose the three M3 screws which connect the pump head and the pump head bracket.
Micro Gear Pump Drive Operating Manual Flow Mode Operation Enter Flow Mode Press and turn the Knob to highlight control mode line. Press Knob to enter next interface, turn Knob to select control mode, press Knob for confirmation. Press Exit Key not to store the selection and return to previous menu.
Micro Gear Pump Drive Operating Manual Flow Mode Operation Procedure Dispensing Mode Operation Enter Dispensing Mode Press and turn the Knob to highlight Control Mode line. Press Knob to enter next interface, turn Knob to select Dispensing , press Knob for confirmation.
Micro Gear Pump Drive Operating Manual Copy No. the total filling number in dispensing mode. Its range is from 0 to 9999. Press and turn Knob to highlight copy number line. Press Knob to enter next interface. Turn Knob to select copy number, press Knob confirmation.
Micro Gear Pump Drive Operating Manual Dispensing Calibration The parameters can not be changed under dispensing volume calibration operation. The defaulted parameters are the parameters set under Dispensing Setting. Pump stops under dispense mode, press and turn Knob to highlight calibration line.
Micro Gear Pump Drive Operating Manual Exter nal Control R 4 8 5 Communication Control External Control Mode 4 2 0 m A (DB15 Plug) Analog Control 1 0 V 1. 0 1 0 kH z 1.Enable the external control function. 2.These five Standard Analog Control Module need to be order separately.
Micro Gear Pump Drive Operating Manual Terminal Definition of Standard External Control Module Analog input in 4 20 mA, 0.5 5 V, 1.0 10 V external control modules. Control the speed of the pump. It is open in 1-10KHz. External control module.
When control computer controls many pumps through RS485 interface, it must identify each pump's I.D. This pump I.D. should be unique. It's the identification of this pump. Maximum 30 pieces WT3000-1FB can be controlled through RS485 at the same time.
Micro Gear Pump Drive Operating Manual Exter nal Control Output Pump is equipped with output port in order to monitor the status of pump. Please see below “External Control Output Port pins Diagram: DB9”. Output signal adopts optoelectronic isolation circuit. A pull-up resistor and power supply are needed when using.
Page 17
Micro Gear Pump Drive Operating Manual Specifications Flow Rate 85.7~2571.4 (Ml/min) 0.1mL~999L Dispensing Volume 0~9999, ‘0’ mean infinite cycle. Copy No. Pause Time 1 second~999 hours, resolution is 0.1second. 30 second~30 minutes, resolution is 1 second. Flow Rate Calibration 4~20 mA, 0.5~5 V, 1.0~10 V modules control Input Function flow rates.
Page 18
Baoding Longer Precision Pump Co.,Ltd. Building A, Chuangye Center Baoding National High-Tech Industrial Development Zone Baoding, Hebei, China 071051 TEL: 86-312-3110087 FAX: 86-312-3168553 E-mail: longer@longerpump.com Http //www.longerpump.com...