ADF-B27 / ADF-B27A*
Blood Glucose Meter
User Manual
*ADF-B27A default unit mmol/L

Advertisement

loading
Need help?

Need help?

Do you have a question about the ADF-B27 and is the answer not in the manual?

Questions and answers

Sondra carr
May 7, 2025

How to set date and time on blood glucose meter

User image 681b70d833ec0
1 comments:
Mr. Anderson
May 19, 2025

To set the date and time on the Andesfit Health blood glucose meter ADF-B27:

1. Press and hold the “S” button for 3 seconds until the last two digits of the year flash.
2. Press the “M” button to adjust the year, then press “S” to confirm.
3. When the month flashes, press “M” to set the correct month, then press “S”.
4. When the day flashes, press “M” to set the correct day, then press “S”.
5. When AM/PM flashes, press “M” to choose the correct one, then press “S”.
6. When the hour flashes, press “M” to set the hour, then press “S”.
7. When the minute flashes, press “M” to set the minute, then press “S”.

This answer is automatically generated

Summary of Contents for Andesfit Health ADF-B27

  • Page 1 ADF-B27 / ADF-B27A* Blood Glucose Meter User Manual *ADF-B27A default unit mmol/L...
  • Page 2 Blood Glucose Monitoring System Owner: Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:...
  • Page 3: Table Of Contents

    DANGER - Misuse of this electrical device can cause electrocution, burns, fire and other HAZARDS. BEFORE YOU BEGIN……………………………………………5 Basic safety precautions should always be taken, including ABOUT YOUR SYSTEM…………………………………………6 all those listed below. INTRODUCTION TO THE SYSTEM……………………..6 Close supervision is necessary when equipment is used by, ABOUT ALTERNATIVE SITE TESTING (AST)………………..7 on, or near children, handicapped persons or invalids.
  • Page 4: Before You Begin

    BEFORE YOU BEGIN Installing the battery ■ The system is intended for use outside the body (in vitro diagnostic use). It is for single-patient use. It should be used only for testing glucose (sugar) and only with fresh capillary whole blood samples taken from the finger and the alternative sites including the palm, the forearm.
  • Page 5: About Alternative Site Testing (Ast)

    AST in this system can be used only during steady-state blood glucose conditions described in the section of “About AST”. Fingertips feel pain more readily because they are full of nerve endings (receptors). At other body sites, since nerve endings are Test principle not so condensed, you will not feel as much pain as at the fingertip.
  • Page 6 ■ A Meter ■ Portable Bag Owner’s Manual ■ ■ Lancing Device ■ Sterile Lancet (Optional) ■ ■ Control Solution (Optional) LOW BATTERY SYMBOL BLOOD DROP SYMBOL ● The Lancing Device can be used for several times however Appears when the battery Flashes when it is ready the lancet should be changed after each use.
  • Page 7 Blood Glucose Meter Test Strip Each strip can be used only once. The test strip consists of the following parts: Absorbent Hole Apply a drop of blood here. The blood will be drawn in automatically. Confirmation window This is where you confirm if enough blood has been applied to the absorbent hole of the strip.
  • Page 8 1.1 Setting the Year Press and Hold “S” button for 3 seconds, until the last two digits of the year will flash on the screen (for example, “14” equals 2014).Press “M” button until the correct year appears. Press “S”. 1.2 Setting the month With the month flashing, press “M”...
  • Page 9: The Two Measuring Modes

    THE TWO MEASURING MODES TESTING WITH BLOOD SAMPLE The meter provides you with two modes for measuring, General and CTL. The CTL mode should be used when a control solution o reduce the chances of infection: is being tested. See the table below. ●...
  • Page 10 On the lancets and the lancing device. If your lancing device differs from the one shown here, please refer to the manufacturer’s manual to ensure proper usage. Stimulating blood perfusion by rubbing the puncture site prior to To reduce the chance of infection: blood extraction has a significant influence on the glucose value ●...
  • Page 11 Fingertip testing Hold the lancing device firmly against the side of your finger. Press the release button. You will hear a click, indicating that the puncture is complete. Step 4 When the blood drop symbol flashes on the screen, you may select the appropriate measure- Choose a different spot each time you test.
  • Page 12 The results obtained from this meter are plasma-calibrated. This The meter provides you with whole blood equivalent results. The helps you and your physician or other qualified healthcare result you obtain from your meter may differ somewhat from your provider to compare your meter results with laboratory tests. laboratory result due to normal variation.
  • Page 13: Control Solution Test

    apply control solution onto a strip. CONTROL SOLUTION TEST Material needed: The result of the control solutions test appears on the display. ● Glucose Meter The result will not be stored under CTL mode. ● Test strips The result of the control solutions test should be within the range ●...
  • Page 14: Care And Storage For Your System

    CARE AND STORAGE FOR YOUR SYSTEM 2.Press “M” again, the most recent test results with date and time ◆ Handle the meter with care. Dropping or throwing the meter will display. Press “M” once more and the next most recent test results will appear.
  • Page 15 ※ The system can be used up to an altitude of 10744 feet. ※ The following substances at levels greater than normal or therapeutic levels may cause significant interference (affect 1. Model: ADF-B27 / ADF-B27A* the result by greater than 10%) resulting in an inaccurate 2. Machine size: 90mm (L)×54mm (W)×13mm(H) 3.
  • Page 16: Display Message

    ※ Patients undergoing oxygen therapy may yield falsely lower result: ascorbic acid, uric acid, hemoglobin,acetaminophen, results. Dopamine, L-dopa, L-dopa and Tolbutamide etc. These sub- ※ Not for use for patients in a hyperglycemic-hyperosmolar stances do not affect test results in normal concentration but state, with or without ketosis.
  • Page 17 4. Test Strip has not 4. Reinsert the Test MESSAGE WHAT IT MEANS ACTION been fully inserted Strip into the Meter. Use a used test strip. Please use a new strip into the Meter. and re-test. Test results 1. Not enough sample 1.
  • Page 18 For glucose concentrations<5.55 mmol/L (<100 mg/dL) Performance Characteristics Measurement Precision Puncture site Fingertip Palm Forearm Repeatability ±15mg/dL (±0.83mmol/L) 96.3% 96.3% 96.3% Mean (mg/dL) 40.3 82.0 For glucose concentrations≥5.55mmol/L (≥100 mg/dL) SD (mg/dL) Puncture site Fingertip Palm Forearm CV (%) Within±15% 99.3% 98.6% 98.6%...

This manual is also suitable for:

Adf-b27a