Page 1
MILLENNIUM II Gas Sensor Oxygen (O2) Sensor User Manual Model: ST340X-25-ASSY Part Number: MAN-0093 Rev 0 June 2008...
Page 2
Safety Monitoring Inc. Contact Net Safety Monitoring Inc. or an authorized representative for details. We welcome your input at Net Safety Monitoring Inc. If you have any comments please contact us at the telephone number or address below or visit our web-site, www.net-safety.com, and complete our on-line customer survey.
5.2.3 Calibrating with the Millennium II Transmitter ....................16 5.3 Sensor Replacement Procedure ........................... 17 5.4 Troubleshoot ................................18 5.5 Spare Parts / Accessories ............................18 MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Page 4
Appendix ....................................20 Appendix A: Electrostatic Sensitive Device (ESD) ......................20 Appendix B: Resistance Table ............................21 Appendix C: Millennium II Oxygen (O2) Sensor Specifications ..................22 MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Millennium II Basic & sensor Transmitter & sensor(AL) Transmitter & sensor(SS) Figure 2 below, shows the dimensions of the Millennium II transmitter with sensor and the Millennium II Basic transmitter with sensor. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Page 7
Figure 2: Sensor with Millennium II series transmitters dimensional drawing MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
(air) and comparing the output results. They should be accurate to the calibration gas concentration used. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Carefully remove all the components from the packaging and check them against the enclosed packing list. Inspect all components for any obvious damage such as broken or loose parts. If you find any components missing or damaged, notify the representative or Net Safety Monitoring immediately. 2.2 Mount The sensor is mounted directly to either transmitter housing or to a separation junction box housing through the ¾”...
A B Earth Ground Ground NOTE: When separating sensor from transmitter using Net Safety separation kit, refer to Multi-purpose Junction Box Manual (MAN-0081) for terminal designations. SECTION 3: Operation 3.1 Configuration Settings All configuration settings are accessed through the Millennium II series transmitter. This is done by setting dip switches on the Millennium II Basic and by selecting menu options in the Millennium Transmitter.
The Oxygen sensor only requires zeroing in ambient air for calibration, so either register 40102 or register 40104 may be used, as both commands will result in a proper calibration cycle. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Table 6: Dip Switch 3 Positions for Alarm Mode DIP Switch 3 Alarm Mode Settings Position 1 ON ON OFF OFF Position 4 ON OFF ON OFF Alarm Mode Above‐Above *Below‐Above * Below‐Above Below‐Below Note: There are two configurations for Alarm Mode (* Below-Above) MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Alarm Mode: Below- Above Alarm Point 1 Position 1 Position 2 19.5% 19.0% 18.5% 18.0% Alarm Mode: Below- Above Alarm Point 2 Position 3 Position 4 22.0% 22.5% 23.0% 23.5% MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Note: A zero / calibration can be done, by releasing the push button at the first solid green in step 5 above and applying clean air. Zeroing the sensor will also result in calibrating, since the sensor is an oxygen sensor. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
4. The existing alarm levels will be displayed, as CH1 Point 1(or CH2 Point 1) and Ch1 Point 2(or Ch2 Point 2). Highlight the level to be changed and select with menu button 3(Reed switch 3) MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
9. “Cal Complete” will be displayed when calibration is complete. Note: Air used for calibration can either be from the gas canister or surrounding air (clean air). MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
6. Install and tighten the bottom part of the sensor housing by turning in a clockwise rotation. 7. Install the locking ring by tightening the set crews with Allen Key tool. 8. Restore power to sensor. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
If the problem persists, please contact Net safety’s service department first by phone to try and resolve the issue. If the issue cannot be resolved, please follow the procedure below on how to return equipment.
5.6 How to Return Equipment A Material Return Authorization number is required in order to return equipment. Please contact Net Safety Monitoring at (403) 219-0688, before returning equipment or consult our Service Department to possibly avoid returning equipment. If you are required to return equipment, include the following information: 1.
In general, exercise accepted and proven precautions normally observed when handling electrostatic sensitive devices. A warning label is placed on the packaging, identifying product using electrostatic sensitive semiconductor devices. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
57.50 36.10 22.70 14.30 8.99 5.65 10000 102.00 63.90 40.20 25.30 15.90 9.99 6.28 Resistance shown is one way. This figure should be doubled when determining closed loop resistance. MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
NEMA 4X Separation Separation Up to 2000 feet / 600 meters Approvals 0575 II 2 G, EEx d IIB + H2, T4 DNV-2006-OSL-ATEX-0475X Approvals -40°C < Ta < +75°C MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.
Page 23
Net Safety Monitoring Inc. 2721 Hopewell Place NE, Calgary, AB Canada T1Y 7J7 1‐866‐FIREGAS (347‐3427) | ph. (403) 219‐0688 | fx. (403) 219‐0694 http://www.net‐safety.com | info@net‐safety.com PRODUCT SERVICES CONTACT INFORMATION Telephone [ 8am ‐ 5pm MDT ]: (403) 769‐6074 | (403) 717‐8219 Fax: (403) 219‐0694 Email: productservices@net‐safety.com http://www.net‐safety.com/service/product_services.html MAN-0093 Rev 0 Oxygen Sensor June 2008 Net Safety Monitoring Inc.