Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table of Contents 1 HOW TO USE THIS MANUAL ..............1-1 1.1 S ....................1-2 AFETY ESSAGES 2 GETTING STARTED.................2-1 2.1 I ......................2-1 NSTALLATION 2.2 E ..............2-1 LECTRICAL AND...
Page 4
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6 OPTIONAL HARDWARE AND SOFTWARE..........6-1 6.1 R ....................6-1 OUNT PTIONS 6.2 Z ....................6-1 ALVES 6.3 I (IZS)..................6-3 NTERNAL 6.4 A ............. 6-3 UTOCAL ETUP OF ALVES 6.5 P...
Page 5
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8 MAINTENANCE ..................8-1 8.1 M ................... 8-1 AINTENANCE CHEDULE 8.2 R ............. 8-2 EPLACING THE AMPLE ARTICULATE ILTER 8.3 R )............... 8-4 EPLACING THE ERMEATION OPTION 8.4 R IZS Z ).............
Page 6
9.5.1 Leak Check...................... 9-60 9.5.2 Pump ....................... 9-60 9.5.3 Kicker....................... 9-60 9.5.4 Z/S Valves & IZS Permeation Tube Oven ............9-62 10 TELEDYNE API M100A SPARE PARTS LIST........10-1 APPENDIX A MAINTENANCE SCHEDULE FOR M100A ......A-1 APPENDIX B ELECTRICAL SCHEMATICS ..........B-1...
Page 7
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G List of Figures 2-1: R ..............2-3 IGURE EMOVAL OF HIPPING CREWS 2-2: R ....................2-4 IGURE ANEL 2-3: R ..........2-5 IGURE ANEL NEUMATIC ECOMMENDATIONS 2-4: F ....................2-10 IGURE RONT ANEL 2-5: A ..................2-11...
Page 8
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G List of Tables 2-1: F ............2-12 ABLE INAL EST AND ALIBRATION ALUES 2-1: F )........2-13 ABLE INAL EST AND ALIBRATION ALUES ONTINUED 4-1: S ................4-6 ABLE YSTEM ODES ISPLAY...
Page 9
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8-1: P ............8-1 ABLE REVENTATIVE AINTENANCE CHEDULE 9-1: T ...................9-3 ABLE UNCTIONS 9-1: T )...............9-4 ABLE UNCTIONS ONTINUED 9-1: T )...............9-5 ABLE UNCTIONS ONTINUED 9-1: T )...............9-6 ABLE UNCTIONS...
Analyzer Instruction Manual, 02164, Rev. G 1 HOW TO USE THIS MANUAL The Model 100A has been designed with serviceability, reliability and ease of operation in mind. The M100A's microprocessor continually checks operating parameters such as temperature, flow, and critical voltages. The instruments modular design uses captive screws to facilitate repair and ease of access.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 1.1 Safety Messages Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol.
2.1 Installation CAUTION To avoid personal injury, always use two persons to lift and carry the Model 100A. Verify that there is no apparent shipping damage. If damage has occurred please advise shipper first, then Teledyne API. Before operation it is necessary to remove the shipping hold-down screws. Remove the instrument cover, then refer to Figure 2-1 for screw location.
Page 14
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G If desired, attach the analog output connections to a strip chart recorder and/or datalogger. Refer to Figure 9-3 for jumper settings for the desired analog output voltage range. Factory default setting is 0-5VDC.
Page 15
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 2-1: Removal of Shipping Screws...
Page 16
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 2-2: Rear Panel...
Page 17
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 2-3: Rear Panel Pneumatic Recommendations...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 2.3 Initial Operation 12. Turn on the instrument power. The display should immediately light, displaying the instrument type (M100A) and the CPU memory configuration. If you are unfamiliar with the M100A, we recommend that you read the overview Section 4 before proceeding.
Page 19
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G The next task is to calibrate the analyzer. There are several ways to do a calibration; they are summarized in Table 7-1. For a preliminary checkout we recommend calibration with zero air and span gas coming in through the sample port.
Page 20
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number Action Comment Input Zero gas Allow Zero gas to enter the sample port on the rear of the instrument.
Page 21
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Step 3 - Review the quality of the calibration: Calibration Quality Check Procedure Step Number Action Comment Scroll the TEST Typical SLOPE value is 1.0 +/- 0.3. If the value is not in this function menu until range, check Section 7.10 or 9.
Page 22
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 2-4: Front Panel 2-10...
Page 23
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 2-5: Assembly Layout 2-11...
Page 24
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 2-1: Final Test and Calibration Values Observed TEST Values Units Nominal Range Reference Section Value RANGE 50 - 20,000 5.3.4 STABIL 0.1 - 2 5.2.1, 9.1.1, Table 9-1, 9.2.5...
Page 25
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 2-1: Final Test and Calibration Values (Continued) Span and Cal Values Observed Parameter Units Nominal Range Reference Section Value Span Conc 20 - 20,000 Table 7-3 Slope 1.0 ± 0.3 7.10...
Page 26
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G INTENTIONALLY BLANK 2-14...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY 3.1 Specifications Ranges In 1ppb increments from 50ppb to 20,000ppb, dual ranges or autoranging Noise at Zero 0.2 ppb RMS Noise at Span <0.5% of reading (above 50 ppb) Lower Detectable Limit 0.4 ppb RMS...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 3.2 EPA Equivalency Designation Advanced Pollution Instrumentation, Inc., Model 100A Sulfur Dioxide Analyzer is designated as Reference Method Number EQSA-0495-100 as defined in 40 CFR, Part 53, when operated under the following conditions: 14.
All replacement parts and repairs are warranted for 90 days after the purchase. If a Product fails to conform to its specifications within the warranty period, Teledyne API shall correct such defect by, in Teledyne API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product.
Page 30
Analyzer Instruction Manual, 02164, Rev. G TERMS AND CONDITIONS All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
4 THE M100A SO ANALYZER 4.1 Principle of Operation The operation of the Teledyne API Model 100A Analyzer is based upon the well proven technology from the measurement of fluorescence of SO due to absorption of UV energy. Sulfur Dioxide absorbs in the 190 nm - 230 nm region free of quenching by air and relatively free of other interferences.
Page 32
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Ultraviolet light is focused through a narrow 214 nm bandpass filter into the reaction chamber. Here it excites the SO molecules, which give off their characteristic decay radiation. A second filter allows only the decay radiation to fall on the PMT.
Page 33
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 4-1: M100A Sulfur Dioxide Analyzer...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 4.2 Instrument Description 4.2.1 Sensor Module, Reaction Cell, Detector The sensor module (Figure 9-7) is where the fluorescence light is generated and detected. It is the most complicated and critical sub-assembly in the entire analyzer. It consists of the following assemblies and functions: 1.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G The M100A is a true computer based instrument. The microprocessor does most of the instrument control functions such as temperature control, valve switching. Data collection and processing are done entirely by the CPU with the final concentration values being sent to a D/A converter to produce the instrument analog output.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 4-1: System Modes Display Mode Meaning SAMPLE xx(1) Sampling normally SAMPLE xx(1) Flashing indicates adaptive filter is on ZERO CAL x(2) Doing a zero check or adjust SPAN CAL x(2)
Page 37
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 4-2: Front Panel Display 4.2.5.2 Status LED's At the right of the display there are 3 status LED's. They can be in three states, OFF, ON, and Blinking. The meanings of the LED's are given in Table 4-2.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 4.2.5.3 Power Switch The power switch has two functions. The rocker switch controls overall power to the instrument; in addition it includes a circuit breaker. If attempts to power up the M100A result in a circuit breaker trip, the switch automatically returns to the off position, and the instrument will not power up.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5 SOFTWARE FEATURES This section covers the software features of M100A which is designed as a computer controlled instrument. All major operations are controlled from the front panel display and keyboard through friendly menu.
Page 40
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 5-2: Setup Menu Tree...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.1.1 Sample Menu Table 5-1: M100A Sample Menu Structure Menu Level Reference Level 1 Level 2 Level 3 Level 4 Description Section TEST Test functions 5.2.1, Table 9-1 TST>...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.1.2 Set-Up Menu Table 5-2: M100A Setup Menu Structure Setup Menu #1 Reference Level 1 Level 2 Level 3 Level 4 Description Section CFG is primarily used for 5.3.1...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 5-3: M100A Menu Structure - Setup Menu #2 Setup Menu #2 Reference Level 1 Level 2 Level 3 Level 4 Description Section RNGE Range control menu 5.3.4 MODE Range mode select - Single, 5.3.4...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 5-4: M100A Menu Structure - Setup Menu #3 Setup Menu #3 Reference Level 1 Level 2 Level 3 Level 4 Description Section MORE Next level of the SETUP...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.2 Sample Mode 5.2.1 Test Functions NOTE In any of the following TEST functions, if XXXX is displayed, that indicates an off scale and therefore meaningless reading. To use the TEST functions to diagnose instrument faults, refer to Troubleshooting Section 9.1.
Page 46
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G PMT Voltage The PMT VOLTAGE measures the PMT signal at the output of the preamp board. The waveform of the PMT voltage can be complex, and vary up to 5000 mV when a high concentration of SO being measured.
Page 47
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G DC Power Supply (DCPS) The DCPS voltage is a composite of the 5 and ±15VDC voltages in the Power Supply Module. This is meant to be a quick indicator to show if the PSM is working correctly. The nominal value is 2500mV ±...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.2.2 CAL, CALS, CALZ, Calibration Functions The calibration and zero/span checking of the M100A analyzer is treated in detail in Section 7. Table 7-1 summarizes types of calibration. The basic function of each of these keys is described here.
Page 49
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.2.2.5 Formula Values The slope and offset terms should be checked after each calibration. The values for these terms contain valuable information about the internal health of the analyzer. The range of acceptable values and their meanings is given in Section 7.10.
The Data Channels can be programmed and customized from the front panel. A set of default Data Channels has been included in the Model 100A software. These are described Section 5.3.3.1. For more information on programming custom Data Channels, a supplementary document containing this information can be requested from Teledyne API.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G <PRM and PRM> will appear. These buttons are used to scroll through the parameters located in each record. The function of each of the default Data Channels is described below:...
Page 52
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G NOTE Only one of the following range choices can be active at any one time. There are 3 range choices: 1. Single Range 18. Auto Range 2. Dual Ranges 5.3.4.1 Single Range...
Page 53
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.3.4.3 Dual Ranges Dual Ranges allows you to select different ranges for REC and DAS analog output channels. To set up Dual Ranges press SETUP-RNGE-MODE-DUAL, then press ENTR. To set the values press SETUP-RNGE-SET.
Page 54
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 3. SELECT RANGE The range selection is the same with dilution as with normal monitoring. See Section 5.3.4 for information on range selection. You should note however, the value entered should be the actual concentration of the calibration gas entering the dilution probe.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.3.5 Password Enable There are two levels of password protection. The most restrictive level requires a password to do instrument calibration. The second level requires a password to do SETUP functions.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5.3.9 Variables Menu (VARS) This menu enables you to change the settings on certain internal variables. The VARS Table 9-5 is located in the Troubleshooting Section 9.1.4. NOTE Before changing the settings on any variables, please make sure you understand the consequences of the change.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6 OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with the M100A includes: Rack mount with slides (P/N 00280). Rack mount without slides, ears only (P/N 01470). Fluorocarbon zero/span valves (P/N 01491).
Adequate zero air can be supplied by connecting a charcoal scrubber and 5 micron particulate filter (Teledyne API P/N 000369) to the zero air inlet tubing. The zero air scrubber used in conjunction with the Zero/Span Valve option provides an inexpensive source of zero air.
The Zero/Span valves or the IZS system can be set up to operate automatically on a timed basis. The Teledyne API model 100A with IZS option offers capability to check any combination of zero and span points automatically on a timed basis, through remote RS-232 operation (see Section 6.8.5), or external contact closure (see Section 7.7).
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 6-2: Setup Automatic Zero/Span Calibration Default Description Timer Enable Enable or disable automatic calibration timer. Starting Date 01-JAN-95 MM:DD:YY Starting Time 00:00 HH:MM, 0 - 23 hours and 0 - 59 minutes...
Page 61
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Example 2: to perform zero calibration adjust once per day retarding 15 minutes everyday starting at 11:30 pm, 12/20/98. 1) MODE: ZERO 2) TIMER ENABLE: ON 3) STARTING DATE: 12/20/98...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6.5 Permeation Tube (used with IZS option) is normally a gas at room temperature and pressure, but can be liquefied at moderate pressures. The permeation tube consists of a small container of SO...
Page 63
0.4ppm @ 0.7 lpm Permeation tube certified 0.8ppm @ 0.7 lpm Permeation tube uncertified 0.8ppm @ 0.7 lpm Teledyne API recommends that you purchase replacement permeation tubes from: VICI METRONICS 2991 Corvin Drive Santa Clara, CA 95051 USA Phone 408-737-0550 Fax 408-737-0346 The M100A uses 700 cc/min of zero air over the perm tube.
Page 64
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 6-1: Permeation Tube Installation...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6.6 4-20 mA, Current Loop Output The current loop option replaces the voltage output of the instrument with an isolated 4-20 mA current output. The current outputs come out on the same terminals that were used for voltage outputs, see Figure 2-2.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6.8 RS-232 Interface The RS-232 communications protocol allows the instrument to be connected to a wide variety of computer based equipment. The interface provides two basic functions in the M100A.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 6-4: RS-232 Port Setup - Front Panel Decimal Value Description Turns on quiet mode (messages suppressed) Places analyzer in computer mode (no echo of chars) Enables Security Features (Logon, Logoff)
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 940331 is the default password. The password can be changed to any number from 0 to 999999 by the variable RS232_PASS. To change the password enter the command V RS232_PASS=xxxxxx which sets the password to the value xxxxxx.
M100A (i.e., operating the port in COMPUTER MODE) we suggest that you order a supplementary manual "The RS-232 Interface", Teledyne API part number 01350. This manual shows additional features of the port designed to support a computer driven interface program.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 6-7: RS-232 Command Summary Commands Definition Print help screen T SET ALL Enable display of all test variables during T LIST T SET name Display only NAME during T LIST...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 6-8: RS-232 Command Summary Terminal Mode Editing Keys Definition V MODE Print current analyzer mode Backspace Erase line Recall last command Execute last command Execute command Switch to computer mode...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 6-9: RS-232 Interface Command Types First Character Message Type Calibration status Diagnostic DAS report Test measurement Variable Warning There are 6 different types of messages output by the M100A. They are grouped below by type in Table 6-9 - Table 6-14.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6.8.3 TEST Commands and Messages Table 6-10: RS-232 Test Messages Name Message Description RANGE RANGE=xxxxx PPB Analyzer range STABILITY STABIL=xxxx.x PPB Std. Deviation of last 25 SO concentration values SAMPPRESS PRES=xxx.x IN-HG-A...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G The T command lists TEST messages. Examples of the T command are: T LIST Lists test messages currently enabled with T SET T LIST ALL Lists all test messages...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in the normal mode (i.e. not in quiet mode) the warning message is also sent to the RS-232 interface.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G An example of an actual sequence of calibration status messages is: C DDD:HH:MM IIII START MULTI-POINT CALIBRATION C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION There are several methods of both checking the calibration and calibrating the M100A, these are discussed in Section 7.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G The following is a summary of the Diagnostic commands. Table 6-14: RS-232 Diagnostic Command Summary Command Description D LIST Prints all I/O signal values. See Table 9-4 for Sig I/O definitions.
Page 78
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G In all of the commands, brackets ([ ]) denote optional parameters. The ID parameter is the instrument ID, useful when the multi-drop protocol is being used. The NAME parameter is the data channel’s name.
Page 79
A data report format looks like the following: 31:10:06 0412 CONC : AVG CONC1=6.8 PPB This report uses the traditional Teledyne API format of a leading first character (“D” in this example), a time stamp (“31:10:06”), and the instrument ID (“0412”). The other fields in the report are the data channel name (“CONC”), the sampling mode (“AVG”), the parameter...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 6.8.8 VARIABLES Commands and Messages Table 6-16: RS-232 Operating Modes Mode Description ZERO CAL A Automatic zero calibration ZERO CAL R Remote zero calibration ZERO CAL M Manual zero calibration...
Page 81
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G A list of variables and their settings can be requested over the RS-232 port by: V LIST Lists internal variables and values The output from this command is quite long and will not be shown here. The general format of the output is: name = value warning_lo warning_hi <data_lo>...
Page 82
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G INTENTIONALLY BLANK 6-26...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7 CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check and adjust the calibration of the M100A. These different methods are summarized in the following Table 7-1. In addition, all of the methods described in this section can be initiated and controlled via the RS-232 port, see Section 6.8 for details.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-1: Types of Zero/Span Checks and Calibrations Section Type of Cal or Check Description Manual Z/S Check - Cal gas This calibration option expects the calibration gas through sample port coming in through the sample port.
Page 85
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 7-1: Model 100A Calibration Setup...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.1 Manual Zero/Span Check or Calibration through the Sample Port The zero and span calibration of the instrument can be checked or adjusted using gases supplied through the normal sample port. This method is often used when the calibration gas is supplied from an external calibrator system.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Enter the expected SO span gas concentration: Table 7-3: Enter Expected Span Gas Concentrations Procedure Step Number Action Comment Press CAL- This key sequence causes the M100A to prompt for the...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.2 Manual Zero/Span Check or Calibration with Zero/Span Valves Option The Zero/Span valve option can be operated from the front panel keyboard. In the Zero/Span valve option the zero and span gas come into the valves through ports on the rear panel of the instrument.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.3 Manual Zero/Span Check with IZS Option The Internal Zero/Span (IZS) system can be operated from the front panel keyboard. When the M100A is in the SAMPLE mode, and the IZS option is installed, the CALS (Calibrate-Span) or CALZ (Calibrate-Zero) buttons will be visible.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.6 Calibrate on SO Permeation Tube If the M100A is equipped with the IZS option, it is possible to span the instrument on the permeation tube(refer to Section 6.5). If this feature is enabled, the instrument internal valves are set so that SO gas is routed through the permeation tube.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-7: IZS or Z/S Valves Modes with Remote Contact Closure Ext Zero CC Ext Span CC Operation Contact Open Contact Open State when in SAMPLE mode, normal monitoring.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G All monitoring instrument systems are subject to some drift and variation in internal parameters and cannot be expected to maintain accurate calibration over long periods of time. Therefore, it is necessary to dynamically check the calibration relationship on a predetermined schedule.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-8: Activity Matrix for Calibration Equipment & Supplies Frequency and method Action if requirements Equipment/supplies Acceptance limits of measurement are not met Recorder Compatible with Check upon receipt...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-9: Activity Matrix for Calibration Procedure Frequency and method Action if requirements Equipment/supplies Acceptance limits of measurement are not met Calibration gases NIST traceable Assayed against an Working gas standard is...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.8.5 SO Calibration Procedure There are two calibration procedures for M100A such as dynamic dilution procedure using compressed gas cylinder or using permeation tube. The procedures for multipoint calibration of the SO analyzer are specified in the Code of Federal Regulation.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 23. Zero calibration procedure. Table 7-10: EPA Zero Calibration Procedure Step Number Action Comment Press CAL The M100A enters the calibrate mode from sample mode. Allow the zero gas through the sample port.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 7-2: Diagram Of Calibration System Table 7-11: EPA Expected Span Gas Concentration Procedure Step Number Action Comment Press CAL- This key sequence causes the M100A to prompt for the...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-12: EPA Span Calibration Procedure Step Number Action Comment Press CAL The M100A enters the calibrate mode from sample mode. Wait 10 min Wait for readings to stabilize at span values.
Page 99
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 1. Zero calibration procedure is identical to Table 7-11. 2. Generate the SO concentration to approximately 80% of the URL of the range. Calculate the expected concentration as follows.
Page 100
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G By rearranging Equation 7.2 it is possible to calculate total air flow which is sum of F and F total × − × × 382 10 Equation 7.3 total Total air flow rate should exceed the analyzer's sample flow rate demand by margin of 10% to 50%.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.8.6 Calibration Frequency To ensure accurate measurements of the SO concentrations, calibrate the analyzer at the time of installation, and re-calibrate it: 1. No later than three months after the most recent calibration or performance audit which indicated analyzer calibration to be acceptable.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-13: Definition of Level 1 and Level 2 Zero and Span Checks (from Section 2.0.9 of Q.A. Handbook for Air Pollution Measurement Systems) LEVEL 1 ZERO AND SPAN CALIBRATION...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.8.8 Summary of Quality Assurance Checks The following items should be checked on a regularly scheduled basis to assure high quality data from the M100A. See Table 7-14 for a summary of activities also the QA Handbook should be checked for specific procedures.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.8.9 ZERO and SPAN Checks A system of Level 1 and Level 2 zero span checks (see Table 7-1) is recommended. These checks must be conducted in accordance with the specific guidance given in Subsection 9.1 of Section 2.0.9 (Q.A.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.8.9.2 Precision Check A periodic check is used to assess the data for precision. A one-point precision check must be carried out at least once every 2 weeks on each analyzer at an SO concentration between 0.08 and...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 7-15: NIST-SRM's Available for Traceability of Calibration and Audit Gas Standards NIST-SRM Type Nominal concentration 1693 in N 50 ppm 1694 in N 100 ppm Permeation Tubes Permeation rate,...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G M100A with IZS option can be used to calibrate zero/span of one Range #, but not both. Other Range # should be calibrated using external zero/span gas through sample port. Pressing CALZ (for zero) or CALS (for span) keys will lead to show #1 and #2 the same way as CAL key except CALZ is dedicated for zero air calibration while CALS is dedicated for span gas calibration.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 7.11 References 1. Environmental Protection Agency, Title 40, Code of Federal Regulations, Part 50, Appendix A, Section 10.3. 2. Quality Assurance Handbook for Air Pollution Measurement Systems - Volume II, Ambient Air Specific Methods, EPA-600/4-77-027a, 1977.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8 MAINTENANCE NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. 8.1 Maintenance Schedule Table 8-1: Preventative Maintenance Schedule Item Maintenance Interval...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8.2 Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to absorb SO , thus causing those readings to be lower than the actual value.
Page 111
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 8-1: Replacing the Particulate Filter...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8.3 Replacing the Permeation Tube (option) The permeation tube is contained in the oven at the rear left of the instrument, refer to Figure 2-5 for its location. 1. Turn off the instrument power.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 3. Replace with charcoal. 4. Re-install felt pad and re-tighten the cap. Make sure the o-ring in the cap is in good shape and squarely seated. 5. Leak check the assembly, then re-attach scrubber to fitting and clip on the bracket.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8.6 Cleaning the Reaction Cell The reaction cell should be cleaned whenever troubleshooting points to it as the cause of the trouble (refer Figure 8-2). A dirty cell will cause excessive noise, unstable span or zero, high stray light, or slow response.
Page 115
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 8-2: Reaction Cell...
Page 116
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 8-3: Pneumatic Diagram...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 8.8 Leak Check Procedure There are two methods of leak checking. First one is the vacuum method which is simplest but it does not show the location of a leak. This vacuum method described below is the general method when a Leak Checker (refer to Figure 9-10) is not readily available.
4. Remove the board, laying it down on an insulating surface such that the board edge pins on the PCB are on the left. The EPROM chip should be at the top center. The current chip should be labeled with something like "API SO H.1 - - -". See Figure 9-1 for location of prom on CPU card.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9 DIAGNOSTIC, TROUBLESHOOTING NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. This section of the manual contains information on diagnosing and corrective action procedure for the instrument performance problems.
Page 120
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 29. Check the TEST functions: A. Compare the TEST functions to the factory values in Table 2-1. This will often provide important clues as to the problem. B. Check for the sign of the drift, particularly the slope and offset readings: The slopes are the software equivalent of the span pot on an analog instrument.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1 Operation Verification - M100A Diagnostic Techniques 9.1.1 Fault Diagnosis with TEST Variables The Table 9-1 indicates possible fault conditions that could cause the TEST functions to be outside the acceptable range.
Page 122
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment PRES Check value in Reaction cell pressure is measured to monitor sample pressure Final Test and to compensate any sample pressure variation. Reaction cell...
Page 123
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment 0 - 5000 mV This is the instantaneous output of the PMT. The PMT voltage values will be relatively constant when: Electric test - variation in the 2000 mV signal observed will be noise of the V/F board and preamp noise.
Page 124
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment STR LGT 0 - 100 ppb Stray light is the background light of the reaction cell expressed in ppb while sampling zero gas. High stray light could be caused by : Aging UV filter.
Page 125
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment SLOPE 1.0 ± 0.3 The slope can be thought as a gain term which determines the steepness of the calibration curve.
Page 126
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment RCELL 50 ± 1 The reaction cell temperature is controlled to 50 C ± 1 C by the TEMP computer.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1.2 Fault Diagnosis with WARNING Messages The M100A monitors several internal values for alarm conditions. If the condition for an alarm is met, the alarm is displayed on the front panel and the warning is transmitted out the RS-232 port.
Page 128
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-2: Front Panel Warning Messages (Continued) Message Description IZS TEMP WARNING IZS temp. out of spec. The warning message is most often present during initial warm-up or poor electrical contact.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1.3 Fault Diagnosis using DIAGNOSTIC Mode Diagnostic mode can be looked at as a tool kit of diagnostics to help troubleshoot the instrument. To enter DIAG mode press: SETUP-MORE-DIAG The diagnostic modes are summarized in Table 9-3.
Page 130
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-3: Summary of Diagnostic Modes DIAG Mode Description SIGNAL I/O Gives access to the digital and analog inputs and outputs on the V/F board. The status or value of all of the signals can be seen.
Page 131
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1.3.1 Signal I/O Diagnostic Table 9-4: Diagnostic Mode - Signal I/O Signal Control Description DSP_BROWNOUT Display brownout is used to keep the display from getting corrupted during low line voltage conditions.
Page 132
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description BROWNOUT_RESET Brownout reset works in conjunction with DSP_BROWNOUT. When DSP_BROWNOUT is set the CPU sends a signal to reset the display and clear the DSP_BROWNOUT.
Page 133
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description ST_DIAG_MODE Status Bit - In Diagnostic mode Logic High = M100A in Diagnostic mode Logic Low = Not in Diagnostic mode...
Page 134
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-4: Diagnostic Mode - Signal I/O (Continued) Signal Control Description DAC_CHAN_2 Output of spare (DAC2) channel in mV. DAC_CHAN_3 Test Channel (DAC3) output. UVLAMP_SIGNAL Current UV lamp signal voltage in mV. Same as UV LAMP in TEST menu.
Page 135
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G NOTE If the SO concentration display is not within the nominal range of 1000 ± 100 ppb, then the lamp ratio may need to be updated. See Section 9.1.6, step 2 of PMT Calibration procedure.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1.3.5 DAC Calibration The Digital to Analog Converters (DAC) are calibrated when the instrument is set up at the factory. Re-calibration is usually not necessary, but is provided here in case the V/F board needs to be replaced and re-calibrated.
Page 137
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 9-5: Model 100A Variables Default Value Name Units Description Value Range DAS_HOLD_OFF 15.0 0.5-20.0 Hold off duration after calibration or diagnostic mode. TPC_ENABLE ON-OFF Temp/Pres compensation enable RCELL_SET...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.1.5 Test Channel Analog Output Many of the TEST functions have an analog voltage associated with them. As a diagnostic aid it is possible to route any one of the various test voltages out the 4th analog output port(see Figure 2-2).
Analyzer Instruction Manual, 02164, Rev. G 9.1.6 Factory Calibration Procedure (Quick Cal) This procedure is referred to in other Teledyne API instruments as the "Quick Cal" procedure. It is used at the factory when the instrument is first set-up. The Factory Cal procedure balances the PMT, preamp, and software gain factors so the instrument has optimum noise, linearity, and dynamic range.
Page 140
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 31. Recalibrate the instrument for both zero and span to calculate the slope and offset. (Refer to Section 7). Above procedure is to assure that the instrument will operate with optimum noise, linearity, and dynamic range.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2 Performance Problems When the response from a span check is outside the control limits, the cause for the drift should be determined, and corrective action should be taken. Some of the causes for drift are listed below:...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2.2 Flow Check 1. Check TEST function SAMP FLOW - Should be 650 ± 65cc/min. 2. Check that pump is running. Observe pump for proper operation. 3. Test that pump is producing vacuum by removing fitting at the inlet of the pump and checking for suction at fitting.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2.4 Negative Concentration Display 1. Mis-calibration. The 'zero' gas that was used to zero the M100A contained some SO gas - that is it had more SO gas than that of the sample air. May also be caused by doing a zero calibration using ambient air.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2.7 Unstable Zero Common causes are: 1. Leak in pneumatic system. 2. Light leak - check the sensor module with strong light. 3. If equipped with IZS, zero air scrubber exhausted - replace charcoal and check broken DFU filter.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2.9 Inability to Zero If the ZERO button is not illuminated when attempting to zero, that means the reading is outside of the software gain ranges allowed. In an analog instrument it would be the equivalent to the zero pot hitting the maximum.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.2.11 Slow Response 38. Contaminated or dirty sample delivery pneumatics A. Dirty/plugged sample filter or sample lines. B. Dirty reaction cell. C. Check leak. 1. Flow rate too low.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3 Electronic Subsystem Troubleshooting and Adjustments 9.3.1 Computer, Display, Keyboard The purpose of this section is to determine if the computer subsystem electronics hardware are working properly. Assessment will make it the board level.
Page 149
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Like the display, the overall functioning of the CPU can be confirmed by a simple test. 1. Locate the CPU board on the mother board by referring to Figure 2-5.
Page 150
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-1: CPU Board Jumper Settings 9-32...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.1.3 Front Panel Keyboard The keyboard consists of 8 keys and 3 LED's. Key strokes are sent to the SBC40 computer's parallel port. The computer software detects the key strokes via interrupts. The bottom line of the display consists of 40 characters which is divided into 8 - 5 character fields.
Page 152
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-2: RS-232 Pin Assignments 9.3.2.1 RS-232 Connection Examples Example 1 Connecting the M100A (using supplied cable) to an IBM-PC AT compatible computer (DB-25 external connector, or DB-25 end of DB-9 to DB-25 Adapter).
Page 153
If the DB-25 connector pinout shows that data is output on pin 2 (from the printer), then it is DTE and the Teledyne API analyzer’s jumpers should be set to DCE mode. If pin 2 of the printer DB-25 is an input to the printer, then set the jumpers of the analyzer to DTE mode.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.2.2 RS-232 Diagnostic Procedures There are several features of the M100A to make connecting to RS-232 and diagnosing RS-232 faults easier. There are two LED's on the rear panel Connector Board which are connected to pin 2 and 3 of the DB-9 connector on the board.
Page 155
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 41. Analog output channel test. In the DIAGNOSTIC menu on the front panel, there is a test that outputs a step voltage to the 4 analog outputs. This test is useful for calibrating chart recorders and dataloggers attached to the M100A.
Page 156
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.3.1 ADC/DAC Calibration Procedure Due to the stability of modern electronics this procedure should not have to be performed more than once a year or whenever a major sub-assembly is exchanged or whenever analog output voltage range is changed.
Page 157
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 4. Pressing SETUP-MORE-DIAG-ENTR-NEXT, and select ANALOG OUTPUT. Verify the quality of ADC/DAC calibration by measuring the test channel output voltage. The signal consists of a scrolling 0%, 20%, 40%, 60%, 80%, 100% of the analog output value. The scrolling may be stopped by pressing the key underneath the % display to hold that value and will display within the square bracket ( [%] ).
Page 158
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.3.2 Changing Output Voltage Ranges Several different output voltage ranges can be selected by DIP switch setting on the V/F board. See Figure 9-3 for the DIP switch settings. If you change the analog output voltage range, then you must reset the power of the instrument and do a ADC/DAC calibration per Section 9.3.3.1.
Page 159
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-3: V/F Board Dip Switch Settings 9-41...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.4 Status/Temp Board The Status/Temp Board is a multifunction board that: 1. Converts the resistance readings of the thermistors to voltages 2. Provides status output circuitry 3. Provides circuitry for contact closure inputs 4.
Page 161
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G These 3 outputs are controlled by a common pot (R34) on the upper edge of the card. If the temperature readouts are in error: 1. Adjust R34 of the Status/Temp board to read 2.577 VDC between test point 6 and test point 2.
Page 162
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 1. Voltage method - preferred method. Digital type Multi-meters (DMM) are ideally suited for this type of measurement because of their high input impedance - usually 10M ohm or greater. The total load resistance should be between a 200ohm (min) to 1000 (max), precision load resistor of .1% tolerance or better is...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.5 Power Supply Module The Power Supply Module consists of several subassemblies described in Table 9-7 below. Table 9-7: Power Supply Module Subassemblies Module Description Linear Power Supply Board The linear power supply board takes multiple voltage inputs from the power transformer and produces +5, +15, -15, +12 VDC outputs.
Page 164
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-4: Power Supply Module Layout 9-46...
Page 165
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-5: Electrical Block Diagram 9-47...
Page 166
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.5.1 PSM Diagnostic Procedures The Linear Power Supply board can be tested by checking the DCPS - TEST function on the front panel. It should read 2500 mV ± 200 mV. If the value is outside this range, individual output voltages can be tested on connector P3, see Schematic in the Appendix for pinouts.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.6 Flow/Pressure Sensor The flow/pressure sensor board consists of 1 pressure sensor and a flow sensor. See Figure 9-6 for a diagram of this board. From these two sensors pressure and flow rate values are computed and displayed on the front panel TEST functions.
Page 168
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-6: Flow/Pressure Sensor 9-50...
Page 169
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-7: SO Sensor Module 9-51...
Page 170
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-8: SO Sensor Module 9-52...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.3.7 Reaction Cell Temp The reaction cell temperature is controlled by the CPU. It operates by reading a thermistor amplifier on the 01086 Status/Temp board. First LED from the front indicates the control power on-off status which can be seen from the cover of the PSM through the slot.
Page 172
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Remove three screws that hold the sensor assembly (refer to Figure 2-1 they are located adjacent to the shipping screws). Remove four screws that hold the fan duct and unplug cable plugs from the fan duct assembly.
Page 173
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-9: PMT Cooler Subsystem 9-55...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 5. Pull out PMT, LED, and thermistor carefully from the cooler block. Be careful not to contaminate the window of the PMT with thermal compound of thermistor. PMT is fragile, so handle carefully.
Page 175
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9. Use Figure 9-10 to check the voltages at each pin of the supply, and the overall voltage. 10. Turn off the instrument power, and re-connect the PMT tube, then re-assemble the sensor.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.4 Optical Sensor Module Troubleshooting 9.4.1 PMT The PMT detects the light emitted by the secondary emission of the SO molecules. It has very high gain and low noise to detect the weak light source optimally. It is not possible to test the detector outside of the instrument in the field.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. If the UV LAMP display is lower than 1000mV after peak adjustment, it is recommended to replace the lamp.
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 50. Without leaving DIAG menu, Scroll until PMT_SIGNAL is shown. It should display near 0 ± 200 mV. If reading is greater than +200 mV, then verify if + 12VDC is supplied for the terminal located at the shutter cover(see Figure 9-7).
Page 179
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Figure 9-11: Kicker Leak Check 9-61...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G 9.5.4 Z/S Valves & IZS Permeation Tube Oven The Z/S Valves and IZS are both options in the instrument. Before troubleshooting this sub- assembly, check that the options were ordered, and that they are enabled in the software (if not call factory).
Analyzer Instruction Manual, 02164, Rev. G 10 TELEDYNE API M100A SPARE PARTS LIST Note: Use of replacement parts other than those supplied by Teledyne API may result in non-compliance with European Standard EN61010-1. Table 10-1: Teledyne API M100A Spare Parts List...
Page 182
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 10-1: Teledyne API M100A Spare Parts List (Continued) PART NO DESCRIPTION 011390200 Assembly, Power Supply Module - 220V/50Hz 011390400 Assembly, Power Supply Module - 240V/50Hz 011900000 37 mm Sample Filter Assembly...
Page 183
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 10-1: Teledyne API M100A Spare Parts List (Continued) PART NO DESCRIPTION 024720000 Tubing: 6’, 1/8” BLK 024750000 Tubing: 6’, 1/4” TYGON FL0000001 Sintered Filter (002-024900) FL0000003 Filter, DFU (036-040180)
Page 184
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 10-1: Teledyne API M100A Spare Parts List (Continued) PART NO DESCRIPTION OR0000046 O-Ring, Permeation Oven PS0000002 High Voltage Power Supply PS0000010 15V Switching Power Supply PU0000020 Pump, 115V 50/60 Hz...
Page 185
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 10-3: Teledyne API M100A 37 mm Expendables Kit PART NO. DESCRIPTION 014740000 M100A 37 MM FILTER EXPENDABLES KIT INCLUDES:` 003690100 Filter, TFE, 37 mm, Qty. 25 (872-006300) FL0000001...
Page 186
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G Table 10-6: Teledyne API M100A Spares Kit For 1 Unit PART NO. DESCRIPTION 014760100 M100A Spares Kit for 1 Unit INCLUDES:` 003290000 Thermistor Assembly (885-071600) 012360000 Fan, Power Supply Module...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G APPENDIX A MAINTENANCE SCHEDULE FOR M100A DATE INSTRUMENT RECEIVED _____________ JUNE JULY SEPT PARTICULATE REPLACE WEEKLY AS NEEDED. FILTER ELEMENT CHARCOAL ZERO REPLACE QUARTERLY. AIR SCRUBBER REPLACE ANNUALLY. PARTICULATE...
Page 188
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G INTENTIONALLY BLANK...
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G APPENDIX B ELECTRICAL SCHEMATICS Part Number Name 00402 Sensor Board Assembly 00403 Sensor Board Schematic 00514 V/F Board Assembly 00515 V/F Board Schematic 00704 Keyboard, non CE 00705 Keyboard Schematic, non CE...
Page 190
Teledyne API Model 100A SO Analyzer Instruction Manual, 02164, Rev. G INTENTIONALLY BLANK...
Need help?
Do you have a question about the 100A and is the answer not in the manual?
Questions and answers