Freezer Parts And Control Panels - Haier DW-86L829BP Operation Manual

Ultra low temperature (ult)freezer
Hide thumbs Also See for DW-86L829BP:

Advertisement

Setting the low temperature alarm
1. In unlocked mode, press "Set" key to select "Low Temp. Alarm", the temperature display flashes
and displays the setting value.
2. Then, press '
' or '
' key, adjust the low temperature alarm setting value.
Temperature setting range: above -91℃ and at most -5℃ above the inner temperature.
3. After adjusting, do not touch the unit for 10 seconds. The unit automatically enters the locked
mode and the temperature display stops flashing which means the value set have been input into
the computer. Otherwise, the setting is invalid
For example: Set the low temperature alarm to -91℃.
User parameters
1. Parameters: dA, T1, T2, P6, IC, PS1, CL1.
2. Entry mode: After the display board unlocks and the unlocking indicator lamp is OFF, long press
the "
" to 5s and enter into user parameters;
3. Press the"
" or "
" key to select dA, T1, T2, P6, IC , PS1 and CL1 after accessing to the
user parameters;
select one of these parameters and then press the "Set" key to modify this parameter.
Regulation mode and scope of parameter value
1. dA: 5 min is the default value for the delayed door open alarm. It is adjustable from 1min to 30min;
Press the "
" or "
" key to modify the set value of delayed door open alarm;
2. Tl: 6 min is the default value for the usb data reading cycle. It is adjustable from 1min to 99min;
Press the "
" or "
" key to modify the set value of usb data reading cycle;
3. T2: usb timing, MM(P2:01~12 ) /DD(P3 : 01~31)/ YYYY(Pl:10~99 )/Hour(P4 : 00-23 )/
Minute(P5 : 00-59);
Press the "
" or "
" key to modify the timing of usb;
4. P6: 12 is the default value for USB derivative mode; it is adjustable from 0 to 12;
The data can be exported in different time frames with parameters of 0~12, among which, 12, as
default, means exporting the data generated from 1 year by one time; 0 means exporting all data
by one time; 1-12 means exporting the data generated from previous months, such as 2 months,
3 months...12 months.
16

Freezer parts and control panels

Freezer Parts
超低温保存箱使用操作步骤
警告:首次开机空箱运行到设定温度,持续运行24小时后,再分批次放入物品。
解锁
按"  "或"  "键,显示板LED箱内温度显示区显示"00"。
按"  "调节至" 06 "(默认值,可调)。
按住"
"键 5秒钟,解锁显示板,显示板锁定指示灯灭。再按一下"
"键,
进入温度设定模式;
温度设定操作
在解锁状态下按下"设置" 键,分别选择"箱内温度设定"、"高温报警设定"、
"低温报警设定"。
相应指示灯亮了以后,按"  "或"  "键调整所需温度。
修改完毕后,按"
"键保存箱内温度设定值;若20秒无按键,箱内温度设定值不保存,
功能锁定后返回显示箱内温度。
注意:其他内容请详见说明书。
Pocket organizer
超低温保存箱注意事项
首次安装或搬动保存箱后,请顺时针旋转水平支脚使其支撑地面,确保保存箱不移动,安装固定
到位后需要静置24小时后再通电使用。
为保证空气流通散热,保存箱四周至少保留30cm的间隙。
保存箱使用环境温度为10~32℃,最佳环境温度为18~25℃,必要时应使用空调系统。湿度应低
于80%RH,环温32℃时,湿度应低于57%RH。
每台保存箱应该独立使用一个插座,且电源插座承受电流应大于16A,并可靠接地。
通电时先打开保存箱电控箱上的电池开关,需长时间关机时,先拔掉电源再关掉此开关。
document storage
保存箱每次开门时间尽量不超过1分钟,关门前如门封条上有冰水,建议擦拭干净,以保证密封效
果良好。
为了延长保存箱使用寿命和降低能耗,使用时,在保证存储物品安全的前提下,建议将保存箱的
温度设定在-50℃~-80℃之间为最优。
本保存箱适用于在低温下存储物品,不宜作为速冻箱使用,不能强制快
速冷冻大量较热物品或大体积液态物品。
如果在通电2到3个小时后保存箱不制冷,请拔掉电源,并尽快与售后
联系。
在遇到报警故障或者其他故障时,请参照说明书按照显示板上的提示
消除故障,如不能自行排除故障,请不要擅自拆卸,及时联系售后维
修人员协助消除故障。
海尔生物医疗 智护生命科学
LED board
Finger print
module(optional)
Door handle
(with lock)
Door card
board (optional)
Front protection
panle of refigeration
compartment
Leveling leg Caster
USB port
13