Page 1
Model 622B01 Platinum Precision Industrial ICP® Accelerometer Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-959-4464 24-hour SensorLine: 716-684-0001 Fax: 716-684-3823 E-mail: imi@pcb.com Web: www.imi-sensors.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
Page 5
Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 6
Piezoelectric ICP® Accelerometers Operating Guide Operating Guide with Enclosed Warranty Information 3424 Walden Avenue, Depew, New York 14043-2495 Phone (716) 684-0003 Fax (716) 684-3823 Toll Free Line 1-800-959-4IMI MANUAL NUMBER: 18405 MANUAL REVISION: A ECN NUMBER: 49766...
Page 7
General OPERATING GUIDE for use with PIEZOELECTRIC ICP® ACCELEROMETERS SPECIFICATION SHEET, INSTALLATION DRAWING AND CALIBRATION INFORMATION ENCLOSED IMI ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT OF PROCEDURES THAT ARE INCONSISTENT WITH THIS OPERATING GUIDE 1.0 INTRODUCTION Congratulations on the purchase of a quality ICP®...
The environment of the application is a critical input voltage. Also, the diode can be changed to supply consideration during program implementation. The higher currents for driving long cable lengths. Constant sensor chosen must be capable of surviving the wide current diodes, as shown in Figure 3, should be used in range of conditions to which it is subjected;...
MS – Mine Safety Approved Sensors 4.1 STANDARD STUD MOUNT TO – Temperature Output Sensor This mounting technique requires smooth, flat contact surfaces for proper operation and is Note: Not all sensors are available with the optional recommended permanent and/or secure prefixes.
semi-permanent mounting, substitute epoxy or another METHOD 2 – Direct Adhesive Mount type of adhesive. For restrictions of space or for convenience, most sensors STEP 3: Hand-tighten the sensor/mounting stud to the (with the exception of integral stud models) can be machine, then secure the sensor with a torque wrench to adhesive-mounted directly to the machine surface.
Page 11
ICP® accelerometers are internally amplified, two-wire accelerometers. Connections to the sensor require two leads: one for the power and signal, and the other for the common and signal return. Often, coaxial cables are used since only two conductors are needed. Coaxial cables are less expensive.
Page 12
(Equation 1) exceed its maximum temperature specification. For this 10 9 max= 2 reason, do not supply excessive current over short cable runs [ −1 ] or when testing at elevated temperatures. Where, f = maximum frequency (hertz) 5.1 CABLE CONNECTOR PROCEDURE C = cable capacitance (picofarads) Care and attention to installation is essential, as the...
6.0 POWERING the system still does not operate properly, consult All ICP® sensors require constant current excitation an IMI Application Engineer. for proper operation. For this reason only use approved constant-current sources. A typical system Note: Always operate the accelerometer within the schematic is shown in Figure 11.
Periodic recalibration may be required by plants with piezoelectric accelerometers. This method relies on strict certification and traceability requirements. It is a simple comparison to a previously calibrated always recommended that the user has the sensor accelerometer, typically referred to as a reference recalibrated periodically, particularly if the sensor standard, as shown in Figure 12.
ICP® sensors are two-wire sensors. They are powered with a constant current DC source. The power supply is typically 18 to 30 volts DC current limited via a constant current between 2 and 20 mA. Typical battery operated supplies offer 2mA of constant current to extend battery life while Figure 14.
11.0 CUSTOMER SERVICE/WARRANTY IMI, a division of PCB Piezotronics, guarantees Total Customer Satisfaction. If, at any time, for any Figure 18. DC Bias Voltage Measurement reason, you are not completely satisfied with any IMI product, IMI will repair, replace or exchange it at no charge.
Page 18
Model Number Revision: B PLATINUM PRECISION INDUSTRIAL ICP® ACCELEROMETER 622B01 ECN #: 47050 Performance ENGLISH OPTIONAL VERSIONS Sensitivity(± 5 %) 100 mV/g 10.2 mV/(m/s²) Optional versions have identical specifications and accessories as listed for the standard model Measurement Range ± 50 g ±...
Page 19
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. REVISED TITLE 48794 SIGNAL/POWER GROUND ELECTRICAL CONNECTOR 5/8-24 UNEF - 2A 2-PIN (MIL-C-5015) 2.06 [52.2]...