Compact high voltage porosity (holiday) detectors (16 pages)
Summary of Contents for PCWI Compact DC30
Page 1
Can be used in accordance with: Australian Standard AS3894.1 ASTM G62, ASTM D5162, ASTM D4787, NACE SP0274, NACE SP0490, NACE SP0188, ANSI/AWWA C213, ANSI/AWWA C214, ISO 2746 CE Marked PCWI Compact Detectors comply with the requirements of EMC Directives 89/336/EEC EMC and its amending directives.
Page 2
POROSITY DETECTOR ACCURACY: +5% of reading or + 50V below 1kV Note: at the time PCWI tested the Porosity Detector at the reference temperature & humidity). Refer to the PCWI Manufacturer’s Calibration Certificate provided with the Porosity Detector. Range: 0kV to 30kV...
Please consult these operating instructions before use. CONTENTS 1.0 Safety Precautions 2.0 Operation 3.0 Specifications 4.0 Control Panel Layout 5.0 Voltage Recommendations 6.0 Troubleshooting 10-11 7.0 Optional Accessories 12-13 8.0 PCWI LiFePO4 Battery 9.0 Warranty 10.0 Service & Maintenance...
Page 4
Electrodes (probes) - extending the versatility of the Compact from large to small and from accessible to inaccessible surfaces. PCWI, in a continuing desire to achieve the maximum in corrosion detection competence, welcomes user enquiries and recommendations. Yours sincerely...
1.0 SAFETY PRECAUTIONS: All hand-held high voltage test equipment should be operated by responsible, trained and authorised personnel. The unit must be earthed to both the item under test and to ground. CAUTION The Detector output can be up to 35,000 volts. Should the operator accidentally make contact with the test electrode, they may experience a mild shock or zap, and in order to avoid this possibility, the wearing of rubber gloves is recommended.
Coating Thickness Range Applied coats should be cured, thickness tested, visually inspected and accepted – before high voltage porosity testing is carried out. Coating thickness should be above 150µm; coatings below this thickness should be tested with a wet sponge unit. 2.0 OPERATION Connect the probe and earth leads to the unit.
Earthing Where the item to be tested is not earthed to ground, a ground spike must be attached. The unit should always be switched off before removing and repositioning the earth lead. After the earth is repositioned, the probe should always be flashed on the substrate to prove a good contact has been made.
4.0 CONTROL PANEL LAYOUT LCD display (including battery condition indicator) Voltage control (10 turn) Visual alarm indicates when fault is found On switch Off/test switch Sensitivity control for the alarm Audible alarm when fault is found High Voltage probe connector Fuse (1.6A slow blow) 5 x 20mm 10 Earth connection point 11 Slide-off power pack...
5.0 VOLTAGE RECOMMENDATIONS International Standards NACE SP0188 derived table Table 1. below is derived from NACE standards and should be used as a guide only. New protective Coatings on Conductive Substrates Table 1 Suggested for High Voltage Spark Testing Total Dry Film Thickness Suggested Inspection (µm) (mil)
Page 10
International Standards NACE SP0490 and SP0274 derived tables Tables 2 & 3 below are derived from NACE standards and should be used as a guide only. Thin Film Pipeline Coatings (FBE) General Pipeline Coatings Table 2 Table 3 kV Values from NACE SP0490 kV Values from NACE SP0274 kiloVolts kiloVolts...
6.0 TROUBLESHOOTING In order to ensure that your Compact Porosity Detector is operating to specification, the Detector handle and lead should be checked prior to use. Use the following procedure: Unscrew the neon holder from the handle. Using a multimeter, complete the circuit and check the resistance at both the handle 1 and the high voltage plug 2.
Page 12
Symptom Cause Solution No Display Flat battery Recharge battery No Power - fuse not fitted Fit fuse Display shows -1 Voltage higher than display Decrease voltage capacity Display erratically shows -1 Display drops to -1 when fault Normal operation found Display voltage drops Slightly conductive, damp or Wash and clean the surface...
DC Detector output voltage. DC Crest Meter includes Calibration Certificate. Brushes & Coils PCWI manufacture brushes and coils to suit your needs. 50mm (2in) to 600mm (24in) wide flat brushes. 25mm (1in) to 500mm (20in) Spiral wound and Disc Internal circular pipeline brushes.
Page 14
60mm, 125mm, 200mm, 450mm connectors for flat or external pipeline brushes. Coil Joiners Fan Brushes Additional Accessories Spare slip-out power packs. 7m Earth lead with clamp. 10m trailing Earth lead (1m plastic encased). Spare neon light holders and neon's See www.pcwi.com.au for more details.
IATA shipping lithium battery guidelines. Check with your freight company to determine any specific requirements. Use the PCWI LiFePO4 Charger to charge LiFePO4 battery. 1) The charger plugs into the side of the battery 2) LED on charger lights up Red when Charging 3) LED on Charger lights up green when battery is charged.
This warranty does not cover any service that is needed after an accident, alterations, misuse, fire or floods. This warranty is the only one given by PCWI and no one has the authority to change, or add to, the obligations and liabilities listed in it.
10.0 SERVICE AND MAINTENANCE AUTHORISED SERVICE REPRESENTATIVE To enable speedy “return to service” whether under warranty or otherwise, PCWI have appointed your distributor as a service centre and have provided all relevant information and recommended parts to be carried to assist distributor’s technical staff carry out this essential part of the PCWI customer service.