Aegis EL-RG-9H Operation Manual

Aegis EL-RG-9H Operation Manual

Pharmaceutical refrigerator

Advertisement

Certificate of Quality
Checker:
AEGIS SCIENTIFIC
1 Ivybrook Blvd, Unit 145R, Ivyland, PA 18974, USA
Tel: 1-800-796-2344 / Fax: 732 879 0248
Website: www.aegisfridge.com
Pharmaceutical Refrigerator
Operation Manual
Read the Operation Manual carefully before using your appliance.
Keep the Operation Manual in a safe place.
Appearance , color and layout of the door may vary.
Translation of the original instruction.
This product is only applicable for storage of pharmaceutical products or drugs. Not
applicable for storage of blood or blood products including biological products, for
example biological vaccine, agents for in vitro diagnosis and etc.
Model:
EL-RG-9H
EL-RG-12H
EL-RG-14H
EL-RG-16H
EL-R-16H
EL-RG-23H
EL-RG-31H
EL-RG-33H
EL-R-33H

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EL-RG-9H and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Aegis EL-RG-9H

  • Page 1 Pharmaceutical Refrigerator Operation Manual Model: EL-RG-9H EL-RG-12H EL-RG-14H EL-RG-16H EL-R-16H Certificate of Quality EL-RG-23H EL-RG-31H Checker: EL-RG-33H EL-R-33H AEGIS SCIENTIFIC 1 Ivybrook Blvd, Unit 145R, Ivyland, PA 18974, USA Tel: 1-800-796-2344 / Fax: 732 879 0248 Website: www.aegisfridge.com Read the Operation Manual carefully before using your appliance.
  • Page 2: Packing List

    EL-RG-23H EL-RG-31H Plastic bag EL-RG-33H EL-R-33H Shelves Label strip Manufactures Name : Aegis Scientific Inc. Drain hole cap Address : 1 Ivybrook Blvd, Unit 145R, Ivyland, PA 18974, USA : 1-800-796-2344 / Fax: 732 879 0248 Fixed card Brackets (for rear wall distance)
  • Page 3 User-friendly design, computer control, smart and carefree, adjustment User-friendly Design not required. High performance thermal insulation. Electric heated glass door with multiple anti-condensation technologies(except EL-RG-9H/14H). Automatic removal of condensate water. Note: Technical information might be somewhat different on your refrigerator than published due to continuous improvement.
  • Page 4: Specification

    Specification Name Pharmaceutical Refrigerator Model EL-RG-33H EL-R-33H Exterior Dimensions 1130×755×1980 (W×D×H) (mm) Interior Dimensions (W×D×H) 1030×590×1425 Content: (mm) Effective Volume 890L Product Feature............Door Glass door with electric heat Solid door Content..............Insulation CFC-Free foamed-in-place urethane Safety Precautions........... Compressor High Quality Hermetically sealed compressor Shelves 12 Polyester coated wire shelves Application guidelines..........
  • Page 5: Safety Precautions

    Safety Precautions Dear users: Specification Thanks for your choice of Aegis pharmaceutical refrigerator, please make sure you have carefully read and observed the contents with following signs in the manual, for better understanding of Name Pharmaceutical Refrigerator this manual and better use of this product, so as to prevent personal injuries and refrigerator...
  • Page 6 Specification Do not place containers of water or heavy objects on the refrigerator. Falling objects may cause personal injury and overflown water may damage the insulation to cause electric shock and fire. Name Pharmaceutical Refrigerator Do not ground the refrigerator through gas pipes, water supply pipes, telephone lines or EL-RG-23H Model lighting conductors.
  • Page 7 Specification Name Pharmaceutical Refrigerator Always check the settings in the controller after restarting a unit from a power outage or shut off. Change of settings may cause damage to the products stored. EL-RG-12H EL-RG-16H EL-R-16H Model Exterior Dimensions If the power is shut off, let the unit sit for 5 minutes before it is powered up again to avoid 665×710×1965 665×710×1965 665×710×1665...
  • Page 8: Application Guidelines

    The refrigerator is designed to operate at a condition of 16 ℃ to 32 ℃ and humidity less than Interior light Fluorescent lamp 15W 85%RH (for EL-RG-9H/14H,the humidity should be less than 70%RH). Small amount of Net Weight 88kg 99kg condensation can occur on the surface of the unit if the actual condition is outside of this range.
  • Page 9: Product Installation

    Ambient temperature: 16 ℃ to 32 ℃, the range of 18 ℃ to 25 ℃ is optimal. Air conditioning system is required as necessary. Ambient humidity: below 85%RH(for EL-RG-9H/14H,the humidity should be less than 70% RH). Avoid excessive dust. Avoid mechanical vibration.
  • Page 10: Preparation Before Usage

    Preparation before usage Circuit diagram(EL-RG-31H) 1. Remove all packaging materials. For safety reason, the equipment of EL-R/G 31H/33H is situated on a wooden pallet and secured by metal brackets. Please remove the screws and place the metal brackets under the unit.(Pic.1). Forklift or specialty lifting equipment should be Pic.1 used to remove the unit off the pallet.
  • Page 11 5. Standing Circuit Diagram(EL-RG-23H) Do not immediately power up the refrigerator after it is installed. Let the unit stand still for 24 hours and then power it up to ensure the unit works properly. 6. Installation of Shelves and Label strip EL-R/G 9H,14H,23H,31H,33H Take out the shelf, label strip and rack fastener from the refrigerator.
  • Page 12 8.Back Bracket(EL-R/G 9H/12H/14H/16H) Circuit diagram(EL-R/G 12H/16H) EL-RG-9H/14H Install the support bolts into the mounting holes on the back of the refrigerator. (See Pic.3) Pic.3 EL-R/G 12H/16H Remove four M5 bolts from the back of cabinet,then fix the back bracket for rear wall distance with the removed bolts.(See Pic.4)...
  • Page 13: Circuit Diagram

    Circuit Diagram Initial Power-on Circuit diagram(EL-RG-9H/14H) Follow the flow chart below to start the refrigerator. While the refrigerator is empty, connect the power line to the appropriate dedicated socket and make sure the power supply in the socket matches the required voltage and frequency.
  • Page 14 If you have any questions about the operation of the refrigerator, please refer should be checked and recorded using appropriate methods. When the refrigerator to the table frequent asked questions as follows. Call Aegis technical support temperature is out of specification, take proper actions to protect the stored samples.
  • Page 15: Component Names Control Panel

    You can purchase a high quality battery from a store for the same size and voltage(9V). You may EL-RG-9H/14H also call Aegis to purchase a high quality replacement battery. The lead time to deliver the battery is Light switch 15 days after receiving the payment. Please call Aegis for details.
  • Page 16: Cleaning And Maintenance

    When you are running short of the chart paper, please contact Aegis to purchase more. The lead time to deliver the charts is 15 days after receiving the payment . Please call Aegis for details.
  • Page 17 If the alarm condition still exists, the buzzer will return to the alarm mode again USB port automatically 20 minutes later. EL-RG-9H/14H series does not have the remote alarm function. Battery switch Remote Alarm Terminals EL-R/G 12H/14H/16H/31H/33H has remote alarm terminals:...
  • Page 18 Alarm Alarm or Safety Phenomenon Alarm Indication Buzzing Alarm EL-RG-23H If upper temperature sensor or lower High temperature sensor perceives Alarm light flashes Impulsive sound alarm temperature temperature of the refrigerator ≥8℃ Light switch Control panel If upper temperature sensor or lower temperature sensor perceives Alarm light flashes Impulsive sound alarm...
  • Page 19 EL-RG-31H 1. If control buttons are not touched in 60 seconds, all changes of parameters will not become effective. The previous settings are restored. Light switch USB port 2. If the unit is powered off before the parameters are saved, the control parameters in Door lock Control panel the RAM will be lost.
  • Page 20 shows“1P” once again. Users can set 1P-5P again. After the setting, press the “Cal ” for 5s, and all s EL-RG-33H ettings will be automatically saved. And then the display area of inner temperature shows the inner temperature normally. Light switch Note: during the said 1P-5P setting process, parameters can be specifically regulated via the Door lock Control panel...
  • Page 21 EL-R-33H USB Function Functions of USB interface The computer has a USB function. Test data can be exported via the USB port. The computer can Light switch Control panel automatically collect and store the test data within the recent 10 years. It collects data every other 6 minutes and stores them automatically.
  • Page 22: Control Panel

    Calibration of Recorder Control Panel The temperature recorder has been accurately adjusted before leaving the factory even if the power supply is interrupted. If there is a need to be adjusted, perform the following procedures. 1. Let the unit run for at least 2 additional hours after it reaches stable condition, EL-R/G 9H/12H/14H/16H/31H/33H 2.
  • Page 23: Application Method

    Application method Maintenance and replacement of the recorder's standby battery Temperature display When the recorder’s LED light blinks green, check the recorder performance as follows. 1) Remove the standby battery. If the LED light is off and the recorder stops working, the This refrigerator has been set capable of reaching appropriate temperature (2~8℃) automatically main power supply of the recorder fails.
  • Page 24 Temperature Recorder (optional) TEMP. SET Chamber temperature recorder with 6 inch standard panel is optional for EL-RG-23H to If you want to reset the temperature, for example, 5℃ is set originally and you want to change the record chamber temperature change. temperature to 5.5℃, please operate according to the following steps: Installation and operation TEMP.

Table of Contents