HOBO® Temp Data Logger (UX100-001) Manual
HOBO Temp Data Logger
UX100-001
Included Items:
• Command™ strip
• Double-sided tape
• Hook & loop strap
Required Items:
• HOBOware 3.4 or later
• USB cable (included with
software)
16208-B
The HOBO Temp data logger records temperature in indoor environments with its integrated
sensor. Using HOBOware®, you can easily configure the logger alarm to trip for specific high or
low temperatures. Or, you can set up burst logging in which the logger records data at a different
interval during certain conditions. The logger can also calculate minimum, maximum, average, and
standard deviation statistics. This compact data logger has a built-in LCD screen to monitor
temperature, logging status, battery use, and memory consumption in between readouts.
Specifications
Temperature Sensor
Range
Accuracy
Resolution
Response Time
Drift
Logger
Logger Operating Range
Logging Rate
Logging Modes
Memory Modes
Start Modes
Stop Modes
Restart Mode
Time Accuracy
Battery Life
Battery Type
Memory
Download Type
Full Memory Download Time
LCD
Size
Weight
Environmental Rating
Plot A: Temperature Accuracy and Resolution
-20° to 70°C (-4° to 158°F)
±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
0.024°C at 25°C (0.04°F at 77°F); see Plot A
8 minutes in air moving 1 m/s (2.2 mph)
<0.1°C (0.18°F) per year
Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing)
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
1 second to 18 hours, 12 minutes, 15 seconds
Normal, burst, or statistics
Wrap when full or stop when full
Immediate, push button, date & time, or next interval
When memory full, push button, or date & time
Push button
±1 minute per month at 25°C (77°F), see Plot B
1 year, typical with logging rate of 1 minute and sampling interval of
15 seconds or greater
One 3V CR2032 lithium battery
128 KB (84,650 measurements, maximum)
USB 2.0 interface
20 seconds
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly
or go blank in temperatures outside this range
3.66 x 5.94 x 1.52 cm (1.44 x 2.34 x 0.6 in.)
23 g (0.81 oz)
IP50
The CE Marking identifies this product as complying with all relevant
directives in the European Union (EU).
Plot B: Time Accuracy

Advertisement

loading

Summary of Contents for Onset Computer Corporation HOBO UX100-001

  • Page 1 HOBO® Temp Data Logger (UX100-001) Manual The HOBO Temp data logger records temperature in indoor environments with its integrated sensor. Using HOBOware®, you can easily configure the logger alarm to trip for specific high or low temperatures. Or, you can set up burst logging in which the logger records data at a different interval during certain conditions.
  • Page 2 HOBO Temp Data Logger (UX100-001) Manual Logger Components and Operation LCD Symbol Description The logger has been configured to stop logging Battery Door when memory fills. The memory bar indicates the Alarm/Stats Button approximate space remaining in the logger to Start/Stop Button record data.
  • Page 3 HOBO Temp Data Logger (UX100-001) Manual Notes: above or below a value you specify. See Setting up Alarms for details. • You can disable the LCD screen when logging. Select 3. Configure filters (optional). Click the Filters button to “Turn LCD off” when setting up the logger as described in create additional filtered data series.
  • Page 4: Setting Up Alarms

    HOBO Temp Data Logger (UX100-001) Manual materials (see Mounting the Logger). After logging begins, “Sensor reading within limits” if you want the alarm to you can read out the logger at any time (see Reading Out clear once the sensor reading returns to the normal the Logger for details).
  • Page 5 HOBO Temp Data Logger (UX100-001) Manual the high level value. Type in the value or drag the red data point at 9:05. Then, the high level was surpassed upper slider. and burst logging began at 9:06. Burst logging then continued until 9:12 when the sensor reading fell back below the high level.
  • Page 6: Battery Information

    HOBO Temp Data Logger (UX100-001) Manual 3. Set the sampling interval, which must be less than and a Mounting the Logger factor of the logging interval. Choose either a preset There are several ways to mount the logger using the materials sampling interval or select Custom and enter your own included: sampling interval.
  • Page 7 Even with the channel disabled, a bad battery event will still be recorded. © 2013 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOware are 1-800-LOGGERS (564-4377) • 508-759-9500 trademarks or registered trademarks of Onset Computer Corporation. All other trademarks are the www.onsetcomp.com •...