SM302 InGaAs Array Spectrometer User Manual 1105811 August 2006 Spectral Products 2659-A Pan American Frwy., NE Albuquerque, NM 87107 Tel (505) 343-9700 Fax (505) 343-9705 www.spectralproducts.com...
Failures or damages resulting from lack of operator attention to proper procedures, failure to follow operating instructions, unauthorized modifications, and natural disasters are not covered under this warranty. The SM302 does not contain any user serviceable parts. Removing its cover, without explicit written permission from Spectral Products , will void any written or implicit warranty.
1.4. Product Overview The SM302 is a versatile, high performance and complete InGaAs array spectrometer that consists of a low noise 256- element InGaAs array detection module, a high precision computer controlled tunable spectrograph optimized for the spectral range of 0.9 to 1.7 µm (.09 to 2.6µm extended), a 16-bit data acquisition board, and Windowsbased operating software.
• InGaAs power supply: *Input voltage = 100-120V AC with DB15 connector. • SM302 spectrograph power supply: UL listed 110/220 V power pack, meets or exceeds UL1950, CSA 1402C, and IEC 950 with 9 pin din connector. • Stray light rejection: better then 10 •...
Getting Started. 2.1. Checking System Requirements Check to make sure that your computer meets the minimum requirements for the SM302 InGaAs system. 2.1.2. Requirements for the Hardware • 1 serial port for spectrograph communication • 1 PCI slot for the data acquisition card •...
HardWare SetUp 2.2.1 Installing the SM302 PC1605E card One of the routine precautions you must be aware of when working with computer components is the problem of static electric discharge Note: Leave the PC16052E card in its static-resistant bag until you are ready to install it.
Spectral Products offers a DK12AT, DK12PS and DK12MA cable for connecting to AT, PS2 and MAC style computers, respectively. Pin Assignments for the Female DB-9 Connector at Rear of SM302 Spectrograph (See Figure 1 and 2) NAME FUNCTION...
Using the LPF 1300 Filter On the SM302 Extended models it might be necessary to use the LPF 1300 filter. Mounted on the side of the unit is a black knob with an In and Out position. The LPF 1300 will filter out any first or second orders below 1300nm (IN position).
Slit changes The standard slit size of the SM302 is 50µ. The slit is located in the back of the housing of the SMA flange on the SM302 entrance. If a different slit size is preferred you must order the new slit by contacting Spectral Products Spectral Products Group at (505) 296-9541.
SM302 Power Supply Jumper Configuration Figure 5 Hardware Jumper setting for SM302 Power Supply To set power supply to run the SM302 standard, jumper to R12 is set for 5V (JP1). To set the SM302 Extended, jumper R4 is set for 6V (JP2).
Spectrometer Power Supply Pin Number Dual Common Common +5.0V Imin 0.5A Imax 7.0A Common +15V Imin 0.1A Imax 3.0A Grating Selection - 12 - Spectral Products 2659-A Pan American Frwy., NE Albuquerque, NM 87107 Tel (505) 343-9700 Fax (505) 343-9705 www.spectralproducts.com...