scigiene BAS Series User Manual

scigiene BAS Series User Manual

Balance analytical scale

Advertisement

Quick Links

Balance Analytical Scale
User Manual
1

Advertisement

Table of Contents
loading

Summary of Contents for scigiene BAS Series

  • Page 1 Balance Analytical Scale User Manual...
  • Page 2: Table Of Contents

    Index ABOUT THE SCALE ..........................................I.INSTALLATION ............................................1.1 Unboxing............................................1.2 Work environment ..................................4 1.3 Installation......................................4 1.4 Button and Display..................................5 II.OPERATION .............................................. 2.1 Calibration ............................................2.1.01Calibration weight..................................6 2.1.02 Calibration Step..................................6 2.2 Function setting ..........................................2.2.01Unit conversion ........................................... 2.2.02 Speed adjustment ........................................
  • Page 3: About The Scale

    OVERVIEW This manual is used for BAS series multifunctional analytical balance. BAS series multifunctional electronic balance is a new generation product newly developed. Using a high-precision electromagnetic force balance sensor. It makes measurement results more accurate, faster response time, and fewer failures. BAS series multifunctional electronic balances have: •...
  • Page 4: I.installation

    I. INSTALLATION 1.1. Unboxing Please carefully open the packing box, gently take out the balance and the contents in the box, and save the packing materials for reuse. Box contents: (1) BALANCE (2) PAN (3) ADAPTER (4) MANUAL (5) CERTIFICATE OF CONFORMITY (6) WEIGHT 1 2 3 5 7 6...
  • Page 5: Button And Display

    1.4. Buttons and display The display and keyboard are like icons. ①POWER ② SET ③TARE(→0/T←),CONFRIM ④CAL ⑤UNIT ⑥ + ⑦ - ⑧Stability indicator ⑨Range progress bar ⑩Weighing ⑪Speed ⑫Stability ⑬UNIT ⑭Animal weighing...
  • Page 6: Ii.operation

    200g 2.1.02 CAL.STEP For BAS series balances, please warm up for 1 hour and wait until the balance is stable before calibrating. Take BAS-2004 as an example, weighing no-load, 1) Press →0← to tare first, the balance will display 0.0000g, 2) Then press the CAL button, the display shows CAL-200 flashing, 3) At this time, put a "200g"...
  • Page 7: Stability Adjustment

    2. Press “UNIT” , lowest speed, press “O/T” to confirm. highest speed, decreasing in order, 2.2.03 SENSITIVITY SETTING 1. Press “SET” until it shows“ STAB”, press “O/T” enter STAB setting. 2. Press “UNIT” to choose sensitivity, High Sensitivity, press “O/T” to confirm. Poorest Sensitivity, 2.2.04 COUNTING There are counting functions in the balance, and you can use this for count the quantity.
  • Page 8: Percentage Function

    2.Press “O/T” to confirm. 2.2.06 PERCENTAGE FUNCTION Press “SET” enter menu, then press “SET” until it shows “PCT”; Press “O/T” enter percentage function, it shows“ PCT 100”; Put 100% sample, press “O/T” enter percentage weighing. 2.2.07 PEAK HOLD The steps are as below: 1.
  • Page 9: Iii.data Output

    III.DATA OUTPUT 3.1 SERIAL PORT CONNECTION BALANCE 2…………………………………… 3 3…………………………………… 2 5…………………………………… 5 3.2 The steps of connecting to computers: 1. Click the “start menu “ Program Attachments Communication HyperTerminal 2. Enter a name: ICON SELECTION, Click OK 3. Ignore the previous column ,Select "connection with a (N) : com 1" ,Click OK: (1) Baud rate :1200 (2) Data bits :8 (3) Parity: No...
  • Page 10: Iv.fault Judgment

    IV. FAULT JUDGMENT FAULT REASON EXCLUDE No display Not connected to the power Plug in the power line; supply; Replace the fuse; Fuse is broken; Replacement of power Power transformer damage; transformer; Contact factory Weighing unstable 1. Bad working environment; Improve the working environment and avoid the interference of 2.
  • Page 11: V.maintenance

    V. MAINTENANCE The BAS series multifunctional electronic balance is a precision mechatronics intelligent measuring instrument, so it must be treated as seriously and carefully maintained as other precision instruments. 1. Do not use sharp objects (such as pencils, ball-point pens) to press keys, only use your fingers to press keys;...