OVERVIEW This manual is used for BAS series multifunctional analytical balance. BAS series multifunctional electronic balance is a new generation product newly developed. Using a high-precision electromagnetic force balance sensor. It makes measurement results more accurate, faster response time, and fewer failures. BAS series multifunctional electronic balances have: •...
I. INSTALLATION 1.1. Unboxing Please carefully open the packing box, gently take out the balance and the contents in the box, and save the packing materials for reuse. Box contents: (1) BALANCE (2) PAN (3) ADAPTER (4) MANUAL (5) CERTIFICATE OF CONFORMITY (6) WEIGHT 1 2 3 5 7 6...
1.4. Buttons and display The display and keyboard are like icons. ①POWER ② SET ③TARE(→0/T←),CONFRIM ④CAL ⑤UNIT ⑥ + ⑦ - ⑧Stability indicator ⑨Range progress bar ⑩Weighing ⑪Speed ⑫Stability ⑬UNIT ⑭Animal weighing...
200g 2.1.02 CAL.STEP For BAS series balances, please warm up for 1 hour and wait until the balance is stable before calibrating. Take BAS-2004 as an example, weighing no-load, 1) Press →0← to tare first, the balance will display 0.0000g, 2) Then press the CAL button, the display shows CAL-200 flashing, 3) At this time, put a "200g"...
2. Press “UNIT” , lowest speed, press “O/T” to confirm. highest speed, decreasing in order, 2.2.03 SENSITIVITY SETTING 1. Press “SET” until it shows“ STAB”, press “O/T” enter STAB setting. 2. Press “UNIT” to choose sensitivity, High Sensitivity, press “O/T” to confirm. Poorest Sensitivity, 2.2.04 COUNTING There are counting functions in the balance, and you can use this for count the quantity.
2.Press “O/T” to confirm. 2.2.06 PERCENTAGE FUNCTION Press “SET” enter menu, then press “SET” until it shows “PCT”; Press “O/T” enter percentage function, it shows“ PCT 100”; Put 100% sample, press “O/T” enter percentage weighing. 2.2.07 PEAK HOLD The steps are as below: 1.
III.DATA OUTPUT 3.1 SERIAL PORT CONNECTION BALANCE 2…………………………………… 3 3…………………………………… 2 5…………………………………… 5 3.2 The steps of connecting to computers: 1. Click the “start menu “ Program Attachments Communication HyperTerminal 2. Enter a name: ICON SELECTION, Click OK 3. Ignore the previous column ,Select "connection with a (N) : com 1" ,Click OK: (1) Baud rate :1200 (2) Data bits :8 (3) Parity: No...
IV. FAULT JUDGMENT FAULT REASON EXCLUDE No display Not connected to the power Plug in the power line; supply; Replace the fuse; Fuse is broken; Replacement of power Power transformer damage; transformer; Contact factory Weighing unstable 1. Bad working environment; Improve the working environment and avoid the interference of 2.
V. MAINTENANCE The BAS series multifunctional electronic balance is a precision mechatronics intelligent measuring instrument, so it must be treated as seriously and carefully maintained as other precision instruments. 1. Do not use sharp objects (such as pencils, ball-point pens) to press keys, only use your fingers to press keys;...