Special Precautions

Electrical Shock Hazard

©Ultrasound Technologies Ltd
You are then prompted to enter the Chart recorder speed.
This sets the initial turn-on speed of the chart recorder. Then once the chart is running
the speed can be further changed to 1 ,2 or 3 cm/min. It is set by pressing the Volume
Up button for a value of 1, 2 or 3 cm/min..
When the correct speed is displayed press the Toco Zero button once.
You are then prompted to select the Data Block on or off.
The Data Block is printed at the start of each recording, this function can be disabled
by setting Data Block to off. It changes from on to off and back again by pressing the
Volume Up button..
When the Data Block is set to your requirements press the Toco Zero button once.
You are then prompted to select the Tachycardia Alarm on or off.
The Tachycardia Alarm is triggered when the system detects a Tachycardia above 170
bpm and an alarm tone sounds. The alarm is tone is stopped by pressing the units
volume button. Tachycardia Alarm can be set to on to off and back again by pressing
the Volume Up button..
When the Tachycardia Alarm is set to your requirements press the Toco Zero button
once.
You are then prompted to select the Bardycardia Alarm on or off.
The Bradycardia Alarm is triggered when the system detects a Bradycardia below 100
bpm and an alarm tone sounds. The alarm is tone is stopped by pressing the units
volume button. Bradycardia Alarm can be set to on to off and back again by pressing
the Volume Up button..
When the Bradycardia Alarm is set to your requirements press the Toco Zero button
once.
The FETATRACK 310 will then restart and operate according to your settings.
The FETATRACK 310 is now operational, and the LCD screen will show US1 --- us2
---
(if the unit is only a single fetus monitor it will only display US1 ---) and a Toco
value.
Your FETATRACK 310 Antenatal Cardiotocograph has been designed for electrical
safety. All the safety and operating instructions should be read before operating the
FETATRACK 310. Failure to do so could result in injury to the user, patient or
damage to the System and accessories.
Do not defeat the grounding integrity of this System. Protection against electrical
shock, in the event of failure of basic insulation, is provided by the connection of the
chassis to the safety ground. Safety grounding occurs only when the 3-wire cable and
plug provided with the system are connected to a properly grounded receptacle.
Do not remove the System cover. The System should be serviced by trained and
qualified personnel only. Contacting the hazardous voltages within the System could
cause serious injury.
Issue 1 December 2004
Ultrasound Technologies Ltd
Page 7