Summary of Contents for Vega Technologies CP-03
Page 1
VEGA INSTRUCTION MANUAL Continuous Positive Airway Pressure MODEL: CP-03 ISO9001 & ISO13485 Certified This device complies with EU EMC standard EN60601-1-2 READ THIS INSTRUCTION Version: 1.0 Date: Nov 04, 2009 R404-1...
Page 2
I. Introduction The product name is CPAP( ). The device uses a Continuous Positive Airway Pressure motor which drives a fan to create a continuous air pressure to force the air into the patient's lung through the nasal cavity, so that the patient's airway is expedite, which is the key to OSA (Obstructive Sleep Apnea Syndrome) therapy.
Page 4
III. Important Safeguards Note: Read all instructions carefully before use. The following basic precautions are needed when using an electrical product: Caution: Failure to read and observe all precautions could result in personal injury or equipment damage. Product cautions: 1. To avoid electrical shock: Keep the unit away from water‧Do not immerse the power cord or the unit in liquid‧Do not use while bathing‧Do not reach for a unit that has fallen into water –...
Page 5
IV Device Specifications Value Description Power Supply 12V DC Current Consuming Max 1.5A Air Pressure Range 3 cmH O – 20 cmH Ramp Time 0 – 60 min ± 3 cmH Air Pressure Calibrate 3 ± 0.3m Working Flow /h @ 9 cmH <...
Page 6
2. Press “START/OFF” at the standby mode, then the CPAP runs with an output of pre-set air pressure, meanwhile, the LCD displays “Pressure xx.x cmH2o”(xx.x is the numeric value of the pre-set air pressure). This is the normal working mode of the device. 3.
Page 7
Mask off warning: In normal working mode or ramp mode, if the mask falls off the patient’s face, there is a “Mask off warning” on the screen: the LCD displays “Mask Off”, meanwhile, the backlight flashes at the frequency of 1 Hz for 1 minute. The following are configuration operations 5.
Page 8
displays “Pressure xx.x cmH2o”, xx.x is the numeric value of the air pressure you want to set to the device. Use “+”and “-”to change this value ——the value will increase/reduce 0.5 once with a single press and will take a fast change with a speed of 0.5 with continues press. Press “MODE”...
Page 9
9. Press “MODE” at the air pressure calibrate mode, the device enters the ramp time setting mode. The LCD displays “Delay Time xx min”, xx is ramp time. Use “+” or “-” to change this value. The ramp time adjusting range is 10 to 60.
Page 10
11. Press “MODE” at the starting air pressure setting mode, the device enters the compliance time checking mode. The LCD displays “Compliance Hrs xxxx.x”, xxxx.x is the compliance time of the device. This value is not to be changed, but only to be reset. Press “+” + “-” for a period of time of 2 seconds to reset it.
Page 11
16. Press “START/OFF” to leave configuration mode and return to standby mode. 17. The configuration can also be done through “CPAP” PC application. Connect the device and your PC with a USB cable, then run “CPAP” software on your PC, this software allows you to do most of the configurations on PC.