Technical Specifications; Operating Conditions; Electrical Details; Performance - Grant ecocool 100R Operating Manual

Combined refrigerated and heating bath circulator
Table of Contents

Advertisement

6.0

Technical specifications

6.1

Operating conditions

Ambient temperature range
Altitude above sea level
Operating environment
Maximum relative humidity
6.2

Electrical details

Mains supply:
Pollution degree:
Installation category:
*220-240V heating circulators can only operate at a single mains frequency. Check the
rating plate of the heating circulator.
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3

Performance

Specification
Typical cooling power at an
ambient temperature of 20°C
Stability (DIN 12876) @ -10°C
Settable temperature range
Tank capacity
Top opening
Tank depth
Liquid depth min/max
Weight
Max pump head pressure
Max pump flow rate
Safety high temperature limit
Max current consumption (both equipment
excluding external pump)
Heater power
*
Combined rating of both the refrigeration bath and heating circulator.
6.4

Storage and transportation

Store and transport in original packaging. Temperature range -20°C to +60°C
ecocool 100R
Operating Manual
5 to 40°C
Up to 2,000m (6,500ft)
Indoor use only
80% RH up to 31°C decreasing to 50% RH at 40°C
220-240V @ 50 or 60Hz* or 110-120V @ 50/60Hz
2
II
@ 20°C
@ 0°C
@ -10°C
@ -20°C
www.grantinstruments.com
ecocool 100R
ecocool 100R AUS
250W
200W
100W
50W
±0.05°C
-20°C to 100°C
5L
120 x 155mm
150mm (usable 140 leaving 10mm clear)
85/145mm
29kg
250mBar
17L/min
Adjustable cut-out above 20°C
7A
1.3kW
ecocool 100R US
18A*
1.4kW
Page 22