6.0
Technical specifications
6.1
Operating conditions
Ambient temperature range
Altitude above sea level
Operating environment
Maximum relative humidity
6.2
Electrical details
Mains supply:
Pollution degree:
Installation category:
*220-240V heating circulators can only operate at a single mains frequency. Check the
rating plate of the heating circulator.
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3
Performance
Specification
Typical cooling power at an
ambient temperature of 20°C
Stability (DIN 12876) @ -10°C
Settable temperature range
Tank capacity
Top opening
Tank depth
Liquid depth min/max
Weight
Max pump head pressure
Max pump flow rate
Safety high temperature limit
Max current consumption (both equipment
excluding external pump)
Heater power
*
Combined rating of both the refrigeration bath and heating circulator.
6.4
Storage and transportation
Store and transport in original packaging. Temperature range -20°C to +60°C
ecocool 100R
Operating Manual
5 to 40°C
Up to 2,000m (6,500ft)
Indoor use only
80% RH up to 31°C decreasing to 50% RH at 40°C
220-240V @ 50 or 60Hz* or 110-120V @ 50/60Hz
2
II
@ 20°C
@ 0°C
@ -10°C
@ -20°C
www.grantinstruments.com
ecocool 100R
ecocool 100R AUS
250W
200W
100W
50W
±0.05°C
-20°C to 100°C
5L
120 x 155mm
150mm (usable 140 leaving 10mm clear)
85/145mm
29kg
250mBar
17L/min
Adjustable cut-out above 20°C
7A
1.3kW
ecocool 100R US
18A*
1.4kW
Page 22
Need help?
Do you have a question about the ecocool 100R and is the answer not in the manual?