NurseCall N86 Room Transmitter Table of Contents | en 3 Table of Contents Identification .................7 Document ......................7 Customer support addresses ............... 8 Generalities ................... 9 NurseCall System ..................9 NurseCall N86 Wall Transmitter ............9 Safety Instructions ..............10 Introduction ....................10 3.1.1 Principle ..........................10 3.1.2...
Page 4
4 en | Table of Contents NurseCall N86 Room Transmitter Options ......................16 4.3.1 Relay contacts for external signals ...............16 4.3.2 Pull Cord Set ........................16 4.3.3 Mounting Plate / Bracket ................... 17 Installation ...................18 Unpacking .....................18 5.1.1 List of contents ........................18 Installation ....................19 5.2.1 Generalities ........................19...
Page 5
NurseCall N86 Room Transmitter Table of Contents | en 5 Daily Message and Battery Low Message ........31 Supervision ....................31 6.10 Bi-directional radio traffic ..............31 6.11 Error Message .....................31 Operation ..................32 Description of the eight different states ........32 Legend of illustrations ................33 7.2.1 LED and relays behaviour ..................33 7.2.2...
8 en | Identification NurseCall N86 Room Transmitter Customer support addresses TeleAlarm SA Rue du Pont 23 2300 La Chaux-de-Fonds Switzerland info@telealarm.com TeleAlarm SA Hinterbüelstrasse 3 8307 Effretikon Switzerland info-ch@telealarm.com TeleAlarm SA Rue du Pont 23 2300 La Chaux-de-Fonds Switzerland www.telealarm.com Copyright TeleAlarm SA, 2020 TeleAlarm Europe GmbH...
NurseCall N86 Room Transmitter Generalities | en 9 Generalities NurseCall System Alarms and Messages arriving from NurseCall Transmitters such as the NurseCall N86 Wall Transmitter are managed and stored by the NurseCall Main Unit (or external third party system using AP80 devices). NOTICE! The document "NurseCall General Overview"...
10 en | Safety Instructions NurseCall N86 Room Transmitter Safety Instructions WARNING! The User / Installer should read and understand this chapter before any intervention on the NurseCall N86 Room Transmitter. Introduction 3.1.1 Principle NOTICE! In case of unclear information, please contact your local representative. 3.1.2 Importance of safety instructions Each safety and protection instruction in this manual must be adhered to in...
NurseCall N86 Room Transmitter Safety Instructions | en 11 CAUTION! The NurseCall N86 Room Transmitter may not be used under exposure to the direct sunlight, to heat, to dust or to an excessive humidity (only use the equipment in a clean environment). >...
12 en | Safety Instructions NurseCall N86 Room Transmitter WARNING! Electro Static Discharge The NurseCall N86 Room Transmitter contains highly sensitive electronic components. It should be opened only in an ESD protected environment with respect to the following precautions: > Discharge yourself from electrostatic loads by touching a grounded conductive surface before opening the unit.
NurseCall N86 Room Transmitter Description | en 13 Description General description 4.1.1 Top view Fig. 4-1:Top view Red call for help button 2. LED indicator for alarm 3. LED indicator for presence 4. Green clear alarm button, to clear an alarm locally 5.
14 en | Description NurseCall N86 Room Transmitter Main functions The main function of the NurseCall N86 Room Transmitter is to enable easier handling of emergencies and help, directly from the patient’s room. 4.2.1 Red call button Its large red button with textured surface makes it easy to identify and to press, launching an help call.
NurseCall N86 Room Transmitter Description | en 15 4.2.5 Clear Alarm with group identification (Multiple Staff mode) This feature "Multiple Staff mode" must be confirmed in the N86 program parameter! When automatic call repetition is enabled, a call must be cleared by a B81 acknowledgment transmitter to stop the alarm repetition and to put the N86 in presence mode.
16 en | Description NurseCall N86 Room Transmitter N86 After the Alarm is cleared the N86 will be in this case directly set back in Stand-by mode by pressing the B81 very close to the N86 Technical alarms like button blocked or Pear push disconnected can only be cleared by pressing the B81 once very close to the N86, or twice when the N86 is in presence mode The green N86 button can only be used to stop the blinking of the white LED, it...
NurseCall N86 Room Transmitter Description | en 17 4.3.3 Mounting Plate / Bracket The mounting plate is made for easy mounting preparation as well to give a certain mobility to the NurseCall N86 Room Transmitter. The mounting plate works as a bracket. To fix the NurseCall N86 Room Transmitter it must be attached to the wall bracket and then pulled down slightly.
18 en | Installation NurseCall N86 Room Transmitter Installation Unpacking The NurseCall N86 Room Transmitter is carefully packed for transportation. The components contained in the box are protected, but should be handled with care. Store the packaging material for further use (storage or transport). NOTICE! In case of defective or missing equipment, do not try to install the NurseCall N86 Room Transmitter.
NurseCall N86 Room Transmitter Installation | en 19 Installation 5.2.1 Generalities > Install the NurseCall N86 Room Transmitter in a dry place and away from any source of heat. CAUTION! Interferences Avoid immediate proximity to other electric devices such as a television. 5.2.2 Wall installation The NurseCall N86 Room Transmitter can be fixed with two or up to five screws...
NurseCall N86 Room Transmitter Installation | en 21 5.2.3 Connecting to the power supply The NurseCall N86 Room Transmitter can be powered by a lithium battery or/ and an external 6-30 VDC power supply. If the device is powered with an external power supply, the battery is not necessary, but it can be used as a backup power supply in case of power loss.
22 en | Installation NurseCall N86 Room Transmitter 5.2.6 Transmitter Identification All NurseCall N86 Room Transmitters are provided with their own unique radio identification code (UID) that is assigned in the factory. For the use with a NurseCall Main Unit only the last three cyphers of the unique serial number is the radio identification code of the transmitter.
NurseCall N86 Room Transmitter Installation | en 23 CAUTION! There is a risk of explosion if battery is replaced by a wrong type or if not connected correctly. The battery should be replaced exclusively by Authorized Personnel. Dispose of used batteries according to instructions and regulations. Battery type is 3V CR2.
24 en | Installation NurseCall N86 Room Transmitter Inputs and Outputs 5.3.1 General inputs The * marks a factory setting. Closing contact causes a Pear Jack button alarm or any other socket 3.5 External trigger programmed call type mm and / Input Pear Input Pear Opening contact causes an any...
NurseCall N86 Room Transmitter Installation | en 25 NO (Normally Open): Trigger selected when closed to GND NC (Normally Close): Trigger selected when open from GND CAUTION! Relays contacts support 0.3A @30 VDC or 15 VAC (RMS) The sum of the input voltage and the voltage at a relay contact and the sum of both relay contact voltages shall not exceed the limits of ES1 (60 VDC or 42 Vpeak) defined in IEC 62368-1:2018 standard §6.2.2.5 5.3.4...
Page 26
26 en | Installation NurseCall N86 Room Transmitter The Assistance Call is also possible when the Care Sent Activated assistance if Personnel has not signalled his/her presence.. only The Assistance Call is possible only if the Care presence is Not activated* Personnel has signalled his/her presence.
NurseCall N86 Room Transmitter Use | en 27 Every time a NurseCall N86 Room Transmitter is used by pressing a button or activating a wired input, the device sends a radio signal and the lamps indicates its action and / or actual state. Optionally, two relays allow the transmission of status to a remote display such as the signal lamps placed on top of the door.
28 en | Use NurseCall N86 Room Transmitter activated. A Call for Assistance is signalled by red double flashes on the signal lamp and by the alternate action/release of the relay Call for Help. Call for Emergency The Call for Emergency is triggered by pushing the Call for Help (red) and the Call for Assistance button (blue) simultaneously.
NurseCall N86 Room Transmitter Use | en 29 Also, when only presence mode is enabled and repeat mode disabled, a B81/ B80A or the N86 green button can be used to clear the call and to put the N86 in presence mode. After the alarm is cleared the N86 can be set in stand-by mode by pressing the N86 green button once or by pressing the B81/B80A once very close to the N86 Technical alarms like button blocked or Pear push disconnected can be cleared...
30 en | Use NurseCall N86 Room Transmitter 6.4.3 Acknowledgment with Nurse identification (Badge only mode) This feature "Badge only mode" must be confirmed in the N86 program parameter! When automatic call repetition and presence is enabled, a call must be cleared B81/B80A acknowledgement transmitter close to the N86, to stop the alarm repetition and to put the N86 in presence mode.
NurseCall N86 Room Transmitter Use | en 31 Caregiver leaving the room The Caregiver "away" message is triggered when the Caregiver triggers the N86 green button or the B81/B80A for Mode 1, when leaving the room. Pressing the green button by leaving the room is signalled by switching off the green flashes on the signal lamp and by releasing the Staff Presence relay.
32 en | Operation NurseCall N86 Room Transmitter Operation Description of the eight different states During operation, the NurseCall N86 Room Transmitter can be set to one of eight following states: State Description Stand-by No call is being processed. Help Call without staff The patient has triggered a Call for Help but presence the Care Personnel has not yet responded.
NurseCall N86 Room Transmitter Operation | en 33 Legend of illustrations 7.2.1 LED and relays behaviour Legend Description The contact of the Call for Help relay is closed LED RED / REL1 ON The contact of the Call for Help relay is open ...
NurseCall N86 Room Transmitter Storage | en 47 Storage One can distinguish between long term storage and short term storage. If the NurseCall N86 Room Transmitter must be temporarily stored, it may not be necessary to wrap it completely. However, if you wish to store the NurseCall N86 Room Transmitter for a longer time, for instance in a storage room or any similar location, it is recommended to use original packing material.
en | Disposal NurseCall N46 Wall Transmitter Disposal This chapter describes the appropriate disposal of the NurseCall N86 Room Transmitter. The NurseCall N86 Room Transmitter is marked with a crossed-out wastebasket symbol. This means that, at the end of its useful lifespan, the product shall be disposed separately from ordinary household wastes in accordance to the EU Directive 2012/19/UE.
NurseCall N46 Wall Transmitter Disposal | en 49 9.4.1 Battery NOTICE! The battery should never be placed in municipal waste. Use a battery disposal facility if available. Fig. 9-1:Crossed-out wastebasket symbol > Please check local regulations for disposal of batteries or call your local representative for information.
Page 50
50 en | Appendix NurseCall N86 Room Transmitter Appendix 10.1 Electrical specifications External power supply 6 - 30 VDC & consumption Transmission: < 25 mA Idle mode: < 10 mA Battery power supply 3V lithium battery type CR2 & consumption Transmission: <...