SCI 440 Series Quick Start Manual

Hipot, insulation resistance, and ground bond testers

Advertisement

Quick Links

HIPOT, INSULATION RESISTANCE, AND GROUND BOND TESTERS
SAFETY CHECKLIST
[S]
urvey the test station. Make sure it is safe & orderly.
[A]
lways keep unqualified/unauthorized personnel away from the test area.
[F]
amiliarize yourself with safety protocols in the event of a problem.
[E ]
xercise caution and never touch products or connections during a test.
[ T]
rain operators. Connect the return lead first and never touch clips directly.
[ Y]
ou should always know when a test is being performed.
WARNING:
THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH ELECTRICAL SAFETY TESTING.
AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC
SHOCK. TO PREVENT ACCIDENTAL INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN
HANDLING AND USING A TEST INSTRUMENT.
440 Series
M o dels: 446 / 44 8
QUICK START GUIDE

Advertisement

loading

Summary of Contents for SCI 440 Series

  • Page 1 QUICK START GUIDE 440 Series HIPOT, INSULATION RESISTANCE, AND GROUND BOND TESTERS M o dels: 446 / 44 8 SAFETY CHECKLIST urvey the test station. Make sure it is safe & orderly. lways keep unqualified/unauthorized personnel away from the test area.
  • Page 2 Check to be sure the correct input line cord is used. Connect the power input plug into its socket on the rear panel of the tester. The SCI 440 Series Testers have an automatic input voltage range selection. Connect the male end of the plug to the grounded Grounded AC outlet.
  • Page 3 SETTING TEST MEMORIES The SCI 440 series testers are equipped with 20 memory programs numbered 1 through 20. Each memory can be connected sequentially to the next consecutive memory. Only one test type can be selected for each memory location. However, all 20 memory locations can be programmed and the parameters will be saved in each memory.
  • Page 4 GO BACK TO MAIN MENU Turn the yellow knob to scroll to When BACK is selected, press the Main Menu. BACK in order to return to the knob. You will be returned to the Main Menu. Main Menu. EDIT TEST PARAMETERS When VOLT is selected, press the Turn the yellow rotary knob to Turn the rotary knob to change...
  • Page 5: Test Connection

    TEST CONNECTION Using the Test leads Connect the black return lead (99-10865-01) to the front panel return output terminal and connect the other end of the lead to the dead metal on the chassis of the DUT. If you are performing a Ground Bond test, connect one end of the high current lead (99-10866-01) to the front panel current output terminal and the other end to the ground pin of the DUT line cord.
  • Page 6: Test Results

    Connect the red current lead of the adapter box to the front panel current output terminal. Connect the high voltage lead of the adapter box to the front panel H.V output terminal. Connect the black return lead (99-10866-01) to the front panel return terminal and connect the other end of the lead to the dead metal on the chassis of the DUT.
  • Page 7 Electrical Safety Compliance Simplified. ™ 28105 N. Keith Drive, Lake Forest, IL 60045 USA +1-847-932-3662 • 800-504-0055 • Fax +1-847-932-3665 • hipot.com • info@hipot.com © 2022 SCI V1.0 | 5/22...