Page 3
!!!DANGER!!! DO NOT USE THIS EQUIPMENT TO PURGE TOXIC OR FLAMMABLE GAS DO NOT USE THIS EQUIPMENT UNDER FLAMMABLE, VOLATILE OR TOXIC ENVIRONMENTAL CONDITIONS...
Additional Nosepiece Adapters are available to provide capabilities of numerous diameter Light Guides. The OPTI-LITE light box also has a high output fan, lamp baffle, and unique Nosepiece design to provide maximum airflow while restricting stray light from heating outer surfaces.
Failure to do so could result in errant operation of the nearby device. D. Avoid using or storing the OPTI-LITE in, or on, areas of excessive vibration. Vibration of the lamp could cause shortening of the life of the Lamp.
Page 7
Route cord so that it will not be pinched, severed or walked upon. K. In the event of loss of power, the OPTI-LITE will remain off when power is restored to the unit. When power is returned, turn the Intensity Knob to OFF and then slowly turn up intensity until light is returned.
OPTI-LITE. 3.1 Receiving and Placement A. Select a space for the assembly and operation of the OPTI-LITE. The location of the machine should allow adequate clearance for ventilation (see Table 3-1 below, Minimum Clearance). Make certain that the mounting feet are on the bottom of the illuminator box.
Page 9
This page intentionally left blank. Section 3 – 2 21-11-1180...
This will allow the Lamp to warm-up, helping to extend the Lamp life. After using the OPTI-LITE, return Intensity Control Knob to low and allow a cool down time of one to two minutes before turning the unit off.
Page 11
This page intentionally left blank. Section 4 - 2 21-11-1180...
5.0 MAINTENANCE 5.1 General A. Other than periodical dusting, and wipe down, the OPTI-LITE and Light Guide requires no other maintenance. B. When cleaning is needed, use a commercial glass cleaner or common household detergent to gently wipe down the exterior of the Illuminator Box and Light Guide.
5.0 MAINTENANCE, continued 5.2 Lamp Replacement, continued Figure 5-1, Lamp Replacement D. Gently pull the Lamp from the ceramic Lamp Socket. See Figure 5-1, Lamp Replacement. Note: Remember to never touch the Bulb or the Reflector in the Lamp. Only touch the sides when inserting a new Lamp! E.
5.4 Nosepiece Adapter Insert A. If use of a different diameter Light Guide is needed a Nosepiece Adapter Insert is necessary. To purchase additional Adapter Inserts, contact Galiso, Inc at (800) 854-3789 or visit www.galiso.com. B. To install the new Nosepiece Adapter, unscrew all three Thumb Screws. Turn the Illuminator Box over to allow the Adapter to fall out.
Page 15
3. Check the fuse. H. The Light Guide is burning 1. Check the fiber type. If a Light Guide is in use that is not a Galiso, Inc. fiber optic Light Guide, it may be plastic and susceptible to burning. Using a glass fiber Galiso, Inc.
Page 16
5.0 MAINTENANCE, continued 5.6 Spare Parts Item Part No. Description Lamp 85-11-0028 Halogen, MR16 Quartz 85-11-0029 Light Guide, 1/4" x 20'L 85-11-0043 Light Guide, 1/4" x 5' ULG 85-11-0050 Light Guide, 1/4" x 6' L 85-11-0061 Light Guide, 1/8" x 5' L 85-11-0062 Light Guide, 1/8"...
Page 17
This page intentionally left blank. Section 5 – 6 21-11-1180...
Page 18
Galiso. In the event of replacements, the replacement unit will continue under the original equipment warranty or carry a 90-day warranty, whichever is longer. No charge will be made for warranty repairs, and/or replacements.
Need help?
Do you have a question about the OPTI-LITE and is the answer not in the manual?
Questions and answers