Measurement Commands:
XRSE
Description:
Command:
Responses:
XRTi
Description:
Command:
Responses:
XRTe
Description:
Command:
Responses:
Confidential
Reads the current raw magnitude of the RST Sensor.
XRSE<CR>
X<s16><CR>
Successfully returns the measurement data value
F<CR>
Failed execution
?<CR>
Syntax failure
Reads the current internal temperature of the MMXI controller. Value is
in units of 1/16 degrees Celsius. Divide the value by 16 to get Celsius.
XRTi<CR>
X<s16><CR>
Successfully returns the temperature data
F<CR>
Failed execution
?<CR>
Syntax failure
Reads the current temperature of the RST Sensor. Value is in units of
1/16 degrees Celsius. Divide the value by 16 to get Celsius.
XRTe<CR>
X<s16><CR>
Successfully returns the temperature data
F<CR>
Failed execution
?<CR>
Syntax failure
www.multimetrixs.com