Guide for
Clinicians
TASKA HandGen1, Gen2 with
Coapt Gen1 or Gen2 System V1.1

Advertisement

loading

Summary of Contents for TASKA Coapt Gen1, Coapt Gen2

  • Page 1 Guide for Clinicians TASKA HandGen1, Gen2 with Coapt Gen1 or Gen2 System V1.1...
  • Page 2: Before You Get Started

    The TASKA Hand works seamlessly with the Coapt Gen1 or Gen2 systems. Please read the following to ensure a successful startup for the TASKA Hand: „ The Bluetooth on the TASKA Hand must be turned off at all times for correct operation of the Coapt system.
  • Page 3 Coapt system. Open Example of connection guide for Coapt ‘H2’ system using TASKA 6-band Coaxial Plug. Take care to connect the Coapt digital line (A) and the TASKA Power Switch cable (1) into the correct plugs to avoid damaging the hand.
  • Page 4: Normal Operation

    Coapt system. If Bluetooth is on during startup, press and hold the hand button until the blue light goes out, then power cycle the prosthesis. Bluetooth off: the Coapt system functions normally with the TASKA Hand, including driving grip changes.
  • Page 5 „ Select and calibrate grip patterns and other prosthesis’ motions „ Firmware and Software updates TASKA HandCal or MyTASKA HandCal or MyTASKA is only required to adjust the following settings: „ Set the hand into economy mode or disable low battery alarm.
  • Page 6 1. Ensure HandCal is installed and the Bluetooth dongle is connected to the PC.* 2. Hold down the hand button to turn Bluetooth on. 3. Start the HandCal software or MyTASKA app and connect to the TASKA Hand. 4. Adjust the required settings as per section 4. Configure**.
  • Page 7 6. Control Grip Cycle buttons The Coapt system changes the function of the button panel. Battery light EMG indicator light Light flashes when the Light flashes when Coapt battery is low. signals are being limited (Grip Triggers or All Sensors). Fault light Bluetooth light Light flashes if there is a...
  • Page 8 The hand will still open and close freely. 1. Press the EMG button. This will pause grip changes from Coapt. 2. On your TASKA Hand select the desired grip using the Hand/Home buttons. The hand will open and close as normal.
  • Page 9 All Sensors mode Change the EMG button setting in HandCal or MyTASKA to All Sensors to allow the hand to be locked and ignore all inputs. 1. Press the EMG button to ignore all inputs from the Coapt system. The light above the Hand button will flash and a beep may be heard.
  • Page 10: Troubleshoot

    A cable is loose or has a poor connection. Check all the connections and re-don the prosthesis if required. Contact your TASKA or Coapt representative for further troubleshooting assistance. For hand specific troubleshooting, refer to the Appropriate Use Guidelines supplied with the TASKA Hand.