LD 554 83 Instruction Sheet

Goniometer

Advertisement

Quick Links

11/01-W97-Sel
1

Description

The goniometer is a self-contained unit for mounting in the X-
ray apparatus and is included in the scope of supply of the X-
ray apparatus (554 811). It is equipped with two independently
controllable stepping motors which move the sensor and target
arms. The motion is defined using the keys in the control panel
of the X-ray apparatus and initiated manually with the ADJUST
knob or automatically with the SCAN key (see instruction sheet
of the X-ray apparatus).
The target stage can accommodate e.g. the LiF crystal for
Bragg reflection (554 77), the NaCl crystal for Bragg reflection
(554 78), the aluminum scattering body from the Compton ac-
cessory X-ray (554 836) or other samples.
The entire goniometer unit can be shifted within the experiment
chamber. The length of the sensor arm is variable, so that the
angular resolution can be altered for the slit of the sensor.
Note
The goniometer is positioned solely by means of electric
stepper motors:
• Do not block the target arm and sensor arm of the
goniometer and do not use force to move them.
Instruction sheet 554 83
Goniometer (554 83)
1 Target arm
with target holder (1a), target stage (1b
2 Sensor arm
with sensor holder (2a), sensor seat (2b)
3 Bottom guide groove
with knurled screws (3a)
4 Terminal pin connector
5 Top guide groove
2

Technical data

Working principle:
Angular resolution:
Angular range of target:
Angular range of sensor:
Sensor arm:
Length of sensor arm:
Width of sensor slit:
Target stage:
Sample clamping width:
Area of platform:
General data:
Dimensions:
Weight:
stepping motors for target and
sensor arms, which can be
electronically coupled
0.1°
unlimited
approx. -10° to 170°
approx. 40-110 mm
1 mm
3-9 mm
25 mm x 28 mm
13.5 cm × 22.5 cm × 12.5 cm
3 kg

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 554 83 and is the answer not in the manual?

Questions and answers

Summary of Contents for LD 554 83

  • Page 1 11/01-W97-Sel Instruction sheet 554 83 Goniometer (554 83) 1 Target arm with target holder (1a), target stage (1b 2 Sensor arm with sensor holder (2a), sensor seat (2b) 3 Bottom guide groove with knurled screws (3a) 4 Terminal pin connector...
  • Page 2 Instruction sheet 554 83 page 2/4 Zero position of the measuring system Mounting and demounting Warning: Each time the zero position of the measuring system is changed the old values of the zero position are overwritten in the microprocessor memory by the new values...
  • Page 3 3/4 Instruction sheet 554 83 Mounting/ demounting the sensor holder Inserting a sensor Mounting: - Slide the goniometer to the left in the experiment chamber as necessary. - Remove the target holder. - Loosen the knurled screw at the sensor seat.
  • Page 4 Instruction sheet 554 83 page 4/4 Mounting/ demounting the target holder Setting up a target - Position the target and sensor arms of the goniometer hori- Mounting: zontally using the ZERO key. - Mount the goniometer and position the target arm of the go-...

Table of Contents