CL96 Gas Detection System Technical Service Manual Revision 16.11 DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-ITD-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
4. Select the System tab 5. Press the DAC button for any active analyzer (see image below ) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
4. Select the System tab 5. Press the ADC button for any active analyzer (see image below) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
If there is an offset voltage outside of the suggested range remove the sample line and press Set Offsets. If the offset is still outside the suggested range refer to the CL96 Flow Troubleshooting Section. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
5. Locate the Point Arrangement Label on the top of the Cl96 enclosure near the inlet manifold. 6. Use the black knobs to adjust and balance inlet flow on each point. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Now locate the exhaust port for the related pump on the side of the CL96 (see image below) and plug it. When the exhaust port is plugged, all points should drop to the bottom of the scale in the yellow region. If the onscreen flowmeter does not drop to the bottom of the scale remove the CL96 rear panel and check for leaks between the exhaust port on the side of the cabinet and the exhaust port on the related pump.
3. Mark the tape after the optic blocks on the left hand side. (See picture below) 4. Advance the tape by pressing Open Gate then Close Gate DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
The CL96 comes with two 24VDC power supplies, one 12VDC power supply and one 5VDC power supply. Use the chart and images below to verify the voltages are within factory specifications. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 12
5VDC for the flow system can be measured at the red terminal block. The DC voltage verification should be done at the yellow terminal blocks mounted on DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 13
Because the pins are located next to each other this test should be done carfully to avoid a short and blowing a fuse DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
The terminal block jumpers are located in the control box and on the pump panel on all of the terminal blocks. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
3. Power cycle the CL96 DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
Page 16
2. If not put the CL96 back into analysis DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
Page 17
The PC cannot communicate with the 96144 Custom Output Module Failure Verification I/O modules Procedure. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
Page 18
None needed 96155 Restarting PLC Connection Restarting PLC Connection None needed DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
4. Select the System tab 5. Press the DAC button for any active analyzer ( see image below ) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 20
ChemLogic cassette out of the way. Next, remove the optic block cover by unscrewing out the two phillips screws (see image below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 21
140. Notice there are two fiber optic lenses per point. Both must be cleaned for the best possible results (see images below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 22
(see image below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
4. Select the System tab 5. Press the DAC button for any active analyzer ( see image below ) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
First, remove the cover for the optic block that is not working properly (see image below). When the cover is removed the power wires will be accessible. Using the DMM, measure the DC voltage between DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 25
Also verify none of the red lights are on, if one is on, this indicates a blown fuse and it must be replaced. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Verify that the orange and brown wires are tight and fully inserted into the optic block (see image below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 27
PC (see image below). Then confirm all optic block communication cables are securely connected to the interconnect board below the analyzer (see image below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Slave gate Slave switch Master gate Master switch DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
Page 29
If the issue cannot be found or resolve using this method please contact DOD Technologies Inc. technical additional support. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Check for any 0 values. This would indicate a communication issue between the flow system boards and the PC. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 31
5.3V at the red (+) and black (-) wires. Also, verify that all interconnect board connections are secured. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 32
0 values please contact DOD Technologies technical support. A17-32 A1-16 B17-32 B1-16 C1-16 C17-32 DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...
If flow is still low, locate the filter at the end of each sample line and verify that it’s not clogged. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
This procedure should be followed when any of the following messages appear in the CL96 Event Log: 1. Output Module Communication Error (Service Fault) 2. Output module failure or not installed (Service Fault) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 35
PC and modules by unplugging the green cable in the control box (see image below) and plugging it back DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 36
Each bank of modules will have (1) EK1101 module at the beginning with (3) address dials ( see image below). DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
To replace a CL96 optic block, follow these steps: 1. Open the gate 2. Remove power from the analyzer by unplugging the green connector. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 38
4. Remove the (2) screws holding down the optic block. 5. Remove all (8) tubes and pay attention to the orientation for reinsertion. (See image below) DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 39
8. Install the new block by following the steps in reverse order 9. Confirm the address on the new block matches the address on the old block. DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
8. Fully insert the inlet and exhaust tubing 9. Reconnect the (2) pin power connector. Inlet and exhaust tubing DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com...
Page 41
Page 38 of 44 Power Connector # 10 Nuts DOD Technologies, INC • 675 Industrial Dr. Bldg. A • Cary, IL 60013 • Phone (815) 788-5200 • Fax (815) 788-5300 DC-MAN-CL96Servicemanual www.dodtec.com Innovative Life Safety Systems & Services...