CO2 Meter TR250Z User Manual

CO2 Meter TR250Z User Manual

Oxygen sensor

Advertisement

Quick Links

 
CM‐0134  25% O2 Transmitter 
CM‐0134‐ WT 25% O2 Transmitter with Tube Sampling 
CM‐0150  25% O2 with Mount Shield 
CM‐0160  95% O2 Transmitter 
CM‐0160‐WT  95% O2 Transmitter with Tube Sampling 
CM‐0161  95% O2 with Mount Shield 
 
User Manual
TR250Z Oxygen Sensor
Copyright © 2015 CO2 Meter. All Rights Reserved.
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TR250Z and is the answer not in the manual?

Questions and answers

Summary of Contents for CO2 Meter TR250Z

  • Page 1 User Manual TR250Z Oxygen Sensor      CM‐0134  25% O2 Transmitter   CM‐0134‐ WT 25% O2 Transmitter with Tube Sampling   CM‐0150  25% O2 with Mount Shield   CM‐0160  95% O2 Transmitter   CM‐0160‐WT  95% O2 Transmitter with Tube Sampling   CM‐0161  95% O2 with Mount Shield    Copyright © 2015 CO2 Meter. All Rights Reserved.
  • Page 2: Table Of Contents

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Table of Contents   WELCOME  ..................... 3   IMPORTANT SAFEGUARDS ................ 3   DEVICE SPECIFICATIONS ................ 4   PACKAGE CONTENTS .................. 5    .................... 5 OFTWARE    ............... 5 INIMUM  YSTEM  EQUIREMENTS    ................... 6 UICK  TART  UIDE    ..................... 6 ESCRIPTION   STANDARD OPERATION  ................. 7  ...
  • Page 3: Welcome

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Welcome Thank you for purchasing the TR250Z Oxygen Sensor. The TR250Z is a long life 25% or 95%  oxygen level transmitter that provides multiple analog linear outputs of O2 concentrations. It’s  designed to be a rugged, maintenance‐free add‐on for OEMs or can be used as a desktop device.  In addition, it can be calibrated in normal air, or can use a calibration gas of any known oxygen  concentration.  Our goal is to make high‐quality, low‐cost gas detection and indoor air quality products  available to you.  Please take some time to read through this manual in order to get the most out of your meter.  Make sure to visit our website regularly to find more information about this product and to  download the software free. Once you download and install the software, you can use the  “Check for updates” function under the Help menu to search for new updates. Please also pay  special attention to the important safeguards shown on the pages ahead.  Important Safeguards To reduce the risk of fire, electrical shock and/or injury to persons, basic safety precautions should  always be followed when using electrical appliances, including the following:  1. READ ALL INSTRUCTIONS BEFORE USING THIS DEVICE!  2. Use only the supplied power supply to operate the unit.   3. Ensure that when sampling in a closed environment that the tubes are securely fashioned to the  device.  4. Do not operate with a blocked off sample path.  5. Do not operate the device if it is malfunctioning.  6. Do not install the device outside in exposed conditions. Due to the nature of the design the  enclosure cannot be made water‐tight and the unit must not be exposed to water.  7. Do not operate the device with the cover detached.   ...
  • Page 4: Device Specifications

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Device Specifications Measuring Range:  25% O2 0‐25% vol.  95% O2 0‐95% vol. Repeatability:  0‐25% CO2 ± 0.05%, ± 1 % of measured value  0‐95% O2 ± 0.02%, ± 1 % of measured value  Accuracy  25% / 95%  ±2% full scale  Power Supply:  Maximum Voltage 24VDC Minimum Voltage 20VDC Power Consumption 600mA @ 24VDC Sensor Ratings:  Life Expectancy >15 years Maintenance Interval No maintenance required Warm‐up Time <15 min (instant measurements)  Outputs:  O2 Output Value  4‐20 mA, linearly scaled ...
  • Page 5: Package Contents

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Package Contents  Please verify that your package contains the following items before using the unit:  Contents:   (1) Meter   (1) Certificate of Calibration   (1) User Manual   (1) Power supply   (1) 2 Meter USB to RS‐485 Cable   (1) Screen Breather      Software Visit CO2Meter.com and go to the “Downloads” page (see link below) to download our DAS or GasLab  software free or for more information on how to connect, configure, and calibrate your meter. For  general information about the program, please see the DAS or Gas‐Lab User Instruction Manual located  http://www.co2meter.com/pages/downloads in the website below.    IMPORTANT:      AKE SURE TO INSTALL SOFTWARE BEFORE CONNECTING                        YOUR DEVICE TO YOUR COMPUTER By installing our software first, you will ensure that the proper driver necessary for your unit is installed  on your computer before connecting your meter. Click on the “Install Now” link and then select “Run”  when prompted by your browser (as shown on below). ...
  • Page 6: Quick Start Guide

    TR250Z Oxygen Transmitter – Manual CO2Meter.com     Install DAS online (Internet Explorer 9 shown)  Quick Start Guide Make sure you read through these instructions thoroughly before using your meter. This guide will help  you familiarize with your meter in order to be as productive as you need to be in the least amount of  time possible.  Description Display – Main interface to view current sensor readings and menu state. Will blink “STRT” upon  applying power while the sensor warms up and stabilizes.   LEDs  – are in for lower right corner of the unit on the main PCB and are labeled S, E and P.  S LED – Status LED. Blinks every time a complete measurement cycle is performed and the value is  updated. Serve’s as a heartbeat for transmitter operation.   E LED – Error LED. Will light if any abnormal conditions are detected within the device, primarily the  sensor falling out of its recommended operating specifications will trigger this lamp.  P LED – Power LED. Will be on continuously when power is applied to the unit.   Buttons ‐ The unit has three surface mount buttons above the LEDs on the PCB. They can be gently and  carefully pressed with your finger. These buttons allow you to access all aspects of sensors operation  and test or calibrate different functional modules.  Rev. 27 April, 2015 Pg. 6 of 13...
  • Page 7: Standard Operation

    TR250Z Oxygen Transmitter – Manual CO2Meter.com   Standard Operation The O2 transmitters comes factory calibrated for easy installation. During its life cycle calibration may  become periodically necessary. We recommend performing a span O2 calibration every month for  guaranteed operation and ensured accuracy. No other calibration is necessary during the life cycle of the  sensor.   Entering the Service Menu To enter the service menu the buttons must be pressed in a factory‐set order. This code can be  customized before shipment but by default will be 1‐2‐3. The buttons must be pressed within a half  second of each other, else the unit will return to standard operating mode. When in the service menu  the unit will not update the current O2 readings. Once you enter the menu the display will read “1234”  or “2222”.   Performing an Atmospheric Span Calibration 1. Allow sensor to stabilize for at least 1 hour  2. Ensure the sensor is exposed to fresh air  3. Enter the service menu when you press 123 one after another in that sequentially.  4. The display will show 1234 flashing indicating that you are in the service menu  5. Press 2 to enter the Oxygen submenu  6. Press 2 again to perform the atmospheric calibration  7. Press 3 to escape ...
  • Page 8: Performing A 0% Oxygen Zero Point Calibration

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Performing a 0% Oxygen Zero Point Calibration 1. Allow sensor to stabilize for at least 1 hour  2. Screw on the calibration adapter  3. Apply N2 calibration gas  4. Wait 5 minutes for stable readings  5. Enter the service menu when you press 123 one after another in that sequentially.  6. The display will show 1234 flashing indicating that you are in the service menu  7. Press 2 to enter the Oxygen submenu  8. Press 1 to calibrate  9. Press 3 to escape  Terminal Connectors 1. VDC = G‐ G+  2. O2 4‐20mA = SIG GND  3. RS‐485 INTERFACE = A B GND  4. RESET = RST ...
  • Page 9: Modbus / Rs-485

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Modbus / RS-485 The sensor support MODBUS protocols over RS‐485, unidirectional transmission at 9600kbps. Address 0  or FE.  The following Modbus registers are available:    Input Registers 1. O2 Value, x10  2. Raw ADC Value  Holding Registers 1. O2 Span Value (4 byte floating number, part 1)  2. O2 Span Value (4 byte floating number, part 2)  3. O2 Pot Value  4. 4‐20mA output  Zero Value  5. 4‐20mA output Span Value (4 byte floating number, part 1)  6. 4‐20mA output Span Value (4 byte floating number, part 2)  7. Control Register  8. Status Register     The following commands are available:  0x03 – Read Holding Register ...
  • Page 10: Theory Of Operation

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Theory of Operation The zirconia oxygen sensor utilizes solid‐state electrochemical reactions to produce a voltage  proportional to the current oxygen. This voltage is then digitalized and filtered to a usable value. The  oxygen sensor has also been factory calibrated to meet specifications.  Calibration The calibration process varies depending on the type of unit and whether it has optional data logging  functionality or not. All units are factory‐calibrated with multiple reference points of gas, and have been  verified to be accurate within their specific functionality before shipment. However, if the unit is  severely jolted or otherwise mechanically disturbed, the sensor can drift requiring recalibration. All  calibration procedures follow a single‐point calibration routine that effectively shifts the zero‐point of  the O2 sensor.  Calibrating using MODBUS or PC Connection You can Download the software directly and free from this link in our website.  (http://www.co2meter.com/pages/downloads)   More precise calibrations can be performed using the PC connection. The unit can be attached to a PC  using an optional USB to RS‐485 cable, available on our website, or via a mini‐USB cable to the internal  USB connection.     Calibration of the Oxygen Sensor    In the desktop software application O2 can be calibrated individually. The O2 sensor can be calibrated to  0% and then to 20.9%.  During normal operation 0% calibration of the oxygen transmitter should not be necessary.   Calibration can also be performed over MODBUS, see additional documentation for MODBUS registers.   Rev. 27 April, 2015...
  • Page 11 TR250Z Oxygen Transmitter – Manual CO2Meter.com Calibration can be performed using either 0% CO2 calibration gas (typically nitrogen, Argon, etc.), or  using a fresh source of air, assumed to be approximately 20.9% O2.  Attach calibration gas to the unit and connect the unit to a personal computer. Open the calibration  screen in the DAS software. Click the “Calibrate” button in the calibration tab for the desired gas,  located in the “Configure Sensor” screen. As long as the gas concentration is stable, the unit should  instantly reflect the calibrated value. This can be confirmed by watching the display. To see the  calibration value in real time, click in the “Collect Real time” button to capture these values before  opening the configuration screen.       Real‐time capture  Zero or Fresh Air Calibration Apply gas and select the appropriate concentration and Click the “Calibrate” button. The sensor reading  should instantly reflect the calibration.   NOTE: DO NOT ADJUST THE SPAN VALUE.  Rev. 27 April, 2015 Pg. 11 of 13...
  • Page 12: Support

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Support The quickest way to obtain technical support is via email. Please send all support enquires to  support@co2meter.com. In your email, please include a clear, concise definition of the problem and any  relevant troubleshooting information or steps taken so far, so we can duplicate the problem and quickly  respond to your inquiry.  Warranty This unit comes with a 1YEAR (warranty period) limited manufacturer’s warranty, starting from the date  the unit was shipped to the buyer.  During this period of time, CO2Meter.com warrants our products to be free from defects in materials  and workmanship when used for their intended purpose and agrees to fix or replace (at our discretion)  any part or product that fails under normal use. To take advantage of this warranty, the product must be  returned to CO2Meter.com at your expense. If, after examination, we determine the product is  defective, we will repair or replace it at no additional cost to you.  This warranty does not cover any products that have been subjected to misuse, neglect, accident,  modifications or repairs by you or by a third party. No employee or reseller of CO2Meter.com’s products  may alter this warranty verbally or in writing.  Returns If the product fails under normal use during the warranty period, an RMA (Return Material  Authorization) number must be obtained from CO2Meter.com. After the item is received,  CO2Meter.com will repair or replace the item at our discretion.   To obtain an RMA number, please call CO2Meter.com at (877) 678‐4259. When requesting an RMA  number, please provide the reason for return and original order number.   If we determine that the product failed due to improper use (water damage, dropping, tampering,  electrical damage etc.) or abuse, or if it is beyond the warranty period, we will inform you of the cost to  fix or replace your device.  If you are returning your device due to a warranty claim (with an RMA number) and you still have the  unit original package, please use it to ship your unit to us. Please make sure to include the provided  RMA number on the outside of the box, preferably on the shipping label. Make sure you secure the unit  inside the package properly to prevent any damage during transit that could void your device’s  warranty. Finally, please ship your device to the address shown under the “Contact Us” section below.  CO2Meter.com will not, under any circumstances, be responsible for your shipment expenses and no  refund will be issued for shipping charges necessary for you to ship the unit to us.  Rev. 27 April, 2015...
  • Page 13: Liability

    TR250Z Oxygen Transmitter – Manual CO2Meter.com Liability All liabilities under this agreement shall be limited to the actual cost of the product paid to  CO2Meter.com. In no event shall CO2Meter.com be liable for any incidental or consequential damages,  lost profits, loss of time, lost sales or loss or damage to data, injury to person or personal property or  any other indirect damages as the result of use of our products.  Contact Us We are here to help!  If the troubleshooting guide above doesn’t help you solving your problem or for more information,  please contact us using the information below.    support@co2meter.com   (877) 678‐4259 Toll free (M‐F 9:00am–6:00pm EST)   (866) 422‐2356   www.co2meter.com  Address:  CO2Meter  131 Business Center  Building A, Unit 3  Ormond Beach, FL 32174  Rev. 27 April, 2015 Pg. 13 of 13...