System Configuration
33 cm HDR
36
37
*
41
Max. fluoro time until alarm:
(only VD02A)
Max. fluoro time until blocking:
(only VD02A)
Max. fluoro time for one patient:
(only VD02A)
Cumulated Fluoro time until alarm:
(only VE00J)
Max. fluoro time without interrupt:
(only VE00J)
NOTICE
*
Depending on the software version, different symbols are available.
Siemens AG
Medical Engineering
Set according to country-specific regulations,
e.g. BRD 4.5 min.; GB 9.5 min.
When should FL be shut down without
buzzer deactivation? e.g. BRD 5 min.;
GB 10 min.
When should FL be shut down completely? (
Set according to country-specific regulations,
no shutdown - only alarm
10.0; disabled
e.g. shutdown after 10 min. continuous fluo-
roscopy
Dose Rate Fluoro, Dose Rate DFR und Dose rate can be edited:
If you wish to set different dose values in zoom mode, you must
adjust the zoom ratios accordingly.
RXL2-120.034.05
Rev. 02
Page 7 of 12
12.99
TD PS 24
4 - 7
38 a..c
39
37 a)
(
37 b
(
37 c
*
40
LITHOSTAR Multiline
System Manual
)
)