3. Specification

Sortation
Gas
Detection Principal
Range
Response Time
Gas Sampling
Input Power
Sample Gas Volume
Accuracy
Output Signal
Display
Alarm Indication
Setting of Alram Value
Alarm Delay Time
Alarm Clearing
Alarm Output
Operating Temperature
Operating Humidity
How to install
Inhalation of gas
Output Option
USB Host
Contents
Oxygen
Optical
0.00 ~ 100%
T90(within 15 seconds)
Suction Type
AC 110 ~220 V (2.4W)
0.2 L/min ~ 3 L/min
≤ ±0.5% / Full Scale
4-20㎃ DC/F.S
7" TFT LED (800 X 480)
Alarm 1 – AL1 LCD (YELLOW)
Alarm 2 – LCD (RED)
Fault Alarm – LCD FAULT(RED)
AL2/AL1 2 stages Alarm-User can optionally set
0~99 seconds with user adjustable
Manually or Automatically
2 Step Alarm Relay Contact
-10℃ ~ 60℃
5 ~ 95%RH (Non-Condensing)
Tabletop type
Female 1/4"
RS-485
USB 2.0 (data log download)