Introduction: The Spectra 2100 Series Portable Automatic Water Sampler has been designed to offer the user a wide range of options in a specialized environmental monitoring instrument. The unit uses a 12 VDC power source and can be operated using a 230/120 VAC to 12 VDC power supply, 12 VDC battery or 12 VDC battery\solar panel package for remote location sampling.
Locating the Sampler: 1. Select a location on level surfaces when possible. 2. Avoid areas where wet or soggy ground or standing water is found. 3. Be aware of locations that may flood or collect water during storm events at the sampler location. 4.
Programming Guide: After system is on, the display will show the following: MODEL 2100 Automatic Water Sampler Spectra Technologies 2020-02-20 20:05:46 With the above display Press ↑(up) to make the pump rotate forward. Press ↓ (down) to make the pump rotate in the reverse direction.
From the main startup screen. MODEL 2100 Automatic Water Sampler Spectra Technologies 2020-02-20 20:05:46 Press the Tab down key. You will see the following screen. MODEL 2100 Automatic Water Sampler Manual Sampling 0100 ml (50-1000ml) Press Enter to confirm. Press to cancel.
Page 8
Menu Sample Setting Flowmeter System Setting Factory Setting About Exit Select “Sample Setting” Sample Setting Set sample Query Exit Press ↵ (Enter) to sample parameter setting interface: Sample Parameter Setting Sampling Delay Time:0001 (1-9999) Sampling Interval Time: 0001 (1-9999) Sampling amount 0300 ml ( 50-500 ) ml Mixed number (1-200) Sampling bottle number 01 Press Enter to save...
Sample Setting Sampling setting Query Exit Query Sampling Time Delay 0001 ( 1-9999) Sampling Interval Time 0001 (1-9999) Sampling amount 0050 ml(50-500) 5. Flow Paced Sampling The input terminal for Flow Paced sampling is found on the right side of the pump enclosure. Input Terminal Pinouts RS-232 Ground...
Page 10
For example, if the 4-20 ma is set to 0 to 5000 units and the Start value is 1000, then the unit will sample every 1000 units. For Pulse outputs Set the desired number of relay closures depending on the flowmeter setting for sampler output.
System setting Time Setting Communication Setting Peristaltic Pump Parameter setting Password Setting Blacklight Exit Press ↵(Enter) to Time Setting: Time Setting 20 -02- 20:35:58 Press Enter to confirm Press to cancel 7. Suction Tube Length Setting Go into Suction Tube Length Setting interface, select the normal length. Suction Tube Length Setting Current tube length:...
How to set the Calibration: 1. Set up the length of suction tube. 2. Press ↑(up) on the start-up interface to make the pump rotate forward. If there’s water is dispensed then the set up was successful; Otherwise, continue to press ↑(up) until water is dispensed. 3.
11. Blacklight Time Blacklight Time Setting Current time set: 300 Setting time: 000 Press Enter to confirm Press M to cancel 12. Refrigerator Connect to the power and press ON/OFF key to run on. Now the “Power” light is on. The digital display will show the current temperature inside the chamber.
About Automatic Water Sampler Hardware Number: Product Number: Usage Counts: 14. Temperature Controller: Heating Mode Setting Note 1:Start up temperature < Stop temperature The program will automatically turn on the heating mode. Note 2: The model 2100 uses a dual zone temperature controller. The display shows the sample chamber temperature and is used for programming the temperature of this location.
Page 15
When the compressor, module, fan, or temperature sensor fails or the input voltage is too low, an error code will be displayed on the control panel. When the compressor stops working. When the system is activated for 3 minutes after the fault is eliminated, the refrigerator will enter the normal working mode.
charge, reverse charge, and short circuit protection for the battery. This technology is designed to intelligently manage the charging of the battery while protecting the battery from damage due to over-charging. The built-in charge controller is equipped with LED status indicators that lets the user know the charging status of the battery as well as diagnostic information should any issue arise.
Page 17
Green Flash 1 Hz Sunlight is sufficient (Battery not connected) None No Output Indicator Flash 1-2 Hz Charging (Actively Charging) Green None Battery is fully charged None Reversed connection/Short Circuit Technical Specifications Rated Power Output 10 W Power Tolerance +/- 5% Operating Voltage 18 V Operating Current...
Warranty Vortex Technologies Inc. warrants products of its own manufacturer to be free from defects in material and workmanship, when installed according to the instructions provided by Vortex Technologies, Inc. and used under normal conditions and service for a period of one year from the date of shipment, provided notice is given within the warranty period.