Installing the probe again Troubleshooting 10.1 Signs of an unsafe operating condition 10.2 Sensor error – Indication of an error on the MLA1000 display 10.3 Fatal error – Indication of an error on the MLA1000 display 10.4 Remedying faults Technical data 11.1...
Pass on the documents to new owners. Safety instructions for use in potentially explosive atmospheres The MLA1000-A display unit and the MLA1000-S probe of meter types MZ, LZ and TZ operated in a potentially explosive atmosphere must bear the following markings:...
Meter types TX and TZ for temporary measurement The types TX and TZ can be easily installed to use this device for temporary measurements. If the location for the measurement changes, the MLA1000 TX and TZ must be completely installed before it is put into operation.
Measuring range (5-2000 means: measuring range from 5 pS/m to 2000 pS/m) V0.95 5-2000 2.2.5 MLA1000-K probe cable The probe cable supplied is specially designed to meet the requirements of the MLA1000 meter. Only the probe cable supplied may be used. ...
Installation may only be carried out by persons who are familiar with the explosion protection regulations at the installation site. The equipment protection level (EPL) of the MLA1000-S and MLA1000-A must match the Ex zones of the installation site. ...
The MLA1000 meter type TZ – additional requirements The display unit of the MLA1000-A type TZ is mounted on a tripod for transport the meter to the measuring site. At the measuring site following is mandatory: ...
Probe cable for MLA1000 TX The probe cable is already connected to the terminals on delivery of the MLA1000 meter type TX. Should it be necessary to replace the probe or probe cable, the terminal assignment diagram can be found in section 6.
The real physical value is higher than the measuring range end value 24 mA (overflow). Supply voltage A fuse must be installed in the supply line for the MLA1000 supply voltage. Fuse rating: max. 5 A Connect the supply voltage as shown in the terminal diagram. ...
Operation The MLA1000 starts to operate as soon as the power supply is switched on. Signs of a safe operating condition The measurement liquid maintains the necessary operating conditions at the measuring probe. The display of the display unit shows measured values.
Notes on explosion protection for meter type TZ The plastic case is only for storage of the MLA1000. The case must not be brought into an Ex zone. The display unit is approved for use in ATEX zone 1. It must not be brought into ATEX zone 0.
Maintenance The MLA1000 has no wear or consumable parts. Nevertheless, the meter should be inspected at regular intervals. Carry out a visual inspection to check that no parts of the MLA1000 are damaged. Notes on explosion protection for maintenance ...
Troubleshooting The MLA1000 is designed for continuous, maintenance-free operation. The following description can help to quickly rectify a fault in the meter. 10.1 Signs of an unsafe operating condition The display of the display unit is blank. The display of the display unit shows a fault.
There is not exchange of the liquid to be measured at the Check the flow velocity and change, if necessary. probe. The measured values of the MLA1000 do not correspond to a The liquid was not homogeneously mixed. The sample The sample is not of the same liquid that was measured reference measurement in the laboratory.