cytiva P-1 Operating Instructions Manual

cytiva P-1 Operating Instructions Manual

Peristaltic pump
Hide thumbs Also See for P-1:

Advertisement

Quick Links

Peristaltic Pump P-1
Operating Instructions
Original instructions
cytiva.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for cytiva P-1

  • Page 1 Peristaltic Pump P-1 Operating Instructions Original instructions cytiva.com...
  • Page 2: Table Of Contents

    Eurasian Economic Union Евразийский экономический союз ....................45 7.4.4 Regulations for North America ....................... 47 7.4.5 Regulatory statements ..........................48 7.4.6 Declaration of Hazardous Substances (DoHS) ................49 Health and Safety Declaration Form ...................... 51 Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 3: Introduction

    1 Introduction Introduction About this chapter This chapter contains important user information, descriptions of safety notices, regu- latory information, intended use of the Peristaltic Pump P-1 instrument, and lists of associated documentation. In this chapter Section See page About this manual...
  • Page 4: About This Manual

    The Operating Instructions provide you with the information needed to install, operate and maintain the product in a safe way. Scope of this manual The Operating Instructions cover the Peristaltic Pump P-1 instrument. Typographical conventions Software items are identified in the text by bold italic text.
  • Page 5: Important User Information

    Peristaltic Pump P-1 is a single channel laboratory pump for use in liquid chromatog- raphy and other applications where accurately controlled liquid flows are required. Peristaltic Pump P-1 is intended for research use only, and shall not be used in any clin- ical procedures, or for diagnostic purposes.
  • Page 6 Note: A note is used to indicate information that is important for trouble-free and optimal use of the product. Tip: A tip contains useful information that can improve or optimize your proce- dures. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 7: Associated Documentation

    Introduction This section describes the user documentation that is delivered with the product, and how to find related literature that can be downloaded or ordered from Cytiva. System-specific documentation In addition to the Operating Instructions manual, the documentation package supplied with Peristaltic Pump P-1 also includes one or more product documentation binders containing detailed specifications and traceability documents.
  • Page 8: Safety Instructions

    Before installing, operating or maintaining the product, all users must read and understand the entire contents of this chapter to become aware of the hazards involved. In this chapter Section See page Safety precautions Labels Emergency procedures Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 9: Safety Precautions

    2.1 Safety precautions Safety precautions Introduction Peristaltic Pump P-1 is powered by mains voltage and handles materials that can be hazardous. Before installing, operating or maintaining the system, you must be aware of the hazards described in this manual. Follow the instructions to avoid injury to the operator or other personnel, damage to samples or other substances handled by the equipment, to the product, or to other equipment in the area.
  • Page 10 2 Safety instructions 2.1 Safety precautions WARNING Do not use Peristaltic Pump P-1 if it is not working properly, or if it has suffered any damage, for example: • damage to the power cord or its plug • damage caused by dropping the equipment •...
  • Page 11 Power cord. Only use power cords with approved plugs delivered or approved by Cytiva. WARNING Access to power cord. Do not block access to the power cord. The power cord must always be easy to disconnect. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 12 Power supply WARNING Electrical shock hazard. All repairs should be done by service personnel authorized by Cytiva. Do not open any covers or replace parts unless specifically stated in the user documentation. WARNING Disconnect power. Always disconnect power from the instru- ment before replacing any component on the instrument, unless stated otherwise in the user documentation.
  • Page 13 WARNING Use only approved parts. Only spare parts and accessories that are approved or supplied by Cytiva may be used for maintaining or servicing the product. WARNING Corrosive substance. NaOH is corrosive and therefore dangerous to health.
  • Page 14: Labels

    Instrument assembly number Serial no Instrument serial number Mfg Year Year (YYYY) and month (MM) of manufacture Voltage Electrical requirements: Frequency • Voltage (VAC Max Power • Frequency (Hz) • Max. power (VA) Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 15: Emergency Procedures

    2.3 Emergency procedures Emergency procedures Introduction This section describes how to shut down Peristaltic Pump P-1 in an emergency situa- tion. The section also describes the result in the event of power failure. Emergency shutdown In an emergency situation, follow the steps below to stop the run:...
  • Page 16: Instrument Description

    3 Instrument description Instrument description About this chapter This chapter gives an overview of the Peristaltic Pump P-1 instrument, and a brief description of its function. Introduction to Peristaltic Pump P-1 Peristaltic Pump P-1 is a single channel laboratory pump for use in liquid chromatog- raphy and other applications where accurately controlled liquid flows are required.
  • Page 17 Side panel for electrical connections flow rate potentiometer Indicators and controls on the front panel Indicator/Switch Description run/stop (switch) Starts and stops the pump. Can be over- ridden by the remote control. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 18 (potentiometer) A continuously variable potentiometer for the control of the flow rate. Gives a linear response in the range from 1 to Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 19 The following illustration shows the side panel viewed when the pump is on its back. Connection Connection Remote: connector for remote Mains power inlet control and pulse counting Fuse holder and voltage selector Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 20: Installation

    4 Installation Installation About this chapter This chapter provides required information to enable users and service personnel to unpack, install, move and transport the Peristaltic Pump P-1 instrument. In this chapter Section See page Site requirements Transport Unpack the product...
  • Page 21: Site Requirements

    Altitude, operation Up to 2000 m Pollution degree of the Pollution degree 2 intended environment Electrical power requirements Parameter Requirement Supply voltage 100-240 V AC Frequency 50-60 Hz Transient overvoltages Overvoltage category II Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 22: Transport

    4 Installation 4.2 Transport Transport Introduction This section gives instructions for moving the Peristaltic Pump P-1 instrument within the laboratory. Moving the instrument Before moving the instrument, disconnect all cables and tubing connected to periph- eral components and liquid containers.
  • Page 23: Unpack The Product

    Set the equipment upright before you start to assemble the product. Visual inspection Inspect all visible parts for damage or missing pieces. If any damage is observed, record this on the receiving documents and inform your Cytiva representative. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 24: Assemble The Product

    Access to power cord. Do not block access to the power cord. The power cord must always be easy to disconnect. Required assembly The following parts must be added to the Peristaltic Pump P-1 instrument before it can be used: • Tubing •...
  • Page 25 4.4 Assemble the product Step Action Two Electrical Power Supply Kits are supplied with Peristaltic Pump P-1, one for 100–120 V and one for 220–240 V. Choose the kit appropriate to your mains supply voltage, and discard the other kit.
  • Page 26 Insert the tubing. Make sure the nuts of the connectors fit correctly into their housings. Close the lid. The correct tubing pressure is set automatically. Note: When the pump is not in use, leave the lid open to prolong the life time of the pump tubing. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 27: Connections

    Connections Flow path Peristaltic Pump P-1 may be connected either before or after a chromatography column. Air bubbles may appear in the bed if the pump is connected after the column and the eluent has not been degassed. However, the advantage of this set-up is that the flow can be controlled by the pump during sample application as well as during elution.
  • Page 28 If the product is connected to laboratory scaffolding, make sure that the laboratory scaffolding is stable. Electrical power Connect the power cord to a grounded power outlet as specified in Section 4.1 Site requirements, on page Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 29: Operation

    This chapter gives instructions on how to operate the product in a safe way. In this chapter Section See page Set the flow rate Start and stop the pump Connections to the remote control socket Calibration of volume information Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 30: Set The Flow Rate

    5.1 Set the flow rate Set the flow rate Peristaltic Pump P-1 has two flow rate ranges for each tubing diameter. The appro- priate range is chosen by setting the flow rate switch to ×1 or ×10 giving a ten-fold variation of the flow rate.
  • Page 31: Start And Stop The Pump

    During the run it is possible to: • Adjust the flow rate with the flow rate potentiometer. • Reverse the flow with the flow direction switch. Press the run/stop switch to the stop position to stop the pump. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 32: Connections To The Remote Control Socket

    Section 5.4 Calibration of volume information, on page Connection to a controller For use in completely automated systems, Peristaltic Pump P-1 can be controlled for example from ÄKTA™ systems using I/O-signals. For further information refer to the relevant controller’s Instruction Manual.
  • Page 33: Calibration Of Volume Information

    5.4 Calibration of volume information Calibration of volume information Below is a step by step procedure for the calibration of the volume information trans- mitted from the Peristaltic Pump P-1 instrument to the fraction collector when connected to the remote control. Step Action Set up the chromatography system as desired.
  • Page 34: Maintenance

    Always clean and drain the equipment completely before moving it. Maintenance schedule Peristaltic Pump P-1 does not need any regular servicing or maintenance other than the replacement of worn out tubing. It is advisable to replace the tubing after 200 hours to minimize the risk of tubing breakage during operation.
  • Page 35 Please complete the checklist in the On Site Service Health and Safety Declaration Form or the Health and Safety Declaration Form for Product Return or Servicing, depending on whether the instrument is going to be serviced on site or returned for service, respectively. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 36 Health and safety declaration forms Health and safety declaration forms are available for copying or printing in the Refer- ence information chapter of this manual, or on digital media supplied with the user documentation. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 37: Reference Information

    7 Reference information Reference information About this chapter This chapter lists the technical specifications of Peristaltic Pump P-1. The chapter also includes ordering information and Health and Safety Declaration forms for service. In this chapter Section See page Specifications Ordering information...
  • Page 38: Specifications

    IP 20 Environmental requirements Parameter Requirement Allowed location Indoor use only Ambient temperature 4°C to 40°C Relative humidity tolerance (non- 20% to 95% condensing) Atmospheric pressure 84 to 106 kPa (840 to 1060 mbar) Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 39: Ordering Information

    7 Reference information 7.2 Ordering information Ordering information Contact details For ordering information visit cytiva.com/products. Spare parts and accessories For correct up to date information on spare parts and accessories visit: cytiva.com/ products Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 40: Recycling Information

    Instructions for disposal Follow the instructions below for disposal of the product: Step Action Separate all electronic components (terminal strips, power supplies, trans- mitters, pumps, probes / sensors, etc.) from the Electrical Cabinet. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 41 Follow local and/or national/federal requirements for disposal of the Peristaltic Pump P-1 Cabinet and the Electrical Cabinet. Dispose of electronic components as specified by local regulations depending on material used in the construction of the components.
  • Page 42: Regulatory Information

    See page 7.4.1 Contact information 7.4.2 European Union and European Economic Area 7.4.3 Eurasian Economic Union Евразийский экономический союз 7.4.4 Regulations for North America 7.4.5 Regulatory statements 7.4.6 Declaration of Hazardous Substances (DoHS) Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 43: Contact Information

    The table below summarizes the required manufacturing information. Requirement Information Name and address of manufacturer Cytiva Sweden AB Björkgatan 30 SE 751 84 Uppsala Sweden Telephone number of manufacturer + 46 771 400 600 Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 44: European Union And European Economic Area

    • used according to the Operating Instructions or user manuals, and • used in the same state as it was delivered, except for alterations described in the Operating Instructions or user manuals. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 45: Eurasian Economic Union

    В следующей таблице приводится сводная информация о производителе и импортере, согласно требованиям Технических регламентов Таможенного союза и (или) Евразийского экономического союза. Требование Информация Информацию об изготовлении Наименование, адрес и номер См. телефона производителя Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 46 Member States of the Customs Union of the Eurasian Economic Union Данный знак о Евразийском соответствии указывает, что изделие одобрено для использования на рынках государств-членов Таможенного союза Евразийского экономического союза Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 47: Regulations For North America

    Note: The user is cautioned that any changes or modifications not expressly approved by Cytiva could void the user’s authority to operate the equip- ment. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 48: Regulatory Statements

    When used in a residential environment, there is a concern of radio interference. 주의사항 A급 기기 (업무용 방송통신 기자재) 이 기기는 업무용환경에서 사용할 목적으로 적합성평가를 받 은 기기 로서 가정용 환경에서 사용하는 경우 전파간섭의 우려가 있습 니다. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 49: Declaration Of Hazardous Substances (Dohs)

    Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be collected sepa- rately and handled properly after decommissioning. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 50 Indicates that this hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in GB/T 26572 • Data listed in the table represents best information available at the time of publication. Peristaltic Pump P-1 Operating Instructions 28962201 AG...
  • Page 51: Health And Safety Declaration Form

    Service Ticket #: To make the mutual protection and safety of Cytiva service personnel and our customers, all equipment and work areas must be clean and free of any hazardous contaminants before a Service Engineer starts a repair. To avoid delays in the servicing of your equipment, complete this checklist and present it to the Service Engineer upon arrival.
  • Page 52 To make sure the mutual protection and safety of Cytiva personnel, our customers, transportation personnel and our environment, all equipment must be clean and free of any hazardous contaminants before shipping to Cytiva. To avoid delays in the processing of your equipment, complete this checklist and include it with your return.
  • Page 53 Page intentionally left blank...
  • Page 54 © 2020 Cytiva All goods and services are sold subject to the terms and conditions of sale of the supplying company operating within the Cytiva business. A copy of those terms and conditions is available on request. Contact your local Cytiva representative for the most current information.