Metron Medical Australia Pty Ltd A.C.N.050 240 527 P.O. Box 2164, 57 Aster Avenue Carrum Downs Victoria Australia 3201 Tel: (03) 9775 1234 Fax: (03) 9775 1990 Int: 61 3 9775 1234 Int: 61 3 9775 1990 TECHNICAL MANUAL METRON ACCUSONIC PLUS...
Page 2
Metron Medical Australia Pty Ltd 57 Aster Avenue Carrum Downs, Australia, 3201 declares that the medical devices described hereafter: Metron Accusonic Plus Ultrasound Therapy Unit Model: AP 100 is in conformity with the essential requirements and provisions of Council Directive 93/42/EEC...
Page 3
Metron Accusonic Plus 0120 The Metron Accusonic Plus Ultrasound Therapy Unit bears the above marking in accordance with the requirements of Council Directive 93/42/EEC. Should you as the user of this technical manual wish to make any comment about the...
Metron Accusonic Plus SPECIFICATIONS MAINS SUPPLY REQUIREMENTS: Voltage 110 - 120 Volts AC or 220 - 240 Volts AC Frequency 50/60 Hz Power 75VA Maximum F U S E S : Primary external 2 of 1A 5 X 20 mm DA205...
Page 6
Metron Accusonic Plus Effective Radiating Area Nominally 0.8 square cms +/- 20% (1 cm head) Nominally 5.0 square cms +/- 20% (2.5 cm head) Power/Intensity Display Accurate to +/- 10% of reading for outputs in excess of 1 Watt. ULTRASOUND MODULATION:...
Page 7
Metron Accusonic Plus ELECTRICAL SAFETY: Designed and manufactured to comply with the following Australian and International standards: AS 3100 - 1985 Definitions and general requirements for electrical materials and equipment. AS 3200.1 - 1990 Approval and test specification - Medical electrical equipment, Part 1 General requirements for safety.
Metron Medical Australia Pty Ltd. Refer to the Metron Accusonic Plus Operator Manual for operator information.
Page 9
Metron Accusonic Plus The degree of acoustic coupling between the ultrasonic treatment applicator and the patient is monitored by the contact sense circuit and the microcontroller. At detection of contact loss the indicator on the selected applicator turns RED, the treatment timer is halted and after 2 seconds of contact loss the ultrasonic output power is reduced from that selected to 0.2 Watts/cm...
Metron Accusonic Plus SCHEMATIC DIAGRAM DESCRIPTION 3 . 1 PRINTED CIRCUIT BOARD (see Schematic, page 11) 3 . 1 . 1 Main Power Supply. The DC voltage from the switchmode power supply is applied to the electronics through the diode bridge BR1. The bridge provides a level of protection against wiring errors.
Metron Accusonic Plus 3 . 1 . 4 Displays & Drivers The seven segment displays, LED 11 and LED 12, and the individual LEDs, LED1 through LED8, are driven by display drivers chips IC5 and IC6. These 35 segment drivers and controlled by serial information sent from the microprocessor to their CK and DATA pins.
Bio-Tek Instruments UW-II UMA Inc UMT-2A or equivalent 4 . 1 . 2 Calibration Procedure Connect the applicator to be calibrated to either input on the Accusonic Plus. NOTE: It is recommended that treatment applicators only be attached and removed from the unit when the power is turned OFF.
Metron Accusonic Plus At the completion of the calibration procedure it is required that the mains power be turned OFF and then turned ON again to return the Accusonic Plus to the normal operation mode. WARNING: If the mains power is removed or turned OFF prior to the completion of the calibration of an applicator the whole procedure will need to be repeated.
Metron Accusonic Plus SCHEMATIC DIAGRAMS 5 . 1 Main Circuit Board Layout __________________________________________________________________ Version 1.6 March 2002 Page 10 of 12...
Metron Accusonic Plus 5 . 2 Schematic Diagram - Main Printed Circuit Board __________________________________________________________________ Version 1.6 March 2002 Page 11 of 12...
Metron Accusonic Plus 5 . 3 Schematic Diagram - Treatment Head Printed Circuit Board __________________________________________________________________ Version 1.6 March 2002 Page 12 of 12...