Page 1
SC28 Monitor for Magnetic Fields, Vibrations, Acoustics, Temperature and Humidity Version 1.0 Beta 1 for Windows 7, 8.1, 10 User Manual SPICER CONSULTING Electron Beams & Electronic Systems...
Contents About this Manual ....................5 Organisation of this Manual ................. 5 Conventions used in this Manual ..............6 Chapter 1 - Introduction to the SC28 System ............7 Hardware ...................... 7 Software ......................7 Sensors ......................7 Getting Started ....................8 Adding Sensor Calibration Files ..............
Page 4
Chapter 7 – Data Logging Controls ............... 30 Chapter 8 – Environmental Specifications ............33 Specifications ..................... 33 Specification Editor ..................34 Chapter 9 – SC28 Calibration ................37 Calibration Panel ..................37 User Defined Units ..................41 Chapter 10 - Menu and Toolbar Functions ............43 Menu ......................
Chapter 3 – Setting up Monitoring, explains how to use the electron beam instrument’s environmental specifications to guide the setting of the SC28 controls. Chapter 4 – Signals and Scales, is a detailed reference on the SC28 signals and scales settings.
The SC28/SI supports a wired Ethernet connection. Software The SC28 software runs on a Windows PC that is connected to the same Local Area Network as the SC28/SI. It automatically detects the SC28/SI hardware without needing to know its IP address in advance. Included with the SC28 software is the SCplot program, which views and prints results or saves graphs that you can use in reports.
When you run the SC28 software, it automatically detects the SC28/SI hardware and connects to it. If no SC28/SI is detected the software will wait until one is connected to the LAN. If more than one SC28/SI is detected on the LAN, the software asks which you wish to connect to.
Sensors section in Chapter 9 - Calibration. How to Run the Programs To run the SC28 programs, select one of the following from: Start Menu > All Programs > Spicer Consulting > SC28 1.0 > SC28 1.0 SCplot 5.6 SC28 is the main monitoring program that performs measurements and logs results to files on the PC.
For full details see the SCPlot User Manual. Getting Started Start the SC28. The SC28 should start up in the task bar. Click the SC28 task bar icon to open the SC28 main window. The status should show Starting… and then Connected to SC28/SI with its serial number, firmware version and IP address.
Help is available in the following ways: Help <F1> • To get help, click the Help button or select Help for SC28 from the Help menu. • Tip strips appear when you move the pointer over the icons and sensor controls, providing additional information.
The methods and units of the specifications vary between manufacturers, but the SC28 is flexible enough to measure the parameters found in all mainstream manufacturers’ specifications. Magnetic Field Sources Power Frequency (50 or 60 Hz) magnetic fields are often produced by current- carrying power cables.
Chapter 3 – Setting up Monitoring Setup Steps The following steps are needed to set up the SC28 to monitor the environment of an electron beam instrument: 1. Check the Specifications 2. Define Position and Orientation 3. Set up Signals and Scales 4.
Lower and Upper bandwidths for each signal. Where possible, be guided by the microscope specification. The maximum upper bandwidth in the SC28 is 1 kHz. Bandwidths are also limited by the sensors in some cases. See Chapter 6 – Chart Recorder Controls for more details.
The SC28 provides two types of data logging files Chart and Spectrum. Each have their own folder, file name and file interval settings. In addition, the SC28 can automatically delete old log files to save disc space. See Chapter 7 - Data Logging Controls for more details.
Chapter 4 – Signals and Scales The SC28 provides 10 channels of data acquisition. In the SC28 Monitor, these signals may be displayed on 6 scales and in 3 display formats (Scope, Spectrum and Chart). Up to 14 different graphs may be displayed giving simultaneous display of the data as it is acquired.
• It is not likely to find temperature or humidity out of range of the sensor. Scales and Ranges The SC28 provides unit and range settings for each scale. The User scale has a selectable sensor, depending on the user defined units available.
Page 21
The SC28 provides the following units and ranges by default. User-defined units may be added. See the User Defined Units section of Chapter 9 – Calibration for how to add user-defined units. SC28 Amplitude Ranges Scale Sensor Unit Ranges Magnetic...
1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 In the SC28 Monitor, the maximum chart length is 3600 points, so it is always at least 1 hour. The chart display normally shows the current chart data log file as it is being built up.
Data Logging control panel. The SC28 provides two types of data logging files Chart and Spectrum. Each have their own folder, file name and file interval settings. In addition, the SC28 can automatically delete old log files to save disc space.
Page 32
Start and Finish refer to the start and finish times of the data in the file. All date and time information, including the time zone, is derived from the PC running the SC28 program. The SC28 starts a new file at the file interval that you set. The following are available: Unit...
Page 38
Reset DC button in the tool bar of the SC28 Monitor. The following window appears: Initially the DC sensor outputs are zero, while it internally zeros its output to the nearest 1 mG at the current X, Y and Z magnetic field levels. After that, the DC sensor outputs turn on and the software uses low-pass filtering to compensate for any residual offset.
Page 39
Add Sensors Sensors are made available for selection in the Calibration panel by adding their calibration files to the SC28 system. Click the Add Sensors… button. Have Calibration Files Spicer Consulting provides calibration files for sensors on a USB memory stick.
Page 40
The file dialog appears. Browse to the location of your calibration files, select the files you wish to add to the SC28 system and click OK. You can select all the files using <ctrl-A> (recommended), individual files using <ctrl-click> or a range of files using <shift-click>.
Under the Units tab, you can enter user-defined units for use with the AUX input (or the MAG or MIC input of the Sensor Interface). For example, you might connect a Spicer Consulting SC22 Magnetic Field Cancelling System BNC output to the AUX input. The SC22 measures magnetic field and outputs 1.0 Volts per milliGauss with 10 kΩ...
Show the SC28 help file. About… Show brief information about the program. Calibration Tool Calibration <F2> Calibration is described in Chapter 9, SC28 Calibration. Specification Tool Specifications <F9> Specifications are described in Chapter 8, Environmental Specifications. Setup Files Open Setup <F3>...
See Setup Files above. Resizing the Window You can re-size the window of the SC28 Monitor by dragging the edges or corners, by clicking on the Maximise button near the top right or by using Full Size on the Window menu.
Bits Input Range Resolution ±10V 2.44 mV Core System: SC28/SI - Sensor Interface (Ethernet-based) Inputs 3-axis magnetic field sensor VIBX, VIBY, VIBZ 3-axis vibration (3 x Wilcoxon 731A) Microphone (B&K 4190/2669L) BNC voltage input, DC coupled, ±10 V range, 100 k input impedance.
Page 47
1.5 Hz - 20 kHz Dynamic Range 110 dB (in this system) Noise Limit 20 dB (in this system) Accuracy ±1 dB 3 Hz - 20 kHz Temperature/Humidity Sensor: SC28/TH Sample rate 1 Hz max Performance Temperature Relative Humidity Dynamic Range 0 - 100 °C...
Page 49
Normal type appears in the file. C-style for loops make repeated lines. Data values are separated by tabs. Italics are substituted as follows: Version Current version of SC28 Program Program that created the file (SC28) FileMode How the file was output (Logged) Date & Time Date and time data logging started SetupFile Name of program setup file (setup.sc28)
Page 51
BWupper Scale upper bandwidth MXclip..Hclip Clipping indicators for all channels (0=not clipping, 1=clipping) The SC28 Chart Datalog may format the horizontal signal data H0 to Hnp as an ISO 8601 compliant time stamp of the form: yyyy-mm-ddTHH:MM:SS+/-ZZ:zz where: Yyyy-mm-dd Date with 4-digit year, 2-digit month and 2-digit day of month...
Default Folders By default, all executables, dynamic link libraries and originals of default setup files for SC28 1.0 are installed in one of the following folders. These files should not be altered. C:\Program Files\Spicer Consulting\SC28 1.0 (32-bit Windows) C:\Program Files (x86)\Spicer Consulting\SC28 1.0 (64-bit Windows)
Appendix 3 - Units This appendix gives some general information about the units used in the SC28 software. Magnetic Field Units The SI unit for magnetic flux density, B is T (Tesla). 1 T is a strong magnetic field. The older unit G (Gauss) is often used.1 G = 10 mG (milliGauss), µT (microTesla) and nT (nanoTesla) are used for typical...
Need help?
Do you have a question about the SC28 and is the answer not in the manual?
Questions and answers