Accessories Service center SEITRON S.p.A. a socio unico - ALL RIGHTS RESERVED - Total or partial reproduction of this document by any means (including photocopying or storage on any electronic medium) and transmittal of same to third parties in any manner, even electronically, is strictly prohibited unless explicitly authorised in writing by SEITRON S.p.A.
IMPORTANT INFORMATION 1.1 Information about this manual This manual describes the operation, the features and maintenance of the external Cooler AACE01. Read this maintenance and operation manual before using the device. The operator must be familiar with the manual and follow its instructions carefully.
All interventions not specified in this manual can only be performed by Seitron service centers. If this latest point is not complied with, Seitron declines all responsibility for the normal operation of the device and the validity of the relevant approvals.
It is recommended to provide this number for any request for technical assistance, spare parts or technical information. Seitron maintains updated at its headquarters an archive with the log of the data of each device. Warranty In the view of a constant development of their products, the manufacturer reserves the right for changing technical data and features without prior notice.
4.0 TECHNICAL FEATURES 4.1 technical features Power supply: 110-240 Vac or 12-16Vdc 5A Power absorption: < 50 W Drying gas method: water is condensed by rapidly cooling down the gas using a cyclonic system (vortex) Method of condensate separation: Peltier unit Set point temperature: +5°C Temperatura di Isteresi:...
5.0 CONNECTION DIAGRAMS 5.1 CONNECTION DIAGRAM USING THE EXTERNAL COOLER POWERED AT 110-240Vac TO INSTRUMENT “IN” INPUT TO INSTRUMENT “T1” INPUT TO INSTRUMENT “P-” INPUT REMOTE WATER DRAIN PARTICULATE FILTER CONNECT TO THE MAINS VOLTAGE 110-240V~ VIA THE SUPPLIED CABLE GAS SAMPLING PROBE...
5.2 CONNECTION DIAGRAM USING THE EXTERNAL COOLER POWERED WITH LOW VOLTAGE 12-16Vdc TO INSTRUMENT “IN” INPUT TO INSTRUMENT “T1” INPUT TO INSTRUMENT “P-” INPUT REMOTE WATER DRAIN PARTICULATE FILTER CONNECT TO A LOW- VOLTAGE SOURCE SUCH AS A CAR BATTERY STARTER GAS SAMPLING PROBE...
5.3 CONNECTION DIAGRAM EXCLUDING THE EXTERNAL COOLER TO INSTRUMENT “IN” INPUT TO INSTRUMENT “T1” INPUT TO INSTRUMENT “P-” INPUT PARTICULATE FEATURED FILTER EXTENTION INSTALL THE PROVIDED FEMALE/FEMALE CONNECTOR GAS SAMPLING PROBE...
6.0 OPERATION In order to operate the entire system, proceed as follows: 1. Carry out the connections, as shown in chapter 5.0 CONNECTION DIAGRAMS. 2. Switch on the instrument. 3. In case the chosen diagram foresees the use of the external cooler: - Power the cooler with the supplied cable.
7.0 MAINTENANCE 7.1 TIGHTNESS TEST OF THE ENTIRE COOLER SYSTEM In order to verify the tightness of the whole Cooler system (Cooler=>Smoke sampling probe=>condensation trap), aiming to exclude pressure leaks because of some damaged or broken parts, do as follows: Run all connections mentioned in the chapters “5.1 and 5.2 CONNECTION DIAGRAM USING THE EXTERNAL COOLER ”.
7.2 PERISTALTIC PUMP ROTOR REPLACEMENT For the replacement of the peristaltic pump rotor, proceed as follows Switch off the instrument and disconnect the power cable from the instrument connector. Locate the rotor of the peristaltic pump. Pull out the rotor of the peristaltic pump, holding inward the two plastic fins indicated by the arrows and at the same time, extract the rotor.
7.3 FUSES REPLACEMENT If it is necessary to replace the fuses of the instrument, proceed as follows. For technical specifications of fuses, see Chapter “2.0 Technical features”. Switch off the instrument and disconnect the power cable from the instrument connector. Locate the fuse holder drawer and pull it out.
7.4 MAINTENANCE OF THE PARTICULATE TRAP / FILTER ASSEMBLY REPLACE THE PARTICULATE FILTER IF IT IS VISIBLY DIRTY OR HUMID. DO NOT PERFORM MEASURES WITHOUT THE FILTER OR WITH A DIRTY FILTER IN ORDER TO AVOID IRREVERSIBLE DAMAGES TO THE SENSORS. Replacing the particulate filter In case the particulate filter is blackened, especially on the outside surface, it becomes necessary the immediate replacement.
750 mm probe for industrial motors, with 3 m. cable. AACEX02S 3 m extension cable for gas sampling probe. AASP01 Heat protective screen for gas sampling probe. 8.3 Service Centers Seitron S.p.A. a socio unico Via del Commercio, 9/11 36065 Mussolente (VI) Tel.: +39.0424.567842 Fax.: +39.0424.567849 E-mail: info@seitron.it http://www.seitron.it...