Guardian™ Jr. Airflow Monitor Models Applies primarily to older Models 4865000, 4865001 Some operation and troubleshooting information applies for newer 97432 Series See separate Instruction Sheets for installation on newer 97432 Series...
Page 2
Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted.
Thank you for displaying confidence in us by selecting a Labconco Guardian Jr. Airflow Monitor. Our design engineers, assemblers and inspectors have utilized their skills and years of experience to ensure that the new Guardian Jr. Airflow Monitor meets our high standards of quality and performance.
Chapter 1: Introduction Performance The Guardian Jr. Airflow Monitor is designed for operation on all new non- explosion proof remote and integral style Protector Fume Hoods. It is also set up for operation for both the new Basic 47 and 70 inch style fume hoods. The monitor itself is available both as a field installed kit or a factory installed option.
Chapter 1: Introduction Component Identification Air Inlet A portion of the air coming into the hood passes through the air inlet and across the flow sensors. Normal Flow Indicator This green light indicates normal flow conditions. Alarm Indicator This red light is activated approximately 6 seconds after the low flow set point is reached.
Disconnect your fume hood from its electrical service prior to beginning the installation of your new monitor. The installation of your monitor is separated into two segments, the alarm module and the power supply assembly. Installation is as follows: Alarm Module 1.
Chapter 2: Installation 6. Locate the alarm module in the electrical cutout in the corner cover or side panel of your hood, being careful not to pinch either the electrical wiring or sensor tube. Lock the bottom of the alarm module in place by sliding the bottom edge of the assembly down over the edge of the cutout.
Chapter 2: Installation Protector Hoods 115 Volt: Locate the black wire, which connects from the incoming AC power (in the junction box) to the hood light switch. Disconnect this black wire from the light switch terminal and connect the solid black wire in wiring harness #48610.
Chapter 2: Installation unused lower switch terminal located near the center of the blower switch. Next, connect the black/white wire, which is a part of wiring harness #48610 to the remaining blower switch terminal. Next connect the white wire from wiring harness #48610 to the terminal of the supplied white wire labeled #48616.
Chapter 2: Installation part of wiring harness #48610. This black wire will have two extra terminals attached – one of these terminals should be reconnected to the light switch terminal. The other terminal should be connected to the unused blower switch terminal located near the center of the blower switch.
Alarm Activation The audio and visual alarm will activate approximately six seconds after an alarm condition is detected. To temporarily mute the audible alarm, press and release the test/reset button. NOTE: After an alarm condition has been detected, the red light will stay on. The audible alarm will remain muted until airflow returns to normal levels.
Constant Audible Alarm Check airflow and calibration. If problems persist with the Guardian Jr. Airflow Monitor, contact Labconco Product Service directly at 1-800-522-7658, for assistance. Product Service 1-800-522-7658...