CO2 Incubator – 29 cu ft. Instruction Manual Cat. CH3850...
Page 2
Preface Covered in this manual Catalog # Model # CH3850 3850 Manual Number 7003850 3/3/14 Original ECR/ECN Date Description Cardinal Health Reach-In Incubator...
Page 3
Cardinal Health makes no representations or warranties with respect to this manual. In no event shall Cardinal Health be held liable for any damages, direct or incidental, arising out of or related to the use of this manual.
Page 4
4 Always use the proper protective equipment (clothing, gloves, goggles, etc.) 4 Always dissipate extreme cold or heat and wear protective clothing. 4 Always follow good hygiene practices. 4 Each individual is responsible for his or her own safety. Cardinal Health Reach-In Incubator...
Scroll for Parameters Arrows - Moves the operator through the choic- es of the selected mode. Inject Indicator - Lights when CO is being injected into the incu- bator. % CO2 Display - Displays CO percentage continuously. Reach-In Incubator Cardinal Health...
Section 1 Installation and Start-Up The Model 3850 Series Reach-In Incubator has four basic modes that Keypad Operation allow incubator setup: Run, Setpoints, Calibration and System Configuration. • Run is the default mode during normal operation. • Set is used to enter system setpoints.
(30 PSIG max.) to 3/8” ID hose Back of Unit Chamber Drain (Overflow) - 3/8” elbow with 3/8” hose barb fitting. Install with fitting sloping slightly downward. Back View Floor Figure 1-7. Top View of Fitting Connections Cardinal Health Reach-In Incubator...
With the water turned on, a float valve inside the incubator regulates the water level in the humidity reservoir. When the reservoir is full, a 10- second audible tone is heard and the ADD WATER message no longer displays. Reach-In Incubator Cardinal Health...
Position the incubator so the unit can be easily disconnected. Plug the provided power cord into the power inlet connector on the back of the cabinet, then into a grounded, dedicated electrical circuit. Cardinal Health Reach-In Incubator...
0-2000 psig range. The low pressure gauge, at the incubator inlet, should have a 0-30 psig range. Input pressure to the incubator must be maintained at 15 psig (103.4 kPa), ±5 psig. Caution Each tank should be connected to one incubator only. Reach-In Incubator Cardinal Health...
5. Press the Mode key until the Run indicator lights for Run mode or press the right/left arrow keys to go to next/previous parameter. Caution Any equipment placed inside chamber must be rated for unit operating temperature and humidity. Cardinal Health Reach-In Incubator...
All Model 3850 Series incubators are equipped with a secondary temperature monitoring system to monitor the air temperature inside the cabinet. This system is designed as a safety device to turn off all heaters in the event of a temperature control failure.
4. Press Enter to save the setpoint. 5. Press the Mode key until the Run indicator lights to go to Run mode, or press the right/left arrow keys to go to next/previous parameter. Cardinal Health Reach-In Incubator 1-11...
Temperature Stabilization Periods Start-Up - Allow 12 hours for the temperature in the cabinet to stabilize before proceeding. Presently Operating - Allow at least 2 hours after the display reaches setpoint for temperature to stabilize before proceeding. Cardinal Health Reach-In Incubator...
Page 21
5. Press the Mode key to return to Run or the right/left arrow to go to next/previous parameter. Calibrating the CO System Model 3850 Series incubators have a CO sensor. The incubator atmosphere is not only affected by the quantity of CO present but also by the air temperature and water vapor present in the incubator atmosphere.
5. Press the up /down arrows to change the display to match the independent instrument. 6. Press Enter to store the calibration. 7. Press the Mode key to return to Run or the right or left arrows to go to the next/ previous parameter. Cardinal Health Reach-In Incubator...
2. Press right arrow until Audible XXX is displayed in message center. 3. Press up/down arrow to toggle Audible ON/OFF. 4. Press Enter to save the setting. 5. Press Mode key to return to Run mode or right/left to go to next/previous parameter. Cardinal Health Reach-In Incubator...
3. Press up/down arrow to change the low temp alarm limit. 4. Press Enter to save the low temp alarm limit. 5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter. Reach-In Incubator Cardinal Health...
Page 26
3. Press up/down arrow to change the low CO alarm limit. 4. Press Enter to save the low CO alarm limit. 5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter. Cardinal Health Reach-In Incubator...
2. Press the right arrow until CO2 RLY XXX is displayed in the message center. 3. Press up/down arrow to toggle the setting ON/OFF. 4. Press Enter to save the setting. 5. Press the Mode key to return to Run Mode or right/left to go to next/previous parameter. Reach-In Incubator Cardinal Health...
3. Press up/down arrow to change the span number to match the sticker. 4. Press Enter to save the setting. 5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter. Cardinal Health Reach-In Incubator...
4. Press Enter to save the setting. 5. Press the Mode key to return to Run mode or right/left to go to next/previous parameter. Note Verify that the ADD WATER alarm is set to ON after enabling the RH HIGH feature. Reach-In Incubator Cardinal Health...
Alarms Section 4 The Model 3850 Series incubator is equipped with a system that notifies the user of an alarm condition inside the incubator. All alarms display in the control panel message center. The following table contains information on all possible system alarms.
Section 4 Alarms Sensor Fault Alarms The microprocessor in 3850 Series incubators continually scans all available sensors to ensure that they are operating properly. Should an error be detected, the incubator will sound an alarm and display the appropriate message. If such an alarm occurs, contact your local distributor or the Technical Services Department.
*Change CO filter (under normal conditions) Check operation of the float assembly and water level switch Replace the AquaTec water cleaning cell, if applicable, every six months. * Qualified service personnel Cardinal Health Reach-In Incubator...
Caution Do not use strong alkaline or caustic agents. Stainless steel is corrosion resistant, not corrosion proof. Do not use solutions of sodium hypochlorite (bleach) as they may cause pitting and rust. Cardinal Health Reach-In Incubator...
Note On reassembly, brass overflow fitting must be checked for leaks. Caution Use of chlorinated water, or decontamination products containing chlorine, will deteriorate the stainless steel and cause rust, voiding the warranty. Reach-In Incubator Cardinal Health...
This electrical circuit is protected by a circuit breaker. However, to avoid this contingency if these conditions occur, open the door of the unit and allow it to stand open until the water vapor disperses. Cardinal Health Reach-In Incubator...
The contacts will trip on a power outage or an overtemperature condition. The contacts may also be programmed to trip or not trip on temperature alarms and CO alarms. See Section 3, Configuration Mode. Figure 6-1. Remote Alarm Outputs Cardinal Health Reach-In Incubator...
Cardinal Health Large Capacity Reach In CO2 Incubator Warranty for United States and Canada Laboratory equipment manufactured by Cardinal Health (hereinafter referred to as “the Company”) is warranted as only as stated below. The Warranty Period starts one week from the date your equipment is shipped from our facility. This allows shipping time so the warranty will go into effect at approximately the same time your equipment is delivered.