  • Page 1 How to operate. Installation & maintenance RAIR-01 Version 6. Approved 23/02/2022 from the makers of...
  • Page 2 PRODUC T CONT ENTS Rediair decontamination device, pre-filter, main combination filter (HEPA 14), instruction manual Rediair decontamination device Pre-filter Main combination filter (HEPA 14)
  • Page 3: Table Of Contents

    CONT ENTS Introduction Product description Intended use Limitations of use Rediair’s 3 stages of decontamination Universal labels, symbols and icons Warnings and safety precautions How to use Rediair Setup LCD touch control panel Disposal, cleaning and maintenance External surface cleaning Filter changing and cleaning Cleaning a pre-filter Cleaning and changing times...
  • Page 4: Introduction

    Always read before commissioning. Please carefully follow the instructions in the operating manual. Non- observance will invalidate the warranty. GAMA Healthcare accepts no liability for any damage or consequential damage arising from this. Product description • Rediair is a mobile air decontamination...
  • Page 5: Rediair's 3 Stages Of Decontamination

    20 nanometres. quick time. The activated carbon absorbs the substances, and when the filter is saturated, the filter must be replaced. GAMA Healthcare recommends replacement every six months. In case of strong odour pollution, a more frequent change is advisable.
  • Page 6: Universal Labels, Symbols And Icons

    INTRODUC TION Universal labels, symbols and icons SYMBOL D ESCRIP TION Distributor Caution Consult accompanying documents Warning Indicates a hazardous situation, which if not avoided, could result in injury to the operator Consult instructions for use AU Regulatory Compliance Mark CE marked product (European Conformity mark) UK Conformity Assessed Do not dispose of in general waste.
  • Page 7 Maximum altitude Double insulated device Handle with care Keep dry Temperature limitation Atmospheric pressure limitation Humidity limitation...
  • Page 8: Warnings And Safety Precautions

    (Only use Filter (RAIR-S01) or humidity, direct sunlight, dust etc. supplied by GAMA Healthcare Ltd.) • If the device accidentally becomes wet, NOT E discontinue use until dry and then test the device for proper operation before reuse.
  • Page 9: Setup

    HOW TO U SE SE TUP Before getting started and switching on the Rediair, please remove all packaging. The Rediair filters must be removed from their Unbox the Rediair and remove protective packaging. all outer packaging. Remove the Remove the blue tape on the external covers external covers of the device.
  • Page 10 Remove the Re-install the main filters plastic wrap (both sides), making sure that from the filters. the tabs are facing outwards. Re-install the external covers (both sides). Note: Ensure external covers are fitted correctly Re-install the pre- or the Rediair will filters (both sides).
  • Page 11: Lcd Touch Control Panel

    HOW TO U SE LCD TOUCH CONTROL PANEL 3.1 - Power on/off • 3.3.3 - Night mode: Rediair will operate To switch Rediair on/off, press the power button. Rediair will start in automatic mode. at its quietest level. All lights and displays In automatic mode, the motor speed will be are turned off and the fan runs at its controlled by a sensor, which is automatically...
  • Page 12 HOW TO U SE LED AIR QUALIT Y INDIC ATOR DISPL AYS (3.7) PM2.5 VALUE (µg/m ) 1 - 35 36 - 75 76 - 150 151 - 500 AIR QUALIT Y Very good Good Not good Blue Green Amber DISPL AY COLOUR 3.6 - Filter indicator 3.7 - Air quality indicator...
  • Page 13 LE D AIR Q UALIT Y INDIC ATOR Very good Good Not good...
  • Page 14: Disposal, Cleaning And Maintenance

    DISPOSAL, CLE ANING AND MAINT ENANCE External surface cleaning Filter changing and cleaning Before performing any filter change, Changing the main Rediair filter is clean/decontaminate the external displayed by the filter indicator button surface of the Rediair device. When located on the digital display. cleaning the exterior, remember to: Pre-filters should be cleaned •...
  • Page 15: Cleaning A Pre-Filter

    Cleaning a pre-filter Cleaning and changing times • The surface of the pre-filter can be • Pre-filters are to be cleaned twice vacuumed with a HEPA filter if available. a month and replaced twice a year. • The pre-filter can also be rinsed with •...
  • Page 16 The pre-filters can be rinsed with clean water. The pre-filters must be completely dry before being used again as any remaining moisture could potentially cause damage to the HEPA 14 filter. To clean AUTO PM2.5 µg/m FILTER POWER TIMER MODE LOCK SPEED 1h 2h...
  • Page 17: Disposal

    DISPOSAL, CLE ANING AND MAINT ENANCE CLE ANING AND CHANGING TIM E S External cover Pre-filter Main filter Clean all external Cleaned twice a Replaced twice surfaces frequently month and replaced a year twice a year Disposal Decommissioning Used Rediair filters can be disposed of Rediair has been designed to operate following local waste guidance policies.
  • Page 18: How To Change The Main Filters

    AUTO PM2.5 µg/m POWER FILTER TIMER MODE SPEED LOCK 1h 2h 4h 8h AUTO PM2.5 µg/m FILTER POWER TIMER MODE LOCK SPEED 1h 2h 4h 8h When filter indicator (3.6 on control panel) lights up continuously red, it is time to change the Switch off device and disconnect pre-filters and main filters.
  • Page 19 Remove the Install the new main filters plastic wrap (both sides), making sure that from the filters. the tabs are facing outwards. 5.10 Re-install the external covers (both sides). Note: Ensure external covers are fitted correctly Install the or the Rediair will new pre-filters not power on.
  • Page 20: Appendix

    APP ENDIX T ECHNIC AL SP ECIFIC ATIONS AIR FLOW R AT E MA X. (M³/ H) C ADR (CLE AN AIR D ELIV ERY R AT E) (M³/ H) Range: 200 – 600 AIR FLOW R AT E WITH H14 FILT ER MA X. (M³/ H) LED AIR QUALIT Y INDIC ATOR DISPL AYS 1 - 35 = Very good (blue)
  • Page 21 Cleaning/Disinfecting NOT E The Rediair device should be cleaned after Use only the cleaning each use (or more frequently, as needed). agents recommended. The following cleaning agents may be used Damage occurring through on external surfaces only: decontamination of the device using agents other than those cited in this CHEMIC AL/ FLUID CONCENTR ATION...
  • Page 22: Declaration Of Conformity

    RAIR-AUS D ESCRIP TION HEPA-14 air purifier GAMA Healthcare Ltd. declares that, if fitted, used and maintained in accordance with the product manual, the above-mentioned product(s) conform to the following EU legislation: • Low Voltage Directive (2014/35/EU) • Electromagnetic Compatibility (EMC) Directive (2014/53/EU) •...
  • Page 23: Contact

    Contact your local distributor for technical support. About GAMA Healthcare GAMA Healthcare is at the forefront of infection prevention technology. Founded in 2004 by two NHS doctors, we develop infection prevention products to help save and improve lives. Today, our Clinell, Carell and Contiplan product ranges protect people in healthcare, businesses and homes across more than 60 countries.
  • Page 24 Scan QR code for more information GAMA Healthcare Ltd., The Maylands Building, Maylands Avenue, Hemel Hempstead, Hertfordshire, HP2 7TG, UK. Tel: +44 (0)20 7993 0030 from the makers of E-mail: info@gamahealthcare.com RAIR-01 Version 6. Approved 23/02/2022 JBN220094...