7 Maintenance and cleaning

G requires no maintenance within the specified operating time. The outside of the filter
®
The Germlyser
can be cleaned using common surface disinfectants.
G, a small amount of unfiltered water will flow out. Over time, this will lead
When replacing the Germlyser
®
to deposits on and in the adapters. Aqua free recommends replacement of the adapters after a period of
use every 6 to 12 months.
In the case of a deviation from the specifications or a damage to the packaging, the product should be
returned to Aqua free, along with a Defect Report (form available to download on the service page of the
Aqua free website www.aqua-free.com).

8 Storage

filter must be properly stored in a dry and hygienic environment at a temperature between
The Germlyser
®
10 and 50°C.

9 Return

reusable filters must be sent back to Aqua free within the agreed time for reprocessing
®
The Germlyser
(form downloads at the service page under www.aqua-free.com).
Aqua free Solutions Ltd.
The Media Centre
7 Northumberland Street
Huddersfield, HD1 1RL, Great Britain
E-Mail:
info@aqua-free.co.uk
+44 (0)1484 483045
Phone:
Headquarters:
Aqua free GmbH
Winsbergring 31
22525 Hamburg
Germany
+49 (0)40-46 89 99 90
Phone:
+49 (0)40-46 89 99 99
Fax:
info@aqua-free.com
www.aqua-free.com
Instructions for use
Germlyser
®
G
Table of contents

1 Product

5 Disassembly of used filter / filter change
6 Accessories
7 Maintenance / cleaning
8 Storage
9 Return
1 Product

1.1 Product description

The Germlyser® G is a
-marked, point-of-use, sterile filter for manual final rinsing of endoscopes to
prevent bacterial nosocomial infectious diseases caused by Pseudomonas aeruginosa, Legionella
pneumophila, Stenotrophomonas maltophilia or Staphylococcus aureus among immunosuppressed and
immunocompromised patients.

1.2 Technical data

Size:
L 77 mm x Ø 55 mm
Retention:
7 log steps Brev. dim.
Pore size:
0.2 µm
Operating time*:
4 weeks
Chlorine resistance**: <10 ppm
* The operating time can be increased up to 8 weeks if sufficient hygienic measures exist and an adequate
hygiene plan is carried out. We recommend monitoring by appropriate clinical microbiological testing.
** Continuous dosage of ≤ 10 ppm over operating time; short-term (1h) high dosage (400,000 ppm)
for chemical disinfection.
*** The Germlyser
®
G is not suitable for connection to low-pressure systems (e.g. boilers).
Flow rate:
Adjustable up to 6 l/min
at 5 bar, at 25°C
Max. operating pressure***:
5 bar at 60°C
Max. operating temperature:
60°C (at 70°C < 10 min.
over operating time)
Connection:
Quick-release coupling
www.aqua-free.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Germlyser G and is the answer not in the manual?

Questions and answers

Summary of Contents for Aqua Free Germlyser G

  • Page 1 In the case of a deviation from the specifications or a damage to the packaging, the product should be Instructions for use returned to Aqua free, along with a Defect Report (form available to download on the service page of the Aqua free website www.aqua-free.com).
  • Page 2 -labelled medical device that is made available to the ® The Germlyser contamination by splashing customer at a fixed rental rate. Reprocessing is carried out exclusively by Aqua free. - Daily, wipe down the filter outlet with a common surface disinfectant 2 Certificates / Declaration of conformity The company Aqua free GmbH is certified according to DIN EN ISO 13485.

Table of Contents