Bante Instruments BI-650 User Manual

Bante Instruments BI-650 User Manual

Industrial conductivity controller

Advertisement

Quick Links

BI-650 Industrial Conductivity Controller
USER MANUAL
Bante Instruments Inc.

Advertisement

Table of Contents
loading

Summary of Contents for Bante Instruments BI-650

  • Page 1 BI-650 Industrial Conductivity Controller USER MANUAL Bante Instruments Inc.
  • Page 2: Installation

    Selectable mounting positions. Installation Safety Warning • BI-650 meter shall be installed and operated only in the manner specified in this user manual. • Only skilled, trained or authorized person should carry out installation, setup and operation of meter.
  • Page 3: Meter Overview

    Overview (−) DC 24V (−), linear power supply 2 DC 24 Connection Relay resting position (NC2) Relay working position (NO2) Relay common (COM2) COM2 (−) DC 24V , linear power supply 2 Relay resting position (NC1) Relay working position (NO1) DC 24V , linear Relay common (COM1)
  • Page 4: Meter Setup

    Analog Output (High) Meter Setup Setting range: 20.00 mS/cm to 0.02 µ S/cm The BI-650 meter contains an integrated setup menu for customizing (default 20.00 mS/cm) the displayed option to meet measurement requirement. The following table describes the functions of each menu item.
  • Page 5: Temperature Compensation

    Temperature Compensation Conductivity Calibration The BI-650 meter is supplied with an industrial conductivity electrode. The BI-650 meter allows 1 to 3 points conductivity calibration. For better When the wires of electrode are connected to the meter, the display. accuracy, we recommend to perform 3 points calibration or select a will show ATC icon immediately.
  • Page 6: Multipoint Calibration

    Calibration and Measurement Communication The BI-650 meter uses a standard Modbus-RTU protocol. All of the data are character type (2 bytes). The response data ranges between -32767 to 32767, hexadecimal. PC Command Definition Length of Byte Data 1.6 Press the Enter key, the meter begins the calibration.
  • Page 7: Troubleshooting

    Appendix 0 × 0009 - 20.00 mA correspond to conductivity value 0 × 000A- Position of decimal point for above value (20.00 mA) Appendix 0 × 000B- Measurement unit of above value (20.00 mA) Decimal Point Response: Troubleshooting 0 × 0000 (Reading × 1) 0 ×...
  • Page 8: Meter Specifications

    Conductivity standard solution 1413 µ S/cm, 480 ml ECCS-1288 Conductivity standard solution 12.88 mS/cm, 480 ml ECCS-1118 Conductivity standard solution 111.8 mS/cm, 480 ml Meter Specifications Model BI-650 Conductivity Range 0.01 µ S/cm to 200.0 mS/cm Resolution 0.001, 0.01, 0.1, 1 Accuracy ±...
  • Page 9: Warranty

    Tel: (+1) 346-762-7358 E-mail: banteinstruments@yahoo.com Factory: F3, Building 2, No.2185, Laifang Rd., Shanghai 201615, China Tel: (+86) 21-6404-1598 E-mail: banteinstrument@hotmail.com The information in this document is subject to change without notice.  www.bante-china.com Copyright © Bante Instruments Inc, 2022. All rights reserved.