99 Washington Street 8/05 Form #270 Melrose, MA 02176 Phone 781-665-1400 OPERATING Toll Free 1-800-517-8431 INSTRUCTION Visit us at www.TestEquipmentDepot.com MODEL 3005 DIGITAL INSULATION-CONTINUITY TESTER...
CONTENTS 1. SAFTY PRECAUTIONS 1. SAFETY PRECAUTIONS ······································································· 1 ○ The instrument is designed to produce in accordance with the following 2. FEATURES ···························································································· 4 standards and supplied in the best conditions after successfully clearing 3. SPECIFICATIONS ················································································· 5 the inspection. 4.
Page 3
DANGER is reserved for conditions and actions that are likely to cause CAUTION ● Always make sure to set the function switch or range selector switch to serious or fatal injury. WARNING is reserved for conditions and actions that can cause the appropriate position before taking measurements.
2. FEATURES 3. SPECIFICATIONS MODEL-3005 are microprocessor controlled insulation-continuity testers. ● Measuring Range and Accuracy (at 23±5°C, relative humidity 45 -75%) ● Designed to safety standards: ○ Insulation Resistance Ranges: IEC 61010-1 Overvoltage CAT III 300V Contamination Degree 2 Nominal Output Voltage...
Page 5
● Typical Number of Measurements. IEC 60529 (IP54) Dust & drip proof (central tendency for supply voltage up to 8V) ● Operating System: Dual integration Insulation Resistance Ranges: Approx. 1000 times min. at load 0.5MΩ ● Display: Liquid crystal display (maximum count: 1999), Continuity Ranges: Unit, Mark Approx.
● Dimensions: 185(L) x 167(W) x 89(D) mm approx. 4. INSTRUMENT LAYOUT 4-1 INSTRUMENT LAYOUT ● Weight: 900g approx. (including batteries) ● Power Source: 8 x R6P, 1.5V AA or equivalent ⑧ ④ ● Auto-power-off Function: Automatically turns off approx. 10 minutes after ① ...
5-1 Removing the Cover 6-1 Disconnection and check of power source of the circuit under test MODEL - 3005 have a dedicated cover to protect against an impact from DANGER the outside and prevent the operation part, LCD, and connector socket ●...
6-2 Insulation Resistance Measurement ④ Read the resistance value from the LCD. DANGER ● Always test the circuit or equipment to ensure it is surely de-energized before measurement according to the instruction of 6-1. ● To avoid electrical shock, measurements must be performed on de- energized circuits only.
●Terminal connection of insulation resistance test ● Principle of Continuity Measurement (Resistance Test) In case of testing insulation of insulated wire and cable against the earth Resistance value can be obtained by applying a certain current to the at direct current, connecting - pole of power to cable conductor, + to the resistance under test and measuring the voltage generated on the both earth obtains smaller measuring value, compared with connecting the sides of the resistor under test.
7. FUNCTIONS 8. BATTERY & FUSE REPLACEMENT DANGER 7-1 AUTO NULL ● Never open the battery compartment cover while making measurement. When conducting continuity tests, the contact resistance of test probes etc. To avoid possible electrical shock, disconnect the test probe before is automatically subtracted before the real resistance is displayed to obtain opening the cover for battery and fuse replacement.
9. CASE AND STRAP BELT ASSEMBLY 10. CLEANING OF THE INSTRUMENT By hanging the instrument around the neck, both hands can be used freely ● When cleaning the instrument, wipe it with a silicon cloth or soft cloth to for easy and safety working. remove dust or dirt.