Progressive Dynamics LIFE-AIR 1000 Service Manual

Hypothermia therapy system

Advertisement

Quick Links

SERVICE MANUAL
MODELS
PD2580, PD1000, PD1000-S, PD1000-1 & PD1000-1S
507 Industrial Rd.
Marshall, Mi 49068
616.781.4241
fax 616.781.7802
www.progressivedynamicsmedical.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the LIFE-AIR 1000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Progressive Dynamics LIFE-AIR 1000

  • Page 1 SERVICE MANUAL MODELS PD2580, PD1000, PD1000-S, PD1000-1 & PD1000-1S 507 Industrial Rd. Marshall, Mi 49068 616.781.4241 fax 616.781.7802 www.progressivedynamicsmedical.com...
  • Page 2 This equipment contains static sensitive components. Proper static handling procedures and equipment must be used when servicing this equipment. ® Perform a safety inspection after making repairs to the LIFE-AIR 1000 and before returning the unit to service. A safety inspection procedure is included in this manual.
  • Page 3 When this happens the flashing lamps and the alarm should shut off. ® There are 2 types of Time-out circuitry used in the LIFE-AIR 1000 , each requires different testing procedures. The procedure to be used can be determined by looking into the air outlet on the top of the unit.
  • Page 4 Safety Inspection Procedure 5. Time-out test procedure #2: a. Connect the hose to the Warm Air Port and to the Hose Storage Port. b. Set the Heater Control Switch to the "100" setting. c. Turn the unit on and allow to stabilize at the "100" setting. d.
  • Page 5 INSTRUCTIONS FOR REMOVAL OF ASSEMBLIES CAUTION: THE PLENUM AND FRONT PANEL ASSEMBLY CONTAIN STATIC SENSITIVE DEVICES. PROPER HANDLING PRECAUTIONS MUST BE USED. PD2580, PD1000/PD1000-S 1. Unplug unit from 120 VAC line. 2. Remove the Rear Filter Cover. 3. Remove the filter element 4.
  • Page 6 BLOWER AND PLENUM ASSEMBLY 1. Support the Blower and Plenum Assembly while removing the 4 screws from top hose interface. 2. Unp lug the 3 pin har ness fro m the te mpera ture sen sor to the Front P anel C ontrol. Some units may have two 3 pin harnesses to be removed.
  • Page 7 Special Instructions for Installation of the following Assemblies: Temperature Controller & Faceplate: When replacing this assembly it is required that a bead of clear Silicone Sealant be placed around the perimeter of the faceplate to provide a seal. This Assembly is STATIC SENSITIVE.
  • Page 20 CONTENTS Safety Inspection Procedures Self Test ............1 Time-out Test #1 ..........1 Time-out Test #2 ..........2 Assembly Removal Filter and Wire Cage ..........3 Blower and Plenum Assy........4 Front Panel Assy..........4 Special Instructions ..........5 Wire Diagrams PD2580 / PD1000 ..........