Summary of Contents for Geotech GT-HR

  • Page 1 Geotech Science Cooperation Ltd 2F., No.200, Sec 3, Datung Rd, Xizhi Dist New Taipei City 221, Taiwan (R.O.C) Model GT-HR Multi-function Readout Operation Manual Copyright Ⓒ 1993-2018 Geotech Cooperation. All right reserved. Doc Rev 1.2 (2018/03/31)
  • Page 3 • GEOTECH is not responsible for any accident, breakdown or other problems due to lack of knowledge and / or non-compliance with the requirements contained in this manual.
  • Page 5 GT-HR Multi-Functional Readout Package includes...
  • Page 7: Table Of Contents

    Table of Contents Notes for Using the Product ......................1 GT-HR Front Panel ..........................1 Product Introduction ......................... 3 Accessories ............................4 Battery ..............................5 How to Charge the Battery ......................5 Taking Care of Lithium Battery ....................... 6 Operational Instructions ........................7 Front Panel Controls .........................
  • Page 8: Notes For Using The Product

    GT-HR Multi-Functional Readout Notes for Using the Product For safety and convenient operation, please carefully read this user manual before using the product. GT-HR Front Panel Display Charge Light Power Light Keypad...
  • Page 9 GT-HR Multi-Functional Readout Label Battery Housing “Sensor” Connector On/Off Switch Micro USB Charger Connector...
  • Page 10: Product Introduction

    GT-HR could generate DC and AC excitation voltages, which could be set at 2.5V, 5V to 12V. The output voltage reading of sensor is then measured and displayed on the screen.
  • Page 11: Accessories

    GT-HR Multi-Functional Readout Accessories The following accessories are come up with GT-HR: 1. A MicroUSB charging cable (Figure 1) 2. An AC Power Adapter with DC-5V,1A output (Figure 1) 3. A set of alligator cables and ground (Figure 2) Figure 1 AC Power Adapter and Micro USB charging cable...
  • Page 12: Battery

    GT-HR Multi-Functional Readout Battery Each unit GT-HR is come with a 1000mA lithium battery. According to the test result, a lithium battery under proper operation has a rated life for 500 or more charging cycles. The battery lifetime is depending on design and quality of the production, normally the capacity will gradually degrade with relative to time, or number of cycling cycles.
  • Page 13: Taking Care Of Lithium Battery

    If the battery will not use for a long time, for example 2-3 years, it would be better to remove it from GT-HR The current battery capacity of GT-HR is displayed in percentage when “measurement” option is selected. The readout has an internal IC which controls the flow of electrical current during re-charging.
  • Page 14: Operational Instructions

    GT-HR Multi-Functional Readout Operational Instructions Every time start up the GT-HR, the readout will display manufacture logo “Geotech” for about 3 seconds, before entering the main menu. Front Panel Controls 1. “Up” and “Down” Keys § At the main menu display option: Choose measurement item.
  • Page 15: Key Note

    GT-HR Multi-Functional Readout 5. Key Note § Press “Enter” key, for exit measurement display and back to main menu § Please carefully notice the excitation voltage for connected sensor. For DC and AC voltage measurement, user has to select the excitation voltage prior to measurement.
  • Page 16: Main Menu

    GT-HR Multi-Functional Readout Main Menu There are 5 options inside main menu as shown in Figure 3 and Figure 4 below: 1. DC Voltage Measurement (Electronic) 2. AC Voltage Measurement (Exchange) 3. Vibrating Wire Measurement (VW Sensor) 4. Current Measurement (4-20 mA) 5.
  • Page 17: Dc Voltage Measurement

    GT-HR Multi-Functional Readout 1. DC Voltage Measurement § Typical sensor for this type of measurement including LVDT and IPI § For Single-End Type: Connect Yellow alligator cable to sensor power input wire (Yellow); Connect Black alligator cable to sensor common ground wire (Black);...
  • Page 18 GT-HR Multi-Functional Readout Figure 5 Selecting Sensor Excitation Voltage Figure 6 Measurement Display. § At the top of measurement display, user could see the sensor type as well as excitation voltage...
  • Page 19: Ac Voltage Measurement

    GT-HR Multi-Functional Readout 2. AC Voltage Measurement § Typical sensor for this type of measurement including ETI and EL-Beam § Connect Yellow alligator cable to sensor power input wire (Yellow); Connect Black alligator cable to sensor common ground wire (Black); Connect Green alligator cable to sensor signal output wire (Green);...
  • Page 20: Vibrating Wire Type Measurement

    GT-HR Multi-Functional Readout 3. Vibrating Wire Type Measurement § Typical sensor for this type of measurement including Strain Gage and VW Piezometer. § Connect White alligator cable to sensor Red wire; Connect Black alligator cable to sensor common ground wire (Black wire); Connect Green alligator cable to sensor Green wire.
  • Page 21 GT-HR Multi-Functional Readout Frequency POS Instrument Name Module Formula Units Range(Hz) F2×10-3 Digits All vibrating wire 450~6000 Hertz sensor μS μSec Rebar Stressmeter、 F2×10-3 Digits Piezometer、 Hertz 1200~3500 Extensometer、 μS μSec Load Cell μE F2×10-3×4.062 μStrain(ε) Arc Weldable Hertz 450~1200 Strain gage μS...
  • Page 22: Current Measurement

    GT-HR Multi-Functional Readout 4. Current Measurement § Typical sensor for this type of measurement including DSM. § Connect Yellow alligator cable to sensor power input wire (Yellow); Connect Black alligator cable to sensor common ground wire (Black wire); Connect Green alligator cable to sensor signal output wire (Green).
  • Page 23: Language Selection

    GT-HR Multi-Functional Readout 5. Language Selection In the current GT-HR version, the display is available in two selected languages (English and Traditional Chinese), as can be seen in Figure 11 to Figure 13 below. Figure 10. Language Selection (English and Traditional Chinese)
  • Page 24 GEOTECH GT HR MULTI FUNCTION HANDHELD READOUT This document contains information that is proprietary to Geotech and is subject to return upon request and may not be reproduced or copied in any form, by photocopy or any other means, including disclosure to outside parties, directly or indirectly, without permission from Geotech.