Page 1
8.5 inch (width) x 11 inch Color Graphics on White Background document should not appear Booklet 20# Bond or Equivalent on produced Labels or IFUs Description/Type: Instructions For Use Stryker Instruments Part Number: Rev. (269) 323-7700 (800) 253-3210 0940-001-025 DSGN-fm-50366 Page 1 of 1...
Page 2
Cast Cutter 0940-000-000 0941-000-000 Instructions For Use ENGLISH (EN) 2016-10 0940-001-025 Rev-L www.stryker.com...
Indications For Use medical devices should install, inspect, service, and test this equipment as described in this manual. The Stryker Cast Cutter is intended to section and remove synthetic and ▪ DO NOT modify any system component or accessory, including the power plaster casts.
Page 4
Standby Mode immunity of the system. NOTE: For a complete list of accessories, contact your Stryker sales Standard Speed Mode representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary.
Page 5
0940-001-025 Rev-L Features Model REF 0940-000-000 Only: Power Cord Plug Cast Cutter Model REF 0941-000-000 Only: Power Source Cord Plug Power Source Cord Appliance Inlet Connector Plug Appliance Inlet Connector Function Switch High Speed Mode – The Cast Cutter blade will oscillate at high speed.
Page 6
Cast Cutter to another person. ▪ DO NOT apply excessive pressure, such as bending or prying, with the blade. Excessive pressure may cause the blade to fracture. If prying on the cast is required, use a cast spreader tool. www.stryker.com...
Page 7
2. Pull the blade off of the bottom two blade guide posts. 3. Slide the blade up behind the screw and remove the blade. 4. Save or dispose of the blade as required. www.stryker.com...
Stryker-authorized service personnel only. 2. If the vacuum hood is blocked with debris, perform the following steps: ▪ For service, contact your Stryker sales representative or call Stryker 2. 1. Use a standard slotted screwdriver to remove the captive screw, customer service.
Contact local 2. Loosen and remove the brush cap, spring and brush. distributor for disposal information. Ensure infected equipment is 3. Dispose of the used spring and brush. decontaminated prior to recycling. www.stryker.com...
0940-001-025 Rev-L Troubleshooting WARNING: DO NOT disassemble or service this equipment, unless otherwise specified. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM CAUSE SOLUTION The Cast Cutter does...
0940-001-025 Rev-L Specifications Model: Stryker Cast Cutter (REF 0940-000-000) Stryker Cast Cutter (REF 0941-000-000) Electrical: 100-120 V , 50-60 Hz, 1.5 A input 220-240 V , 50-60 Hz, 0.75 A input Speed: 100 V Nominal 120 V Nominal 230 V...
Page 12
0940-001-025 Rev-L Model: Stryker Cast Cutter (REF 0940-000-000) Stryker Cast Cutter (REF 0941-000-000) European Not applicable Conformity: Product Safety CSA International Certification: Canadian Standards Association (CSA) Canadian Standards Association (CSA) CAN/CSA-C22.2 No. 601.1-M90, Medical Electrical Equipment CAN/CSA-C22.2 No. 601.1-M90, Medical Electrical Equipment —...
Page 13
RF compliance level above, the Cast Cutter (REF 0941-000-000) should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating Cast Cutter (REF 0941-000-000). Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. www.stryker.com...
Page 14
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. www.stryker.com...