Thank you for purchasing Torque wrench testing machine. ■The testing machine purpose for use inspection/measurement torque of torque wrench. ■Easy operation by routine laboratory at open close. ■Measurement result can be checked correctly on LED red display. ■Switching by three kinds of measurement mode(Peak/truck/peak-hold). If the manual is lost or becomes illegible, or if additional manuals are required, contact your distributor. If you have any questions about the products or the contents of this instruction manual, contact your distributor. Instructions are classified by degree of risk and described as follows.
Page 21
■The product purpose for use inspection/measurement of torques wrench only. Please use only for intended purposes ●Must be rigthway ○Much load is applied during the mesurement. ○Do not place the machine where it will be install on stable vibration and instability. place. To avoid the risk of personal injury from breakdown of the attached stand during measurement ●Must be used within ○AC100V 〜 240V. ○Electrical captaincy (Ampere) for a wall range of rated voltage. socket is decided To avoid the fire and electric shock. ●Do not apply load more ○Please use within power limit. than the maximum torque. To avoid the risk of personal injury from breakdown ●Do not touch the power plug with wet To avoid the risk of person injury from electric shock. hands. ●Do not touch the power ○Do not touch the electric plug right after use. cable and metal parts of connector . To avoid the risk of person injury from electric shock. ●Please clean the electric plug and the wall socket when they are with foreign To avoid the risk of person injury from substances.
Page 22
●Measurement direction: ○This testing machine does not work rotation Rotation to the right to the left (counter-clockwise) (clockwise rotation) To avoid the risk of person injury and breakdown, damage of testing machine. ●Do not use in the places ○The machine is unable to cope with under the such as water, sea, great environment mentioned humidity, or near oil, left. Leaking, rusting chemicals, solvent. worsens the performance and accuracy of the machine. To avoid the risk of person injury and breakdown, damage of testing machine. ●Please do not vibrate ○Torque Checker is a sensitive measuring instrument. or shock that machine It can cause breakdown of the machine or malfanction. ●Can not be used to ○The machine is not for air impacts or measure power tools electric wrenches. ITo avoid the risk of person injury and breakdown , damage of testing machine. ●Do not dismantle and Prohibition of dismantle. remodeling. The performance or the life of the machine declines, which causes breakdown or injury.
Page 23
●Pull out AC adapter ○Leaving for a long time can heat up the condenser. ○Do not pull the code when unplug from outlet. from wall socket when it is not used for a long time. To avoid the fire and electric leakage. ●Use standard accessory ○Use attachements described in the by maker-classified . instruction manual. To avoid the risk of person injury and accident ●Do not place the machine where it will be strong magnetism and near magnetism. The performance goes bad. ●Do not use around ○The signals or noises of electric circuit can worsen the performance. electrical equipment's in operation. It can cause breakdown or injury. ●Do not use in a poor ○Using in a poor carriage can harm your health. Use in a good carriage. carriage. ●Please keep the working ○It is possible to slide when the floor surface is wet with oil etc. floor always clean. It can cause injury. ●Please take care of ○Remove dust, dirt, mud, oil, water to avoid breakdown or rusting after use. the machine.
Page 24
Contents Products Number TA419TQ Number Contents Main body ○ 1 AC Adapter Set TS419TQ−AD ○ 1 Hexagon sockt head ○ 4 Hexagon sockt head cap screws(M10×45) cap screws Set Nut(M10) ○ 4 TS419TQ−BT Washer ○ 4 TA419TQ−204 (CA0204) ○ 1 (Square internal 6.35×Square external 12.7) TA419TQ−308 (CA0308) ○ 1 (Square internal 9.5×Square external 25.4) TA419TQ−408 (CA0408) ○ 1 (Square internal 12.7×Square external 25.4) TA419TQ−836 (MA8-36) ○ 1...
Name of each part Holes for fixed bolts (φ10/4 parts) Display or Control Panel Square Drive (Internal square) TA419TQ 25.4mm Power Switch (ON/ OFF) Terminal outlet (DC9V, 1.6A, Center Plus) 「Display or Control Panel 」 Display (Red:LED) TA419TQ 10〜200 Zero / Enter Botton Press button when resetting the result displayed to "Zero" or switching of each mode. Reset Button Press when changing the set number (1, 2, 4, 8 seconds) with Peak mode (Auto 「Zero」 Reset). Mode Button Measurement mode ①Track Mode ②Peak Mode ③Peak Hold When press switching.
Page 26
Precaution 1.Prepare stand to fix. ●Choose the place to fix ○Components which are hard and strong, thicker than 6mm and install it firmly. are needed. ○Much load is applied to the machine and installed area. ○Avoid setting on an unstable stand, a slope and shaking place. To avoid the risk of personal injury from breakdown of the attached stand during measurement 2.Please drill holes on the fixed stand. 172mm Because the checker uses φ10.5〜φ11 attached hexagon socket head cap screws, drill a hole on the stand as illustration. φ10.5〜φ11(recommended) 3.Fix tightly with 4 haxagon socket head cap screws. Hexagon sockt head cap screws(M10×45) Hexagon internal width across flat: 8mm ※Prepare a 8mm-hex key wrench Components should be ①Firm and durable ②Thicker than 6mm Nut: M10 Hexagon external width across flat: 14mm ※Prepare a spanner or a offset wrench which is width of 14mm...
Page 27
Precaution 4.Please insert AC adapter Insert AC adapter into the connection terminal at back of the Checker after confirming power switch is OFF . To a wall socket ●Pull out AC adapter from ○Pull out from a wall socket when not used. Do not pull the code when pulling out. The code will be damaged wall socket for safety and can cause electric leakage. when it is not used for Leaving for a long time can heat up the a long time. condenser and will be cause of the fire. Instruction 1.Turn on the power switch 「OFF」 Turn on the power switch after inserting AC adapter into a wall socket. The process from starting to measurement display is same as 「ON」 illustration below. It takes about 4 seconds. 「Common to all models」 「Models Number」 「Mesurement display」...
Page 28
Instruction 2.Select to the operation mode. Types of operation mode. Peak Mode ● Auto 「Zero」 Reset will be activated after Peak Hold in set seconds(Choose 1, 2, 4, 8 seconds) ●Peak Hold Measurement result will be displayed continually unless the button is pressed after Peak Hold Pressing the button reset to 「Zero」 manually. ●Track Mode 「Zero display」 is up when applying stops, and the result changes depending on added torque. 2−1.Setting Procedure of operation modes Display is switched when the button is pressed(with buzzer ring). TA419TQ 10〜200 Pressing the button, the mode will be changed when buzzer rings. TA419TQ TA419TQ TA419TQ 10〜200 10〜200 10〜200 Measurement mode will be return when pressing button of the operation mode. TA419TQ TA419TQ 10〜200 10〜200...
Page 29
Instruction 2−2.Setting of display seconds by Auto 「Zero」 Reset in Peak mode. Display will be switched by pressing the button(with buzzer ring). TA419TQ 10〜200 Display will be switched each time the button pressed(with buzzer ring). TA419TQ 10〜200 TA419TQ TA419TQ TA419TQ 10〜200 10〜200 10〜200 Please press button to get back to measuring screen after you entered seconds for Auto Reset TA419TQ TA419TQ 10〜200 10〜200...
Page 30
Instruction 3.Measurement methods of each mode 3−1.Measurement method of Peak Mode 3−1−1.Torque tool, preset type Please apply force after checking set ● torque of torque wrench ex) Torque wrench, preset type Set torque : 50N・m Auto 「Zero」 Reset after 8 seconds TA419TQ 10〜200 The result will be displayed with ● clicking sound.However, 「Zero」 reset will be automatically done in the set seconds if you stopped applying force before it reaches the set torque. 「When it clicks」 TA419TQ 10〜200 「When it stops in progress」 TA419TQ 10〜200 「Zero」 Reset will be automatically done in 8 seconds TA419TQ 10〜200...
Page 31
Instruction 3−1−2.Indicator torque tool (Plate-shape, torque wrench, preset type) Please apply force as an example below ● ex)Torque wrench, preset type Set torque : 50N・m Auto 「Zero」 Reset TA419TQ 10〜200 Please keep applying force till the indicator ● reaches 50N・m. However, 「Zero」 reset will be automatically done in the set seconds if you stopped applying force before it reaches the set torque. 「When torque same with set torque applied」 TA419TQ 10〜200 「When it stops in progress」 TA419TQ 10〜200 「Zero」 Reset will be automatically done in 8 seconds TA419TQ 10〜200...
Page 32
Instruction 3−2.Measuring method of Peak Hold 3−2−1.Torque tool, preset type Apply force after the set torque of torque wrench ● ex) Torque wrench, preset type Set torque : 30N・m Manual 「Zero」 Reset TA419TQ 10〜200 The result will be displayed with clicking ● sound. The result will be also displayed at the time when you stopped before it sounds 「When it clicks」 TA419TQ 10〜200 The displayed result will not be changed ● if you apply force over 30.7N・m. 「in case of applying force more」 TA419TQ 10〜200 Numbers will be displayed when you stopped ● applying force 「when you stopped applying force」 TA419TQ 10〜200 It will be reset (with buzzer ● sounds) when you push the button of TA419TQ TA419TQ 10〜200...
Page 33
Instruction 3−2−2.Indicator set torque tool(Plate type, Dial type set) Please apply force after checking set ● torque of torque wrench ex.) Dial set torque wrench Set torque : 30N・m Manual Zero Reset TA419TQ 10〜200 The result will be displayed at the time ● when you stopped before it reaches the set torque. 「When torque same with set torque applied」 TA419TQ 10〜200 The displayed result will not be changed ● if you apply force over 30.7N・m. 「in case of applying force more」 TA419TQ 10〜200 Numbers will be displayed when ● you stopped applying force 「when you stopped applying force」 TA419TQ 10〜200 It will be reset (with buzzer ● sounds) when you push the button of TA419TQ TA419TQ 10〜200 10〜200...
Repair・Inspection ■The machine should be repaired when display is broken, the result is not loaded or incorrect. (Non-free) It depends on the condition product whether it can be repaired or not. ■Rough use, keeping for a long time, ecessive use can make the accuracy bad. Proofreading regularly is necessary to keep precision For recommended calibration in a year (for profit) ■If you need additional assistance for operation Specification 2 0 5 φ 1 0 1 7 2 φ D S ※Not included weight by AC adapter and conversion adapter. Power supply:100V 〜 240V ● Mesurement mode : Peak mode, Track mode ● Zero Reset : Manual Reset, Auto Reset (1, 2, 4, 8seconds : adjustable) ● Measuring direction: Rotation to the right (clockwise rotation) ● ...