  • Page 1 II  Operating instructions...
  • Page 2 2171 Executive Drive Suite 200 Addison, Illinois 60101 Telephone: +1 833 367 6442 E-mail: usa@miha-bodytec.com Internet: www.miha-bodytec.us TD150170 – Rev. 8 This manual was created by: kothes GmbH Internet: www.kothes.com © miha bodytec Inc. 2021 miha-bodytec II EMS training device 2021-07-28...
  • Page 3 Supplemental directives Conscientious handling of the miha bodytec II Conscientious handling is the prerequisite for successful and safe use. The trainer must take part in a training session held by the manufacturer to be able to safely operate the EMS training device, hereinafter referred to as “device”.
  • Page 4 The contents of this manual are protected by copyright. Use of these contents is permissible within the framework of use of the device. Any other use is not permitted without the written permis- sion of the manufacturer. miha-bodytec II EMS training device 2021-07-28...
  • Page 5: Table Of Contents

    2.4 A team for your safety..........19 2.5 Warnings..............20 2.6 Reporting of adverse events........26 2.7 Symbols on the miha bodytec II........27 2.7.1 Symbols on the front..........27 2.7.2 Symbols on the back..........28 2.7.3 Rating plates............29 2.7.4 Cleaning symbols............
  • Page 6 13.1 Symbols on the packaging........105 13.2 Transport and storage..........106 13.3 Environmental protection........106 Ensuring optimum function........... 108 14.1 Maintaining the device..........108 14.2 Maintaining the Electrodes........109 14.3 Handling error messages........109 Index................. 110 miha-bodytec II EMS training device 2021-07-28...
  • Page 7: Overview And Scope Of Delivery

    Overview and scope of delivery Overview and scope of delivery 1.1 Scope of delivery miha bodytec II Fig. 1: Overview of miha bodytec II Ä “Control unit (top side)” Control unit ( Column on page 9) Base plate Power cable...
  • Page 8 With EMS, electrical pulses from outside activate the muscles. It makes no dif- ference to the muscle whether electrical stimuli are sent from the brain or from electrodes: It reacts by contracting. miha-bodytec II EMS training device 2021-07-28...
  • Page 9 14 Main controller for setting the pulse strength All controls with the exception of the on/off switch, which is located on the underside (1), are located on the top side of the control unit. 2021-07-28 miha-bodytec II EMS training device...
  • Page 10 Connection to cable socket USB port Speaker The on/off switch (Fig. 4/1) is embedded in the case on the under- side of the control unit. A built-in terminal strip is located in a depression on the underside. miha-bodytec II EMS training device 2021-07-28...
  • Page 11: Equipment

    ® “i-body “. The miha bodytec USB flash drive can be inserted into the USB port (Fig. 4/4) for performing updates and for saving device settings. A speaker (Fig. 4/5) is located behind the perforations in the case.
  • Page 12 Size 2, size 3 and size V2 in in² 2 electrodes for 22.85 in² 27.56 in² abdomen 2 electrodes for 10.31 in² 12.81 in² ® Fig. 8: i-body chest 2 electrodes for 16.25 in² 20.98 in² upper back miha-bodytec II EMS training device 2021-07-28...
  • Page 13 Table 3: Electrode sizes for each i-body belt size: ® Electrode size Size of the i-body belt Size 1: 2 buttocks electrodes each 14.7 in² ® Fig. 10: i-body belt Size 2: 2 buttocks electrodes each 18.34 in² 2021-07-28 miha-bodytec II EMS training device...
  • Page 14 Pump spray bottle for moistening the electrodes. Fig. 13: Pump spray bottle miha bodytec USB flash drive The miha bodytec USB flash drive is delivered with the device and serves as a data carrier for performing updates as well as saving and transferring settings.
  • Page 15 Overview and scope of delivery In the event of important updates, a miha bodytec USB flash drive will be sent to users of the device who perform the update them- selves or have it performed on the device by the trainers.
  • Page 16: Safety

    II may only be used by persons above the age of 21. miha-bodytec II EMS training device 2021-07-28...
  • Page 17: Misuse

    The irritation can usually be reduced by alter- nate electrode placement using an alternate conductive medium. – Electrode placement and stimulation settings must follow guidance of the prescribing practitioner. 2021-07-28 miha-bodytec II EMS training device...
  • Page 18: Risks Due To Physical Condition: Contraindications

    Safety – The long-term effects of repeated electrical stimulation are unknown. – Only use genuine miha bodytec equipment and spare parts. Misuse of the device can cause hazardous situations. The device is exclusively intended for use as an electronic muscle stimulator for the usecases described in this manual and must only be used in a dry environment.
  • Page 19: A Team For Your Safety

    18) apply to them before they start with the training. Trainer The trainer must have taken part in a basic miha bodytec training session and can prove their qualification by presenting a certificate issued in their name. The participation in an additional higher-level EMS training session is recommended.
  • Page 20: Warnings

    Function, operation, and programming of the control unit Performing updates The trainer applies the information they learned in the basic miha bodytec training session so that everyone training with the device has a safe EMS workout. In particular, the trainer is responsible for the following: Ensuring that updates are integrated promptly following release.
  • Page 21 – Administer CPR until emergency medical services arrive. – If you have access to a defibrillator, defibrillate the ath- lete/patient. The EMS training involves a residual risk which may lead to cardiac arrhythmia. If untreated, cardiac arrhythmia may result in death. 2021-07-28 miha-bodytec II EMS training device...
  • Page 22 Improper routing of connection cables can lead to trip hazards or strangulation of babies and toddlers/children as well as handicapped persons. miha-bodytec II EMS training device 2021-07-28...
  • Page 23 – Keep cable clips in a safe place out of the reach of babies, infants/children, and handicapped persons. Inhaling or swallowing cable clips can lead to injuries. Allergic reaction There are currently no known allergic reactions to materials used by miha bodytec. 2021-07-28 miha-bodytec II EMS training device...
  • Page 24 Danger due to insufficient level of knowledge! L WARNING – miha bodytec II may only be used by persons above 21 years. – Electrodes may only be connected and detached by the trainer or under their supervision.
  • Page 25 – Never switch on the device if condensation is visible. Condensation can form in the event of changes in humidity and temperature. Moisture in the control unit can result in damage or even total destruction. 2021-07-28 miha-bodytec II EMS training device...
  • Page 26: Reporting Of Adverse Events

    ReportsManualsForms/Forms/UCM349464.pdf Follow the instructions on the form to either fax or mail it in for submission. For help filling out the form, see MedWatchLearn: http://www.accessdata.fda.gov/scripts/MedWatchLearn/ Call FDA at 1-800-FDA-1088 to report by telephone. miha-bodytec II EMS training device 2021-07-28...
  • Page 27: Symbols On The Miha Bodytec Ii

    Safety 2.7 Symbols on the miha bodytec II 2.7.1 Symbols on the front Fig. 14: Symbols on the front of the miha bodytec II Symbol Standard Reference Standard Title Symbol Title Explanatory Text ISO 7010 Second Edi- Graphical Symbols –...
  • Page 28: Symbols On The Back

    Safety 2.7.2 Symbols on the back Fig. 15: Symbols on the back of the miha bodytec II Symbol Standard Reference Standard Title Symbol Title Explanatory Text ES60601-1:2005/ Medical electrical General warning General warning sign (R)2012 And A1:2012, equipment – Part 1: sign Table D.2, Symbol 2...
  • Page 29: Rating Plates

    FCC IDs IC IDs Hardware version identification number (HVIN) Mandatory sign: Read manual Mandatory sign: Prohibited for persons fitted with cardiac pace- Fig. 16: Rating plate miha bodytec II, makers example Mandatory sign: Waste of electrical and electronic equipment (WEEE)
  • Page 30 Type BF applied Identifies a type BF (R)2012 And A1:2012, equipment – Part 1: part applied part complying Table D.1, Symbol 20 General require- with IEC 60601-1. ments for basic safety and essential performance. miha-bodytec II EMS training device 2021-07-28...
  • Page 31 Enclosure protection, where... safety and essential N1= degree of protec- performance. tion from particulates (scale of 0 – 6); and N2 = degree of protec- tion from water (scale of 0 – 8) 2021-07-28 miha-bodytec II EMS training device...
  • Page 32: Cleaning Symbols

    Do not tumble dry. 2014-01-15, 3109 For Use On Equip- ment - Registered Symbols 7000 Fifth Edition Graphical Symbols Do not bleach Do not bleach. 2014-01-15, 3124 For Use On Equip- ment - Registered Symbols miha-bodytec II EMS training device 2021-07-28...
  • Page 33: Environmental Protection

    Electronic components, such as liquid crystal display, printed circuit boards and cabling, can contain toxic substances. These must not get into the environment. Disposal must be carried out by a speci- alized disposal company. 2021-07-28 miha-bodytec II EMS training device...
  • Page 34: Technical Data

    Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. miha bodytec II is intended to stimulate muscles in order to improve or facilitate muscle performance. In addition, it is indicated for the following conditions:...
  • Page 35 10 hours down the device from the max- imum permis- sible storage temperature (+158 °F / +70 °C) to the operating tem- perature for intended use of +68 °F (+20 °C) ambient tem- perature 2021-07-28 miha-bodytec II EMS training device...
  • Page 36 Ueff = 15 V at 500 ohm (30 mA) value Assuming a load impedance of 500 ohm, this produces a current intensity level of 30 mA r.m.s. The output voltage does not contain any DC component. Output waveform – bipolar miha-bodytec II EMS training device 2021-07-28...
  • Page 37 The service life of the equipment depends on the functionality during training sessions. life of equipment As soon as the functionality is no longer ensured or the hygienic requirements are no (contact points, longer met, the equipment must be replaced. Velcro fasteners, etc.) 2021-07-28 miha-bodytec II EMS training device...
  • Page 38 Technical data Compliances miha bodytec II medical complies with the applicable requirements of the following international and national standards: IEC 60601-2-10:2012 – Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators IEC 60601-1-2:2014 –...
  • Page 39 Environment The miha bodytec II medical is designed for use in domestic envi- ronments and professional health facilities but not near active sys- tems of high-frequency surgery devices or in high-frequency shielding rooms used for magnetic resonance imaging where high- intensity electromagnetic interferences occur.
  • Page 40 We do not provide technical support for this product if this was changed. miha-bodytec II EMS training device 2021-07-28...
  • Page 41: Electromagnetic Compatibility

    The miha bodytec II EMS training device has been designed for use in electromagnetic environments as outlined below. The buyer or user of the miha bodytec II device must make sure that the device is used under such condi- tions.
  • Page 42 The miha bodytec II EMS training device has been designed for use in electromagnetic environments as outlined below. The buyer or user of the miha bodytec II device must make sure that the device is used under such condi- tions.
  • Page 43 The miha bodytec II EMS training device has been designed for use in electromagnetic environments as outlined below. The buyer or user of the miha bodytec II device must make sure that the device is used under such condi- tions.
  • Page 44 ; 250/300 cycles 0 % U ; 250/300 cycles ration during power mains interruptions, an IEC 61000-4-11 uninterruptible power supply (UPS) sufficiently fast and powerful to pro- vide the energy needed must be used. miha-bodytec II EMS training device 2021-07-28...
  • Page 45: Emc Warnings

    21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz. Replaceable parts: All cables and equipment must be obtained from miha bodytec or Ä Chapter authorized dealers. For replaceable parts available, see 1.2 “Equipment”...
  • Page 46 Portable RF communications equipment (including periph- erals such as antenna cables and external antennas) must be used no closer than 30 cm (12 inches) to any part of the miha- bodytec II EMS training device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment can result.
  • Page 47: Basic Information On Ems Training

    Positive and negative electrodes are not on the same muscle. Agonist and antagonist are stimulated simultaneously. EMS training can cause more intense muscle contractions than classic strength training. At the same time, there is compara- tively very little stress on the joints. 2021-07-28 miha-bodytec II EMS training device...
  • Page 48: Training Frequency And Regeneration

    Illness and training Just as with intense conventional workouts, EMS training may only be performed if the body is in adequate physical condition. miha-bodytec II EMS training device 2021-07-28...
  • Page 49 This means that the effects of EMS training on the body's creatine kinase and myo- globin levels are comparable to the effect of a very intense conven- tional muscular workout 2021-07-28 miha-bodytec II EMS training device...
  • Page 50 You should drink a high-protein energy drink such as a protein shake for better regeneration and for stimulation of muscle building. Take a break after the workout and do not complete any further workouts to guarantee best possible regeneration and to avoid overtraining. miha-bodytec II EMS training device 2021-07-28...
  • Page 51: Understanding The Device

    Creating a summary in the favorites menu also makes it possible to speed up the selection process, which facilitates the use of transponder cards. 2021-07-28 miha-bodytec II EMS training device...
  • Page 52 The brand code (parameterization) is used for selecting or adding default factory settings such as color settings, graphics settings, and network functions. This option may only be used by miha bodytec service employees or under their supervision. Risk of injury from improperly entering brand codes! L WARNING –...
  • Page 53 Channel 2 – buttocks Channel 3 – lower back Channel 4 – upper back Channel 5 – side back Channel 6 – abdomen Channel 7 – chest Channel 8 – arms Channel 9 Channel 10 2021-07-28 miha-bodytec II EMS training device...
  • Page 54: Opening Menu Tabs

    Turn the multi-function button counterclockwise to select the menu tab that is displayed to the left of the selected menu tab. In the example shown, this would be tab “A” (Fig. 21/A). Fig. 21: Selected menu tab miha-bodytec II EMS training device 2021-07-28...
  • Page 55: Opening Menus

    Color Example Meaning Black background Menu can be selected. Fig. 23: Example menu Gray background Menu is locked. Locked menus cannot be opened. Colored back- Menu is currently ground* selected. 2021-07-28 miha-bodytec II EMS training device...
  • Page 56 (example here “E1 è E2 è E3”). Press the multi-function button. Fig. 25: Setting has been selected ð The displayed submenu setting (here Fig. 26/E3) is applied. Fig. 26: Setting has been changed miha-bodytec II EMS training device 2021-07-28...
  • Page 57 With transponder card: Data is written on the card and the programs or training plans saved on the card are dis- played. The main menu is only displayed when the card is removed from the device. Fig. 28: Main menu 2021-07-28 miha-bodytec II EMS training device...
  • Page 58: Entering Text

    Select the letters by turning the individual level controllers (Fig. 29/1 – 10) (channels 1 – 10) and copy them into the respective text boxes by pressing the corresponding controllers. Copy the entered text by pressing the multi-function button (Fig. 29/11). miha-bodytec II EMS training device 2021-07-28...
  • Page 59: Customization And Setup

    “Setup” menu. Pause transitions can be used for instance to demonstrate to the athlete/patient when they have to contract their muscles. 7.2 Selecting the language The device lets you choose from the following menu language options: German English French Spanish Italian 2021-07-28 miha-bodytec II EMS training device...
  • Page 60: Program Memory Settings

    7.3 Program memory settings Displaying the program memory Personnel: Trainer Open the “Setup” menu tab. Open the “Program memory settings” main menu item Ä Further information on page 58. ð The “Program memory settings” opens. miha-bodytec II EMS training device 2021-07-28...
  • Page 61: Setting Up The Training Plan Memory

    Turn the multi-function button to the right to select the “Create training plan memory” menu item. ð The “Create training plan” menu item is selected. Confirm your selection by pressing the multi-function button. ð The “training plan” menu opens. 2021-07-28 miha-bodytec II EMS training device...
  • Page 62: Setting Up Synchronized Start

    2 steps are required to set up synchronized start: Ä “Setting up the master device” Setting up the master device on page 63 Ä “Setting up a managed device” Setting up managed devices on page 63 miha-bodytec II EMS training device 2021-07-28...
  • Page 63 All managed devices that are supposed to participate in the training session have to be set up as described below. Personnel: Trainer Open the “Setup” menu tab. ð The “Setup” menu tab opens. Open the “Synchronized start” menu item. 2021-07-28 miha-bodytec II EMS training device...
  • Page 64: Adjusting The Favorites Menu

    Confirm the assignment by pressing the multi-function button. Use the multi-function button to select the “Setup menu” menu item and press the button to confirm your selection. ð The “Favorites menu settings” menu is closed. miha-bodytec II EMS training device 2021-07-28...
  • Page 65: Backing Up Data

    Trainer Materials: miha bodytec USB flash drive “Backup” will not be enabled until you insert the miha bodytec USB flash drive into the device before switching to the “Setup” menu tab! Insert the miha bodytec USB flash drive into the USB port (Fig.
  • Page 66 Starting data restoration After selecting the data packages you wish to import, you can start restoring the data stored on the miha bodytec USB flash drive. Use the multi-function button to select the “Start process”...
  • Page 67: Installing Updates

    – Only use computers with updated anti-virus software when handling update files – Before copying the update file to the miha bodytec USB flash drive, scan the computer for viruses. Computer viruses can change the update files. This can result in errors in the control unit that affect the electrical outputs.
  • Page 68: Using Networks

    If the “Update” menu item has a grey background, the Fig. 32: USB port update file on the miha bodytec USB stick is defective, or the miha bodytec USB stick has not been fully inserted. Ensure that the update file is valid and that the miha bodytec USB flash drive is fully inserted.
  • Page 69: Resetting To Factory Settings

    Customization and setup miha bodytec LOGX portal The device’s network capability is used for communicating with the miha bodytec LOGX portal. The miha bodytec LOGX portal is a control center for commercial use. It is provided by miha bodytec. 7.11...
  • Page 70: Before The Training

    (low blood sugar). 8.2 Selecting undergarments and shoes Wearing miha bodytec undergar- The miha bodytec undergarments available as equipment have ments been specially developed for using the device with respect to mate- rial composition and strength. It is mandatory to wear the genuine Ä...
  • Page 71: Moistening And Applying Electrodes

    Ä “Selecting the size of the i-body with the instructions straps flex” on page 72. ® Wearing accessories that are too small or too large (i-body ® ® i-body straps flex, i-body belt) can lead to contact injuries. 2021-07-28 miha-bodytec II EMS training device...
  • Page 72 If the buttocks belt is loose, select a smaller size. If the buttocks belt is too tight, select a larger size or use an extension piece to enlarge it. miha-bodytec II EMS training device 2021-07-28...
  • Page 73 +86 °F. Only fill the pump spray bottle up to the max. marking visible on the side. Otherwise, no pressure can be built up. Pump air into the bottle by moving the pumping piston up and down (Fig. 33/1). 2021-07-28 miha-bodytec II EMS training device...
  • Page 74 Putting on the i-body ® Put on the i-body Close the “fitting aid” (hook and loop in the area of the chest and abdomen) in accordance with Fig. 35. ® Fig. 35: Wearing the i-body miha-bodytec II EMS training device 2021-07-28...
  • Page 75 However, never pull on the silicone on the side to smooth out the ® i-body ® Fig. 37: Smoothing out the i-body ® Putting on and closing the i-body straps flex (arms) 2021-07-28 miha-bodytec II EMS training device...
  • Page 76 Insert the hook into one of the loops (Fig. 40). Adjust the width so that it is comfortable for the athlete and at the same time provides enough contact pressure. ® Fig. 40: Closing the i-body straps flex miha-bodytec II EMS training device 2021-07-28...
  • Page 77 Fig. 42. ® Fig. 42: Connecting the i-body straps flex ® Putting on and closing the i-body straps flex (legs) ® Moisten 2 more i-body straps flex (Fig. 43). ® Fig. 43: i-body straps flex 2021-07-28 miha-bodytec II EMS training device...
  • Page 78 ® different i-body straps flex size. ® Put the i-body straps flex on the right leg in the same way as described for the left leg. ® Fig. 46: Positioning the i-body straps flex miha-bodytec II EMS training device 2021-07-28...
  • Page 79 (Fig. 49) with the i-body cable. ® When connecting the i-body belt, always ensure that the corresponding cables (shortest length) are used in accord- ance with Fig. 47. ® Fig. 49: Connecting the i-body belt 2021-07-28 miha-bodytec II EMS training device...
  • Page 80 ð All electrodes are moistened, applied, and connected as instructed. The connection cable is securely connected to ® the control unit or the i-body connect wireless. Training can now start. miha-bodytec II EMS training device 2021-07-28...
  • Page 81: Training With The Device

    – Never switch on the device if condensation is visible. Condensation can form in the event of changes in humidity and temperature. Moisture in the control unit can result in damage or even total destruction. 2021-07-28 miha-bodytec II EMS training device...
  • Page 82: Connecting The Device

    Connect the power cable to the power supply. ð The control unit can be mounted on the column. Fig. 53: Connecting the control unit miha-bodytec II EMS training device 2021-07-28...
  • Page 83: Individual Training Settings

    This is why these settings must always be determined by a trainer. Every person reacts differently to current intensity. Therefore, the current intensity must be regulated individually depending on the respective current sensitivity. 2021-07-28 miha-bodytec II EMS training device...
  • Page 84: Adjusting The Height Of The Grab Handle

    Tighten the clamping lever clockwise. Fig. 55: Adjusting the handle height The handle must not be loose after tightening and must sit firmly. Check to make sure that the screw connection is tight prior to every training session. miha-bodytec II EMS training device 2021-07-28...
  • Page 85: Overview Of Programs/Training Plans

    Pulse rise The pulse rise specifies in which time a current pulse increases to its maximum value. Rectangular pulses are preferred due to the rising slope of the pulse and the resulting effective stimulation. 2021-07-28 miha-bodytec II EMS training device...
  • Page 86: Training Programs

    The basic option (2) allows beginners to experience an adequate workout feeling by providing for a more gently rising pulse wave. When selecting the advanced option (3), users will experience an instantaneous pulse rise, resulting in an even stronger contraction of the muscles. miha-bodytec II EMS training device 2021-07-28...
  • Page 87: Training Plans

    The training plans provided by the device combine the training pro- Ä Chapter 11.3 “Training programs” on page 86) with spe- grams ( cially designed exercises. The sequences perfectly complement the intended training targets. 2021-07-28 miha-bodytec II EMS training device...
  • Page 88 Dynamic forward lunges, left Dynamic forward lunges, right Dynamic diagonal crunches, left Dynamic diagonal crunches, right Dynamic knee compression Dynamic body rotation, left Dynamic body rotation, right Dynamic crunches Dynamic forward fold Dynamic biceps miha-bodytec II EMS training device 2021-07-28...
  • Page 89 Dynamic butterfly reverse diag- onal, left Dynamic butterfly reverse diag- onal, right Dynamic body rotation, left Dynamic body rotation, right Dynamic diagonal crunches, left Dynamic diagonal crunches, right Dynamic crunches Dynamic overstretching Dynamic forward fold 2021-07-28 miha-bodytec II EMS training device...
  • Page 90 Dynamic forward lunges, right Dynamic side lunges, left Dynamic side lunges, right Dynamic knee compression Dynamic diagonal crunches, left Dynamic diagonal crunches, right Dynamic body rotation, left Dynamic body rotation, right Dynamic crunches Dynamic forward fold miha-bodytec II EMS training device 2021-07-28...
  • Page 91 Dynamic forward lunges, left Dynamic forward lunges, right Dynamic side lunges, left Dynamic side lunges, right Dynamic diagonal crunches, left Dynamic diagonal crunches, right Dynamic crunches Dynamic body rotation, left Dynamic body rotation, right Dynamic forward fold 2021-07-28 miha-bodytec II EMS training device...
  • Page 92: Training

    A basic requirement of EMS training is that the athlete/patient is tions? supervised by their trainer. The trainer can therefore intervene appropriately in the event of an emergency. The trainer must act immediately if the athlete/patient loses consciousness: miha-bodytec II EMS training device 2021-07-28...
  • Page 93 ð The main level jumps to “0” and all LEDs are flashing. The training time continues running during the pause. Turn the main controller to end the pause. ð The training session continues as usual. 2021-07-28 miha-bodytec II EMS training device...
  • Page 94 (Fig. 57/13). Adjust the application time if necessary by turning the multi- function button. Press the multi-function button to start the program. ð The time at the top right in the display starts running. miha-bodytec II EMS training device 2021-07-28...
  • Page 95: Transponder System

    The device comes with a transponder card reader to maximize operation convenience. The following personalized data can be saved on the transponder card: Program and/or training plan Time specification Last value of the main controller Last value of the individual level controllers 2021-07-28 miha-bodytec II EMS training device...
  • Page 96 “Card management” menu and save them to the transponder card using the “Write card” function. ð The transponder card is now formatted with the defined values and ready for use. Fig. 59: Transponder card miha-bodytec II EMS training device 2021-07-28...
  • Page 97 These user settings are linked to the transponder card associated with the athlete. This ensures that athletes will only participate in the training with synchronized start if they wish to do so. 2021-07-28 miha-bodytec II EMS training device...
  • Page 98 ð The transponder card of the athlete/patient is linked to an individual language setting. Display texts will be shown in the individually selected language when the transponder card has been inserted. Use the multi-function button to select “Write card” . miha-bodytec II EMS training device 2021-07-28...
  • Page 99: Training With Synchronized Start

    Start the workout by pressing the multi-function button on the master device once. ð The countdown begins on all the participating devices. The synchronized start was successful. 2021-07-28 miha-bodytec II EMS training device...
  • Page 100: After The Training

    Disconnect the connection between the control unit and the ® i-body (Fig. 60/1). Disconnect the connection between the ® ® i-body connect wireless and the i-body (Fig. 60/2). See the ® “i-body connect wireless” manual for further information. miha-bodytec II EMS training device 2021-07-28...
  • Page 101: Cleaning And Storage

    – Do not tumble dry. – Do not iron or treat with steam. – Do not spin-dry. – Do not wring out. ® Incorrect cleaning can cause material damage of the i-body electrode vest and individual electrodes. 2021-07-28 miha-bodytec II EMS training device...
  • Page 102 105 °F / 40 °C. If a washer with a hand wash setting is available, you can ® also wash the i-body straps flex in the washer using the supplied laundry net. miha-bodytec II EMS training device 2021-07-28...
  • Page 103: Cleaning The External Cables

    Incorrect cleaning can cause material damage of the cables. Only use a soft, lightly moistened cloth for cleaning the control unit connection cable and the cables for the external electrodes. Never use abrasives or solvents. 2021-07-28 miha-bodytec II EMS training device...
  • Page 104: Cleaning The Control Unit

    Dry all control unit surfaces with a soft, dry cloth. Wipe the LC display with a soft polishing cloth. ð All control unit surfaces are now dry and clean. This com- pletes the cleaning of the control unit. miha-bodytec II EMS training device 2021-07-28...
  • Page 105: Packaging And Storage

    And Information To humidity within the Be Supplied – Part 1: temperature limits a General Require- device may be ments exposed to for storage purposes Ä Chapter 13.2 “Transport and storage” on page 106. 2021-07-28 miha-bodytec II EMS training device...
  • Page 106: Transport And Storage

    Packaging materials are valuable raw materials. In many cases, they can be re-used or recycled for other uses. Envi- ronmental hazards can result from the incorrect disposal of packaging materials. miha-bodytec II EMS training device 2021-07-28...
  • Page 107 The following type of battery is installed: Lithium battery CR2032 (3 V, 220 mAh) The installed battery must not be disposed of via domestic waste. Return to the manufacturer for proper disposal. 2021-07-28 miha-bodytec II EMS training device...
  • Page 108: Ensuring Optimum Function

    Ä Chapter 10 “Adjusting the height of the grab handle” on page 84). Monthly Check to make sure the mounting screws between the base plate Trainer (Fig. 62/6) and the column (Fig. 62/5) are tight. Every 24 months Check the electrical system Licensed elec- trician miha-bodytec II EMS training device 2021-07-28...
  • Page 109: Maintaining The Electrodes

    If the error is no longer displayed, continue with the workout. If the error is displayed again or another error code is dis- played, notify the manufacturer or Customer Service and report the error codes ( Ä “Customer service” on page 15). 2021-07-28 miha-bodytec II EMS training device...
  • Page 110: Index

    Diabetes mellitus ......18 Dimensions ......34 miha-bodytec II EMS training device 2021-07-28...
  • Page 111 Storing ......103 Error messages ......109 2021-07-28 miha-bodytec II EMS training device...
  • Page 112 Having the device checked ... . . 108 Sending in the device ....108 miha-bodytec II EMS training device 2021-07-28...
  • Page 113 Rhabdomyolysis ......18 Training frequency ..... 47 2021-07-28 miha-bodytec II EMS training device...
  • Page 114 Body forming ......91 miha bodytec USB flash drive ... . 14 Fitness .

Table of Contents