55619 – PERISTALTIC PUMP SERVICE KIT
To suit Pool Lab ASP (Auto Sampling Photometer) and Pool Lab Hybrid Systems
acid and liquid chlorine feed peristaltic pumps. Replace every 12 months
PACKAGE CONTENTS
Qty
Part No. / Description
1
DRUM CAP ASSEMBLED
WITH TUBING AND WEIGHTS
CONSISTS OF:
5.0 meters x Part No. 55621
PERI PUMP CLEAR TUBING
1 x Part No. 55620
CAP ASSEMBLY
4 x Part No. 55617-CW
CERAMIC WEIGHT
1 x Part No. 55617-WR
WEIGHT RETAINER (RED)
1
32554-12-BLACK / TUBE & ROLLER KIT
1
55506-BLACK / INJECTOR VALVE
1
55615-LUB / 5g SILICONE LUBRICANT
Image
Rev 6

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 55619 and is the answer not in the manual?

Questions and answers

Summary of Contents for POOL LAB 55619

  • Page 1 55619 – PERISTALTIC PUMP SERVICE KIT To suit Pool Lab ASP (Auto Sampling Photometer) and Pool Lab Hybrid Systems acid and liquid chlorine feed peristaltic pumps. Replace every 12 months PACKAGE CONTENTS Part No. / Description Image DRUM CAP ASSEMBLED...
  • Page 2 RINSE ANY SPILLED CHEMICALS, OR TOOLS / CLOTHING / SKIN / EYES ETC. EXPOSED TO CHEMICALS. – THE POOL LAB PERISTALTIC PUMP HAS BEEN DESIGNED FOR USE WITH 16.5% HYDROCHLORIC ACID ONLY. USE OF HIGHER STRENGTH HYDROCHLORIC ACID (OR OTHER ACID TYPES) MAY DAMAGE EQUIPMENT, AND RESULT IN A SAFETY HAZARD.
  • Page 3 Hydrochloric (Muriatic) Acid Warnings ➔ Read warnings on the label. Do not allow acid to come in contact with skin, eyes or mouth, and follow directions on label if accidental contact occurs. ➔ Hydrochloric acid is highly corrosive, and will corrode metallic surfaces. ➔...
  • Page 4 Peristaltic Pump Service Procedure 1. Flush the feed pump and tubing with fresh water. The filtration system must be running to perform this task. Change filtration mode to ON if required, and use the ACID PRIME or LCHL PRIME function in the ASP TOOLS menu to flush water through the tubing. a) Loosen the fume stop cap to relieve any pressure in the drum.
  • Page 5 4. Use a cloth to wipe clean the pump housing and cover to remove all the old grease. 5. Install and lubricate new squeeze tube and roller parts, replace cover. a) Lubricate the backing plate where it contacts the inner rotor half. b) Install the inner rotor half (inner and outer rotor halves are identical parts).
  • Page 6 7. Replace feed tubing. a) Cut 2x lengths of new feed tubing as required: 1 x for pump to injector connection. 1 x for pump to reservoir connection b) Connect tubing from pump outlet to injector, and tighten lock nuts by hand only. While tightening the injector lock nut, hold the injector body in place with a spanner to prevent over-tightening the cell housing thread.